Pachymetric measurements with a new Scheimpflug photography-based system Intraobserver repeatability and agreement with optical coherence tomography pachymetry

被引:67
|
作者
Milla, Myriam [1 ]
Pinero, David P. [1 ,2 ]
Amparo, Francisco [1 ]
Alio, Jorge L. [1 ,3 ]
机构
[1] Vissum Corp, Inst Oftalmol Alicante, Alicante 03016, Spain
[2] Univ Alicante, Dept Opt, E-03080 Alicante, Spain
[3] Univ Miguel Hernandez, Div Ophthalmol, Alicante, Spain
来源
关键词
CORNEAL THICKNESS MEASUREMENTS; IN-SITU KERATOMILEUSIS; ULTRASOUND PACHYMETRY; ANTERIOR SEGMENT; OCULUS-PENTACAM; ORBSCAN-II; REPRODUCIBILITY; VISANTE; TOPOGRAPHY; MYOPIA;
D O I
10.1016/j.jcrs.2010.08.038
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the intraobserver repeatability of pachymetric measurements obtained with a new Scheimpflug photography-based system and the agreement of the measurements with those obtained by optical coherence tomography (OCT). SETTING: Vissum Corporation, Alicante, Spain. DESIGN: Evaluation of diagnostic technology. METHODS: Corneal thickness was measured at different locations (center; 2.5 mm and 4.0 mm nasal, temporal, superior, and inferior) in 18 eyes of 18 patients with no ocular pathology or previous surgery with a Scheimpflug photography-based system (Sirius). Three consecutive measurements at each location were obtained with the system to evaluate intraobserver repeatability. Afterward, pachymetric measurements were obtained at the same locations with an OCT system (Visante) to evaluate the agreement between techniques using the Bland-Altman method. RESULTS: The intraclass correlation coefficient of repeated measures ranged from 0.990 to 0.997. The coefficient of variation was lower than 1% and the standard deviations of the repeated measurements (S-w) were below 6 mu m at all corneal locations. Statistically significant differences were found between the coefficient of variation and S-w, values of the repeated measurements corresponding to the 2.5 mm and 4.0 mm locations (P <=.037). The agreement with OCT pachymetric measurements was poor, with ranges of agreement larger than 20 mu m at all corneal locations. Statistically significant differences were found between standard deviations of the differences between the 2 devices at the 2.5 mm and 4.0 mm locations (P <=.017). CONCLUSION: Scheimpflug photography-based technology provided repeatable pachymetric measurements that were not equivalent to those obtained with OCT.
引用
收藏
页码:310 / 316
页数:7
相关论文
共 50 条
  • [1] Evaluation of Pachymetric Measurements with Scheimpflug Photography-Based System and Optical Coherence Tomography Pachymetry at Different Stages of Keratoconus
    Akcay, Betul Ilkay Sezgin
    Ozgurhan, Engin Bilge
    Bozkurt, Ercument
    Kurt, Tugba
    Yildirim, Yusuf
    Cosar, Mediha Gulen
    Yildirim, Aydin
    Akkan, Julide Canan Umurhan
    Demirok, Ahmet
    JOURNAL OF OPHTHALMOLOGY, 2014, 2014
  • [2] Intrasubject repeatability of corneal morphology measurements obtained with a new Scheimpflug photography-based system
    Montalban, Raul
    Pinero, David P.
    Javaloy, Jaime
    Alio, Jorge L.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (06): : 971 - 977
  • [3] Intrasubject repeatability in keratoconus-eye measurements obtained with a new Scheimpflug photography-based system
    Montalban, Raul
    Alio, Jorge L.
    Javaloy, Jaime
    Pinero, David P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (02): : 211 - 218
  • [4] Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography: Repeatability, reproducibility, and agreement
    Wang, Qinmei
    Ding, Xixia
    Savini, Giacomo
    Chen, Hao
    Feng, Yifan
    Pan, Chao
    Hua, Yanjun
    Huang, Jinhai
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (01): : 178 - 185
  • [5] Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes
    Clara Ruiz-Belda
    David P. Piñero
    Pedro Ruiz-Fortes
    Roberto Soto-Negro
    Myriam Moya
    Rafael J Pérez-Cambrodí
    Alberto Artola
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2016, 254 : 169 - 175
  • [6] Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes
    Ruiz-Belda, Clara
    Pinero, David P.
    Ruiz-Fortes, Pedro
    Soto-Negro, Roberto
    Moya, Myriam
    Perez-Cambrodi, Rafael J.
    Artola, Alberto
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2016, 254 (01) : 169 - 175
  • [7] Repeatability and agreement of biometric measurements using spectral domain anterior segment optical coherence tomography and Scheimpflug tomography in keratoconus
    Li, Ye
    Gokul, Akilesh
    McGhee, Charles
    Ziaei, Mo
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2022, 49 (08): : 884 - 884
  • [8] Repeatability and agreement of biometric measurements using spectral domain anterior segment optical coherence tomography and Scheimpflug tomography in keratoconus
    Li, Ye
    Gokul, Akilesh
    McGhee, Charles
    Ziaei, Mohammed
    PLOS ONE, 2021, 16 (05):
  • [9] Intraobserver Repeatability of Tomographic, Pachymetric, and Anatomical Measurements in Healthy Eyes Using a New Swept-Source Optical Coherence Topographer
    Escolano Serrano, Jaime
    Tarazona Jaimes, Claudia Patricia
    Monera Lucas, Carlos Enrique
    Romero Valero, Daniel
    Moya Martinez, Alejandro
    Martinez Toldos, Jose Juan
    CORNEA, 2022, 41 (05) : 598 - 603
  • [10] Repeatability and Reproducibility of Slit Lamp, Optical Coherence Tomography, and Scheimpflug Measurements of Corneal Scars
    Das, Manoranjan
    Menda, Shivali A.
    Panigrahi, Arun K.
    Prajna, N. Venkatesh
    Yen, Michael
    Tsang, Betty
    Kumar, Alisha
    Rose-Nussbaumer, Jennifer
    Acharya, Nisha R.
    McCulloch, Charles E.
    Lietman, Thomas M.
    McLeod, Stephen D.
    Keenan, Jeremy D.
    OPHTHALMIC EPIDEMIOLOGY, 2019, 26 (04) : 251 - 256