Dose reconstruction for real-time patient-specific dose estimation in CT

被引:11
|
作者
De Man, Bruno [1 ]
Wu, Mingye [2 ]
FitzGerald, Paul [3 ]
Kalra, Mannudeep [4 ,5 ]
Yin, Zhye [1 ]
机构
[1] GE Global Res, Image Reconstruct Lab, Niskayuna, NY 12309 USA
[2] GE Global Res, Xray & CT Lab, Shanghai 201203, Peoples R China
[3] GE Global Res, Radiat Syst Lab, Niskayuna, NY 12309 USA
[4] Massachusetts Gen Hosp, Div Thorac, Boston, MA 02114 USA
[5] Massachusetts Gen Hosp, Div Cardiac Imaging, Boston, MA 02114 USA
关键词
computed tomography; radiation dose estimation; MONTE-CARLO-CALCULATION; COMPUTED-TOMOGRAPHY; TUBE CURRENT; SUPERPOSITION/CONVOLUTION; MODULATION; RADIATION; NOISE;
D O I
10.1118/1.4921066
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Many recent computed tomography (CT) dose reduction approaches belong to one of three categories: statistical reconstruction algorithms, efficient x-ray detectors, and optimized CT acquisition schemes with precise control over the x-ray distribution. The latter category could greatly benefit from fast and accurate methods for dose estimation, which would enable real-time patient-specific protocol optimization. Methods: The authors present a new method for volumetrically reconstructing absorbed dose on a per-voxel basis, directly from the actual CT images. The authors' specific implementation combines a distance-driven pencil-beam approach to model the first-order x-ray interactions with a set of Gaussian convolution kernels to model the higher-order x-ray interactions. The authors performed a number of 3D simulation experiments comparing the proposed method to a Monte Carlo based ground truth. Results: The authors' results indicate that the proposed approach offers a good trade-off between accuracy and computational efficiency. The images show a good qualitative correspondence to Monte Carlo estimates. Preliminary quantitative results show errors below 10%, except in bone regions, where the authors see a bigger model mismatch. The computational complexity is similar to that of a low-resolution filtered-backprojection algorithm. Conclusions: The authors present a method for analytic dose reconstruction in CT, similar to the techniques used in radiation therapy planning with megavoltage energies. Future work will include refinements of the proposed method to improve the accuracy as well as a more extensive validation study. The proposed method is not intended to replace methods that track individual x-ray photons, but the authors expect that it may prove useful in applications where real-time patient-specific dose estimation is required. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:2740 / 2751
页数:12
相关论文
共 50 条
  • [21] Patient-specific dose calculations for pediatric CT of the chest, abdomen and pelvis
    Kost, Susan D.
    Fraser, Nicholas D.
    Carver, Diana E.
    Pickens, David R.
    Price, Ronald R.
    Hernanz-Schulman, Marta
    Stabin, Michael G.
    [J]. PEDIATRIC RADIOLOGY, 2015, 45 (12) : 1771 - 1780
  • [22] Patient-Specific Organ and Effective Dose Estimates in Adult Oncologic CT
    Gao, Yiming
    Mahmood, Usman
    Liu, Tianyu
    Quinn, Brian
    Gollub, Marc J.
    Xu, X. George
    Dauer, Lawrence T.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 214 (04) : 738 - 746
  • [23] Improvements on a patient-specific dose estimation system in nuclear medicine examination
    Chuang, K. S.
    Lu, J. C.
    Lin, H. H.
    Dong, S. L.
    Yang, H. J.
    Shih, C. T.
    Lin, C. H.
    Yao, W. J.
    Ni, Y. C.
    Jan, M. L.
    Chang, S. J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2014, 158 (01) : 1 - 7
  • [24] Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization
    Bacher, K
    Bogaert, E
    Lapere, R
    De Wolf, D
    Thierens, H
    [J]. CIRCULATION, 2005, 111 (01) : 83 - 89
  • [25] Patient-specific radiation dose and cancer risk estimation in CT: Part II. Application to patients
    Li, Xiang
    Samei, Ehsan
    Segars, W. Paul
    Sturgeon, Gregory M.
    Colsher, James G.
    Toncheva, Greta
    Yoshizumi, Terry T.
    Frush, Donald P.
    [J]. MEDICAL PHYSICS, 2011, 38 (01) : 408 - 419
  • [26] Real-time, acquisition parameter-free voxel-wise patient-specific Monte Carlo dose reconstruction in whole-body CT scanning using deep neural networks
    Yazdan Salimi
    Azadeh Akhavanallaf
    Zahra Mansouri
    Isaac Shiri
    Habib Zaidi
    [J]. European Radiology, 2023, 33 : 9411 - 9424
  • [27] Patient-Specific Radiation Dose and Cancer Risk Estimation in Pediatric Chest CT: A Study in 30 Patients
    Lia, Xiang
    Samei, Ehsan
    Segars, W. Paul
    Sturgeon, Gregory M.
    Colsher, James G.
    Frush, Donald P.
    [J]. MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [28] Real-time, acquisition parameter-free voxel-wise patient-specific Monte Carlo dose reconstruction in whole-body CT scanning using deep neural networks
    Salimi, Yazdan
    Akhavanallaf, Azadeh
    Mansouri, Zahra
    Shiri, Isaac
    Zaidi, Habib
    [J]. EUROPEAN RADIOLOGY, 2023, 33 (12) : 9411 - 9424
  • [29] The role of Size-Specific Dose Estimate (SSDE) in patient-specific organ dose and cancer risk estimation in paediatric chest and abdominopelvic CT examinations
    Franck, Caro
    Vandevoorde, Charlot
    Goethals, Ingeborg
    Smeets, Peter
    Achten, Eric
    Verstraete, Koenraad
    Thierens, Hubert
    Bacher, Klaus
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (08) : 2646 - 2655
  • [30] The role of Size-Specific Dose Estimate (SSDE) in patient-specific organ dose and cancer risk estimation in paediatric chest and abdominopelvic CT examinations
    Caro Franck
    Charlot Vandevoorde
    Ingeborg Goethals
    Peter Smeets
    Eric Achten
    Koenraad Verstraete
    Hubert Thierens
    Klaus Bacher
    [J]. European Radiology, 2016, 26 : 2646 - 2655