Objective: The purpose of this study was to determine if upper body (UB) and lower body (LB) bioelectrical impedance analysis (BIA) devices yield similar estimates of BF% as whole body (WB) BIA. Approach: Ninety-four men and women had their body composition evaluated using a Baseline 12-1190 Body Fat Scale (Fabrication Enterprises Inc.) as the LB BIA device, an Omron HBF-306C Handheld Body Fat Loss Monitor (Omron Healthcare, Inc.) as the UB BIA device, and a Seca Medical Body Composition Analyzer 514 (Seca North America) as the WB BIA device as well as via air displacement plethysmography (ADP; BodPod; Life Measurement Inc.). Main results: in women, the BF% was greater for WB BIA compared to UB BIA and LB BIA (28.9 5.0 versus 23.8 4.4 versus 20.4 5.9%; p < 0.001) with no differences between the WB BIA and ADP BF% (28.9 5.0 versus 28.3 7.5%; p = 0.982). In men, BF% was greater for WB BIA compared to UB BIA and LB BIA (19.1 6.7 versus 15.5 5.3 versus 14.9 4.0; p < 0.001) with no differences between the WB BIA and ADP BF% (19.1 6.7 versus 17.6 7.8; p = 0.394). Significance: Our findings suggest that the UB and LB BIA devices used in this study underestimate BF% compared to a WB BIA device and ADP in men and women.
引用
收藏
页数:5
相关论文
共 50 条
  • [31] Accuracy of Five Bioelectrical Impedance Analysis Instruments for the Measurement of Body Composition in College Females
    Barnes, Jeremy T.
    Pujol, Thomas J.
    Williams, Ronald D., Jr.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2008, 40 (05): : S271 - S271
  • [32] Bioelectrical impedance analysis for body composition assessment: reflections on accuracy, clinical utility, and standardisation
    Leigh C. Ward
    European Journal of Clinical Nutrition, 2019, 73 : 194 - 199
  • [33] PREDICTIVE ACCURACY OF BIOELECTRICAL IMPEDANCE EQUATIONS CORRECTED FOR FAT-FREE BODY SIZE
    WILSON, WL
    HEYWARD, VH
    COOK, KL
    HICKS, VL
    JENKINS, KA
    QUATROCHI, JA
    COLVILLE, BC
    AMERICAN JOURNAL OF HUMAN BIOLOGY, 1992, 4 (03) : 319 - 326
  • [34] Prognostic Significance of Bioelectrical Impedance Analysis Defined Body Composition in Upper Gastrointestinal Cancer
    Davies, Llion
    Blake, Paul A.
    Beamish, Andrew
    Karran, Alex
    Thomas, Charlotte E.
    Witherspoon, Jolene
    Howell, Gary
    Barlow, Rachael C.
    Lewis, Wyn G.
    GASTROENTEROLOGY, 2014, 146 (05) : S689 - S689
  • [35] Proposal of new body composition prediction equations from bioelectrical impedance for Indonesian men
    J Hastuti
    M Kagawa
    N M Byrne
    A P Hills
    European Journal of Clinical Nutrition, 2016, 70 : 1271 - 1277
  • [36] Proposal of new body composition prediction equations from bioelectrical impedance for Indonesian men
    Hastuti, J.
    Kagawa, M.
    Byrne, N. M.
    Hills, A. P.
    EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2016, 70 (11) : 1271 - 1277
  • [37] Total body water estimation using bioelectrical impedance: a meta-analysis of the data available in the literature
    R. Martinoli
    E. I. Mohamed
    C. Maiolo
    R. Cianci
    F. Denoth
    S. Salvadori
    L. Iacopino
    Acta Diabetologica, 2003, 40 : s203 - s206
  • [38] Total body water estimation using bioelectrical impedance: a meta-analysis of the data available in the literature
    Martinoli, R
    Mohamed, EI
    Maiolo, C
    Cianci, R
    Denoth, F
    Salvadori, S
    Iacopino, L
    ACTA DIABETOLOGICA, 2003, 40 (Suppl 1) : S203 - S206
  • [39] Accuracy of bioelectrical impedance spectroscopy in measuring changes in body composition during severe weight loss
    Cox-Reijven, PLM
    van Kreel, B
    Soeters, PB
    JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2002, 26 (02) : 120 - 127
  • [40] Accuracy of skinfold and bioelectrical impedance assessments of body fat percentage in ambulatory individuals with cerebral palsy
    Oeffinger, Donna J.
    Gurka, Matthew J.
    Kuperminc, Michelle
    Hassani, Sahar
    Buhr, Neeley
    Tylkowski, Chester
    DEVELOPMENTAL MEDICINE AND CHILD NEUROLOGY, 2014, 56 (05): : 475 - 481