Calibration of Near-Infrared Detectors Using a Wavelength Tunable Light Source

被引:6
|
作者
Maham, Kinza [1 ]
Vaskuri, Anna [1 ,3 ]
Manoocheri, Farshid [1 ]
Ikonen, Erkki [1 ,2 ]
机构
[1] Aalto Univ, Metrol Res Inst, POB 15500, FI-00076 Aalto, Finland
[2] VTT MIKES, POB 1000, FI-02044 Espoo, Finland
[3] NIST, Boulder, CO USA
关键词
Near-infrared detectors; Spectral responsivity; Supercontinuum laser; InGaAs; Ge; SUPERCONTINUUM LASER; IRRADIANCE; RADIOMETERS; FILTER; GE;
D O I
10.1007/s10043-020-00586-9
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
This paper presents the spectral responsivity calibrations of two indium gallium arsenide (InGaAs) and one germanium based near-infrared photovoltaic detectors using a wavelength tunable laser source based on a supercontinuum laser developed at the Metrology Research Institute, Aalto University. The setup consists of a supercontinuum laser based on a photonic crystal fiber as the light source, a laser line tunable filter, and coupling optics. These responsivity calibrations are performed against a pyroelectric radiometer over a wide spectral range of 800-2000 nm. Our wavelength tunable laser source has a high spectral power up to 2.5 mW with a narrow spectral full-width-at-half-maximum of 3 nm at a wavelength of 1100 nm. Despite the sharp spectral intensity variations, no artifacts are observed in the spectral responsivities of the detectors. Comparison of the spectral responsivities of the InGaAs detectors measured using the wavelength tunable laser and the earlier calibrations performed at the Metrology Research Institute in 2010 and 2016, shows that the higher spectral power of wavelength tunable light source decreases the expanded uncertainty from approximately 4% to 2.2-2.6% over the spectral range of 820-1600 nm. Temperature dependence of the spectral responsivities near the band gap edges are also measured and analysed.
引用
收藏
页码:183 / 189
页数:7
相关论文
共 50 条
  • [41] The System of Portable Fat Detector With Dual-wavelength Near-infrared Light
    Song, Weijun
    Zhang, Song
    Yang, Yimin
    Yang, Lin
    [J]. 2009 3RD INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING, VOLS 1-11, 2009, : 1254 - 1257
  • [42] Is Near-Infrared Light Neuroprotective?
    Bezard, Erwan
    [J]. ANNALS OF NEUROLOGY, 2016, 80 (02) : 310 - 310
  • [43] Ultrabright near-infrared light
    Andries Meijerink
    [J]. Light: Science & Applications, 13
  • [44] INFRARED NEAR-INFRARED SPECTROMETRY AND CHEMOMETRY - MULTIVARIATE CALIBRATION
    MENARDO, C
    [J]. ANALUSIS, 1993, 21 (10) : M32 - M34
  • [45] Monitoring brain activity using near-infrared light
    Fantini, S
    Aggarwal, P
    Chen, K
    Franceschini, MA
    [J]. AMERICAN LABORATORY, 2001, 33 (20) : 15 - +
  • [46] Digital reconstruction of teeth using near-infrared light
    Angelino, Keith
    Yauney, Gregory
    Rana, Aman
    Edlund, David
    Shah, Pratik
    [J]. 2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 4414 - 4418
  • [48] Ultrabright near-infrared light
    Meijerink, Andries
    [J]. LIGHT-SCIENCE & APPLICATIONS, 2024, 13 (01)
  • [49] Using Near-Infrared Ⅰ/Ⅱ Light to Regulate the Performance of Nanozymes
    Chaoyu Fan
    Jiachen Zhao
    Yonghua Tang
    Youhui Lin
    [J]. Journal of Analysis and Testing, 2023, 7 (03) : 272 - 284
  • [50] Measurement of Reciprocity Failure in Near-Infrared Detectors
    Biesiadzinski, T.
    Lorenzon, W.
    Newman, R.
    Schubnell, M.
    Tarle, G.
    Weaverdyck, C.
    [J]. PUBLICATIONS OF THE ASTRONOMICAL SOCIETY OF THE PACIFIC, 2011, 123 (900) : 179 - 186