Monte Carlo-based QA for IMRT of head and neck cancers

被引:2
|
作者
Tang, F. [1 ,2 ]
Sham, J. [2 ]
Ma, C-M [3 ]
Li, J-S [2 ,3 ]
机构
[1] Queen Mary Hosp, Dept Clin Oncol, Pokfulam, Hong Kong, Peoples R China
[2] Univ Hong Kong, Dept Clin Oncol, Pokfulam, Peoples R China
[3] Fox Chase Canc Ctr, Dept Radiat Phys, Philadelphia, PA 19111 USA
关键词
D O I
10.1088/1742-6596/74/1/012021
中图分类号
O59 [应用物理学];
学科分类号
摘要
It is well-known that the presence of large air cavity in a dense medium (or patient) introduces significant electronic disequilibrium when irradiated with megavoltage X-ray field. This condition may worsen by the possible use of tiny beamlets in intensity-modulated radiation therapy (IMRT). Commercial treatment planning systems (TPSs), in particular those based on the pencil-beam method, do not provide accurate dose computation for the lungs and other cavity-laden body sites such as the head and neck. In this paper we present the use of Monte Carlo (MC) technique for dose re-calculation of IMRT of head and neck cancers. In our clinic, a turn-key software system is set up for MC calculation and comparison with TPS-calculated treatment plans as part of the quality assurance (QA) programme for IMRT delivery. A set of 10 off-the-self PCs is employed as the MC calculation engine with treatment plan parameters imported from the TPS via a graphical user interface (GUI) which also provides a platform for launching remote MC simulation and subsequent dose comparison with the TPS. The TPS-segmented intensity maps are used as input for the simulation hence skipping the time-consuming simulation of the multi-leaf collimator (MLC). The primary objective of this approach is to assess the accuracy of the TPS calculations in the presence of air cavities in the head and neck whereas the accuracy of leaf segmentation is verified by fluence measurement using a fluoroscopic camera-based imaging device. This measurement can also validate the correct transfer of intensity maps to the record and verify system. Comparisons between TPS and MC calculations of 6 MV IMRT for typical head and neck treatments review regional consistency in dose distribution except at and around the sinuses where our pencil-beam-based TPS sometimes over-predicts the dose by up to 10%, depending on the size of the cavities. In addition, dose re-buildup of up to 4% is observed at the posterior nasopharyngeal mucosa for some treatments with heavily-weighted anterior fields.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Monte Carlo-based dosimetry of head-and-neck patients treated with SIB-IMRT
    Sakthi, N
    Keall, P
    Mihaylov, I
    Wu, QW
    Wu, Y
    Williamson, JF
    Schmidt-Ullrich, R
    Siebers, JV
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (03): : 968 - 977
  • [2] An independent Monte Carlo-based IMRT QA tool for a 0.35 T MRI-guided linear accelerator
    Khan, Ahtesham Ullah
    Simiele, Eric A.
    Lotey, Rajiv
    DeWerd, Larry A.
    Yadav, Poonam
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (02):
  • [3] A Monte Carlo-based IMRT plan re-calculator
    Zakarian, K
    Alaly, J
    Low, D
    Wood, R
    Zhang, L
    Deasy, J
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1979 - 1979
  • [4] Clinical implementation of a Monte Carlo based QA process for IMRT
    Popescu, IA
    Shaw, C
    Cranmer-Sargison, G
    Wells, D
    Zavgorodni, S
    Mann, R
    Beckham, W
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S87 - S87
  • [5] Monte Carlo investigation of heterogeneity effect for head and neck IMRT
    Yang, J
    Li, J
    Chen, L
    Chen, Z
    Luo, W
    Fan, J
    Stathakis, S
    Price, R
    Ma, C
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2166 - 2166
  • [6] A prototype Monte Carlo-Based IMRT treatment planning research system
    Deasy, J
    Lee, E
    Kawrakow, I
    Zakarian, C
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1254 - 1254
  • [7] A GPU-Based Monte Carlo QA Tool for IMRT and VMAT
    Graves, Y.
    Kim, G.
    Folkerts, M.
    Teke, T.
    Popescu, I.
    Cervino, L.
    Tian, Z.
    Jia, X.
    Jiang, S.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3957 - 3958
  • [8] Clinically Applicable Monte Carlo-based Biological Dose Optimization for the Treatment of Head and Neck Cancers With Spot-Scanning Proton Therapy
    Tseung, Hok Seum Wan Chan
    Ma, Jiasen
    Kreofsky, Cole R.
    Ma, Daniel J.
    Beltran, Chris
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (05): : 1535 - 1543
  • [9] Monte Carlo evaluation of IMRT plans for head and neck and lung treatments
    Wang, L
    Yorke, E
    Chui, C
    [J]. RADIOLOGY, 2001, 221 : 600 - 601
  • [10] Monte Carlo-Based Patient Specific QA Procedure for Proteus One at UFHPTI
    Liu, C.
    Li, Z.
    Song, Y.
    Ho, M.
    Hsi, W.
    Park, J.
    Liang, X.
    Feng, H.