Multi-projection fluorescence optical tomography using a handheld-probe-based optical imager: phantom studies

被引:5
|
作者
Ge, Jiajia [1 ]
Erickson, Sarah J. [1 ]
Godavarty, Anuradha [1 ]
机构
[1] Florida Int Univ, Opt Imaging Lab, Dept Biomed Engn, Miami, FL 33174 USA
基金
美国国家卫生研究院;
关键词
BREAST-CANCER DETECTION; PHYSIOLOGICAL-PROPERTIES; SPECTROSCOPY;
D O I
10.1364/AO.49.004343
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A handheld-probe-based optical imager has recently been developed toward three-dimensional tomography. In this study, the improvement of target depth recovery was demonstrated using a multi-projection technique on large slab phantoms using 0.45 cc fluorescing target(s) (with 1:0 contrast ratio) of 1.5 to 2.5 cm deep. Tomographic results using single-and multi-(here dual) projection measurements (with and without a priori information of target location) were compared. In all experimental cases, the use of multi- projection measurements along with a priori information recovered target depth and location closer to their true values, demonstrating its applicability for clinical translation. (C) 2010 Optical Society of America
引用
收藏
页码:4343 / 4354
页数:12
相关论文
共 50 条
  • [1] Fluorescence tomographic imaging using a handheld-probe-based optical imager: extensive phantom studies
    Ge, Jiajia
    Erickson, Sarah J.
    Godavarty, Anuradha
    [J]. APPLIED OPTICS, 2009, 48 (33) : 6408 - 6416
  • [2] Multi-projection based fluorescence optical tomography using a hand-held probe based optical imager
    Ge, Jiajia
    Erickson, Sarah J.
    Godavarty, Anuradha
    [J]. OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE VIII, 2009, 7174
  • [3] Handheld-probe-based optical Doppler tomography for blood flow imaging
    Jeon, Deokmin
    Ravichandran, Naresh Kumar
    Jung, Unsang
    Jeon, Mansik
    Kim, Jeehyun
    [J]. INFRARED PHYSICS & TECHNOLOGY, 2018, 95 : 183 - 188
  • [4] Deep 360° Optical Flow Estimation Based on Multi-projection Fusion
    Li, Yiheng
    Barnes, Connelly
    Huang, Kun
    Zhang, Fang-Lue
    [J]. COMPUTER VISION - ECCV 2022, PT XXXV, 2022, 13695 : 336 - 352
  • [5] Resolution of a Gen-2 handheld optical imager: diffuse and fluorescence imaging studies
    Roman, Manuela
    Gonzalez, Jean
    Carrasquilla, Jennifer
    Erickson, Sarah J.
    Akhter, Rad
    Godavarty, Anuradha
    [J]. APPLIED OPTICS, 2013, 52 (33) : 8060 - 8066
  • [6] Diagnostic imaging of breast cancer using fluorescence-enhanced optical tomography: phantom studies
    Godavarty, A
    Thompson, AB
    Roy, R
    Gurfinkel, M
    Eppstein, MJ
    Zhang, C
    Sevick-Muraca, EM
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (03) : 488 - 496
  • [7] MARGIN ASSESSMENT IN RENAL SURGERY USING A HANDHELD OPTICAL COHERENCE TOMOGRAPHY PROBE
    Ludwig, Wesley W.
    Wobker, Sara E.
    Gorin, Michael A.
    Ball, Mark W.
    Zysk, Adam M.
    Pierorazio, Philip M.
    Allaf, Mohamad E.
    [J]. JOURNAL OF UROLOGY, 2016, 195 (04): : E265 - E266
  • [8] Margin Assessment in Renal Surgery Using a Handheld Optical Coherence Tomography Probe
    Ludwig, Wesley W.
    Wobker, Sara E.
    Ball, Mark W.
    Zysk, Adam M.
    Yemul, Kiran S.
    Pierorazio, Philip M.
    Gorin, Michael A.
    Allaf, Mohamad E.
    [J]. UROLOGY, 2018, 113 : 241 - 245
  • [9] Noise filtration in fluorescence-enhanced optical tomography: breast phantom studies
    Zhu, Banghe
    Sevick-Muraca, Eva M.
    Eppstein, Margaret J.
    Godavarty, Anuradha
    [J]. INVERSE PROBLEMS IN SCIENCE AND ENGINEERING, 2009, 17 (01) : 97 - 104
  • [10] Imaging of molecular probe activity with Born-normalized fluorescence optical projection tomography
    Vinegoni, Claudio
    Fumene Feruglio, Paolo
    Cortez-Retamozo, Virna
    Razansky, Daniel
    Medoff, Benjamin D.
    Ntziachristos, Vasilis
    Sbarbati, Andrea
    Pittet, Mikael
    Weissleder, Ralph
    [J]. OPTICS LETTERS, 2010, 35 (07) : 1088 - 1090