Long-Term Durability of Bioprosthetic Aortic Valves: Implications From 12,569 Implants

被引:347
|
作者
Johnston, Douglas R.
Soltesz, Edward G.
Vakil, Nakul
Rajeswaran, Jeevanantham
Roselli, Eric E.
Sabik, Joseph F., III
Smedira, Nicholas G.
Svensson, Lars G.
Lytle, Bruce W.
Blackstone, Eugene H.
机构
[1] Cleveland Clin, Dept Thorac & Cardiovasc Surg, Inst Heart & Vasc, Cleveland, OH 44195 USA
[2] Cleveland Clin, Dept Quantitat Hlth Sci, Res Inst, Cleveland, OH 44195 USA
来源
ANNALS OF THORACIC SURGERY | 2015年 / 99卷 / 04期
关键词
EDWARDS PERICARDIAL BIOPROSTHESIS; PROSTHESIS-PATIENT MISMATCH; REPLACEMENT; SURVIVAL; AGE;
D O I
10.1016/j.athoracsur.2014.10.070
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Increased life expectancy and younger patients' desire to avoid lifelong anticoagulation requires a better understanding of bioprosthetic valve failure. This study evaluates risk factors associated with explantation for structural valve deterioration (SVD) in a long-term series of Carpentier-Edwards PERIMOUNT aortic valves (AV). Methods. From June 1982 to January 2011, 12,569 patients underwent AV replacement with Edwards Life-sciences Carpentier-Edwards PERIMOUNT stented bovine pericardial prostheses, models 2700PM (n [310) or 2700 (n [12,259). Mean age was 71 +/- 11 years (range, 18 to 98 years). 93% had native AV disease, 48% underwent concomitant coronary artery bypass grafting, and 26% had additional valve surgery. There were 81,706 patient-years of systematic follow-up data available for analysis. Demographics, intraoperative variables, and 27,386 echocardiographic records were used to identify risks for explant for SVD and assess longitudinal changes in transprosthesis gradients using time-varying covariable analyses. Results. Three hundred fifty-four explants were performed, with 41% related to endocarditis and 44% to SVD. Actuarial estimates of explant for SVD at 10 and 20 years were 1.9% and 15% overall, respectively, and in patients younger than 60 years, 5.6% and 46%, respectively. Younger age (p < 0.0001), lipid-lowering drugs (p = 0.002), prosthesis-patient mismatch (p = 0.001), and higher postoperative peak and mean AV gradients were associated with explant for SVD (p < 0.0001). The effect of gradient on SVD was greatest in patients younger than 60 years. Conclusions. Durability of the Carpentier-Edwards PERIMOUNT aortic valve is excellent even in younger patients. Explant for SVD is related to gradient at implantation, especially in younger patients. Strategies to reduce early postoperative AV gradients, such as root enlargement or more efficient prostheses, should be considered. (C) 2015 by The Society of Thoracic Surgeons
引用
收藏
页码:1239 / 1246
页数:8
相关论文
共 50 条
  • [21] Long-Term Mechanical Durability of Surgical Aortic Valves: A 2 Billion Cycle In Vitro Study
    Jae, Hyun Jae
    Bhat, Sanchita
    Karnik, Shweta
    Bagheri, Maryam
    Dasi, Lakshmi
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 84 (18) : B404 - B405
  • [22] Long-term durability of orthodontic mini-implants
    Abbassy, Mona Aly
    Bakry, Ahmed Samir
    Zawawi, Khalid Hashim
    Hassan, Ali Habib
    ODONTOLOGY, 2018, 106 (02) : 208 - 214
  • [23] Long-term durability of orthodontic mini-implants
    Mona Aly Abbassy
    Ahmed Samir Bakry
    Khalid Hashim Zawawi
    Ali Habib Hassan
    Odontology, 2018, 106 : 208 - 214
  • [24] Long-Term Durability of Transcatheter Valves The Importance of Accurate Data
    Butchart, Eric G.
    Chambers, John
    Borer, Jeffrey S.
    Grunkemeier, Gary
    Yoganathan, Ajit
    JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (02) : 253 - 256
  • [25] Risk-corrected impact of mechanical versus bioprosthetic valves on long-term mortality after aortic valve replacement
    Lund, Ole
    Bland, Martin
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2006, 132 (01): : 20 - U41
  • [26] THE LONG TERM DURABILITY OF EDWARDS SAPIEN TRANSCATHETER AORTIC VALVES: A HISTOPATHOLOGY STUDY
    Yahagi, Kazuyuki
    Lutter, Christoph
    Olson, Michelle
    Kutys, Robert
    Virmani, Renu
    Ladich, Elena
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (13) : 122 - 122
  • [27] Long Term Durability of Edwards SAPIEN Transcatheter Aortic Valves: A Pathology Study
    Ladich, Elena
    Yahagi, Kazuyuki
    Kutys, Robert
    Olson, Michelle
    Virmani, Renu
    LABORATORY INVESTIGATION, 2016, 96 : 82A - 82A
  • [28] Long Term Durability of Edwards SAPIEN Transcatheter Aortic Valves: A Pathology Study
    Ladich, Elena
    Yahagi, Kazuyuki
    Kutys, Robert
    Olson, Michelle
    Virmani, Renu
    MODERN PATHOLOGY, 2016, 29 : 82A - 82A
  • [29] COMPARISON OF LONG-TERM RESULTS OF CARPENTIER-EDWARDS AND HANCOCK BIOPROSTHETIC VALVES
    BOLOOKI, H
    KAISER, GA
    MALLON, SM
    PALATIANOS, GM
    ANNALS OF THORACIC SURGERY, 1986, 42 (05): : 494 - 499
  • [30] Long-Term Durability of Transcatheter Aortic Valve Prostheses
    Blackman, Daniel J.
    Saraf, Smriti
    MacCarthy, Philip A.
    Myat, Aung
    Anderson, Simon G.
    Malkin, Christopher J.
    Cunnington, Michael S.
    Somers, Kathryn
    Brennan, Paul
    Manoharan, Ganesh
    Parker, Jessica
    Aldalati, Omar
    Brecker, Stephen J.
    Dowling, Cameron
    Hoole, Stephen P.
    Dorman, Stephen
    Mullen, Michael
    Kennon, Simon
    Jerrum, Melanie
    Chandrala, Pavan
    Roberts, David H.
    Tay, Justin
    Doshi, Sagar N.
    Ludman, Peter F.
    Fairbairn, Timothy A.
    Crowe, Joanne
    Levy, Richard D.
    Banning, Adrian P.
    Ruparelia, Neil
    Spence, Mark S.
    Hildick-Smith, David
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (05) : 537 - 545