A comparison of hydroxyapatite (HA)-coated threaded, HA-coated cylindric, and titanium threaded endosseous dental implants

被引:0
|
作者
Jeffcoat, MK
McGlumphy, EA
Reddy, MS
Geurs, NC
Proskin, HM
机构
[1] Univ Alabama, Sch Dent, Dept Periodont, Birmingham, AL 35294 USA
[2] Ohio State Univ, Coll Dent, Dept Prosthodont, Columbus, OH 43210 USA
[3] Howard Proskin Associates, Rochester, NY USA
关键词
endosseous dental implants; human clinical trials; hydroxyapatites;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study was to compare the success of hydroxyapatite (HA)-coated and machined titanium (Ti) implants in a 5-year randomized, controlled clinical trial conducted at 2 centers. Materials and Methods: Each of 120 edentulous patients received HA-coated threaded, HA-coated cylindric, and machined Ti threaded implants in a randomized design using 5 or 6 implants. Digital radiographs allowed for yearly measurements of bone loss. Calibrated clinicians also measured mobility, Gingival index, Plaque Index, probing depth, and recession. A Kaplan-Meier analysis was used to compare the proportion of ailing implants (defined as less than 2 mm of alveolar bone loss over 5 years) for each type of implant design. The criteria employed to assess implant outcome included the need for successful implants to lose less than 2 mm of bone support over the 5 years following placement of the prosthesis. Results: This analysis revealed that 95.2% of machined Ti threaded implants and 97.92% of HA-coated threaded implants were successful, while 99.0% of HA-coated cylindric implants experienced less than 2 mm of bone loss (P < .06). Discussion: All types of implants placed in this study had success rates above 95%. Conclusion: Over 5 years, the success rate tended to favor HA-coated implants.
引用
收藏
页码:406 / 410
页数:5
相关论文
共 50 条
  • [1] GUIDED TISSUE REGENERATION ON DEHISCED HA-COATED AND TITANIUM ENDOSSEOUS IMPLANTS
    ZABLOTSKY, MH
    MEFFERT, RM
    JOURNAL OF DENTAL RESEARCH, 1990, 69 : 347 - 347
  • [2] CHARACTERIZATION OF COMMERCIAL HA-COATED DENTAL IMPLANTS
    KIM, Y
    LEGEROS, J
    LEGEROS, R
    JOURNAL OF DENTAL RESEARCH, 1994, 73 : 137 - 137
  • [3] THE EFFECT OF SURFACE MACROTEXTURE ON HA-COATED DENTAL IMPLANTS
    COOK, SD
    SALKELD, SL
    JOURNAL OF DENTAL RESEARCH, 1994, 73 : 137 - 137
  • [4] ANALYSIS OF OSSEOINTEGRATION IN HA-COATED AND NON-HA COATED IMPLANTS
    MONTGOMERY, MT
    NICKELL, JC
    NUMMIKOSKI, P
    WATKINS, RT
    JONES, DL
    JOURNAL OF DENTAL RESEARCH, 1994, 73 : 138 - 138
  • [5] HA-Coated and non HA-coated implants: Periodontal responses to 48 months.
    Morris, HF
    Ochi, S
    Truhlar, RS
    Spray, JR
    JOURNAL OF DENTAL RESEARCH, 2000, 79 : 518 - 518
  • [6] Immediate loading of single threaded HA-coated root form implants at the maxillary premolar area.
    Proussaefs, P
    Lozada, J
    Kan, J
    Farnos, A
    Kleinman, A
    JOURNAL OF DENTAL RESEARCH, 2002, 81 : A393 - A393
  • [7] Microstructural changes in bone of HA-coated implants
    Ogiso, M
    Yamashita, Y
    Matsumoto, T
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1998, 39 (01): : 23 - 31
  • [8] Peri-implant tissue response of immediately loaded, threaded, HA-coated implants: 1-year results
    Rungcharassaeng, K
    Lozada, JL
    Kan, JYK
    Kim, JS
    Campagni, WV
    Munoz, CA
    JOURNAL OF PROSTHETIC DENTISTRY, 2002, 87 (02): : 173 - 181
  • [9] Histologic evaluation of threaded HA-coated root-form implants after 3.5 to 11 years of function: A report of three cases
    Proussaefs, P
    Lozada, J
    Ojano, M
    INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2001, 21 (01) : 21 - 29
  • [10] Ti implants with nanostructured and HA-coated surfaces for improved osseointegration
    Sirin, Hasret Tolga
    Vargel, Ibrahim
    Kutsal, Tulin
    Korkusuz, Petek
    Piskin, Erhan
    ARTIFICIAL CELLS NANOMEDICINE AND BIOTECHNOLOGY, 2016, 44 (03) : 1023 - 1030