Influence of ocular chromatic aberration and pupil size on transverse resolution in ophthalmic adaptive optics optical coherence tomography

被引:31
|
作者
Fernández, EJ
Drexler, W
机构
[1] Vienna Univ Med, Ctr Biomed Engn & Phys, Vienna, Austria
[2] Univ Murcia, Dept Phys, Lab Opt, Murcia, Spain
[3] Vienna Univ Med, Ctr Biomed Engn & Phys, Vienna, Austria
来源
OPTICS EXPRESS | 2005年 / 13卷 / 20期
关键词
D O I
10.1364/OPEX.13.008184
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Optical coherence tomography (OCT) enables visualization of the living human retina with unprecedented high axial resolution. The transverse resolution of existing OCT approaches is relatively modest as compared to other retinal imaging techniques. In this context, the use of adaptive optics (AO) to correct for ocular aberrations in combination with OCT has recently been demonstrated to notably increase the transverse resolution of the retinal OCT tomograms. AO is required when imaging is performed through moderate and large pupil sizes. A fundamental difference of OCT as compared to other imaging techniques is the demand of polychromatic light to accomplish high axial resolution. In ophthalmic OCT applications, the performance is therefore also limited by ocular chromatic aberrations. In the current work, the effects of chromatic and monochromatic ocular aberrations on the quality of retinal OCT tomograms, especially concerning transverse resolution, sensitivity and contrast, are theoretically studied and characterized. The repercussion of the chosen spectral bandwidth and pupil size on the final transverse resolution of OCT tomograms is quantitatively examined. It is found that losses in the intensity of OCT images obtained with monochromatic aberration correction can be up to 80 %, using a pupil size of 8 mm diameter in combination with a spectral bandwidth of 120 nm full width at half maximum for AO ultrahigh resolution OCT. The limits to the performance of AO for correction of monochromatic aberrations in OCT are established. The reduction of the detected signal and the resulting transverse resolution caused by chromatic aberration of the human eye is found to be strongly dependent on the employed bandwidth and pupil size. Comparison of theoretical results with experimental findings obtained in living human eyes is also provided. (c) 2005 Optical Society of America.
引用
收藏
页码:8184 / 8197
页数:14
相关论文
共 50 条
  • [1] Adaptive optics ultrahigh resolution optical coherence tomography
    Drexler, W
    Fernandez, EJ
    Hermann, B
    Unterhuber, B
    Sattmann, H
    Prieto, PM
    Fercher, AF
    Artal, P
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U929 - U929
  • [2] Improved cellular resolution retinal imaging with aberration free and pupil wander free reflective Adaptive Optics - Optical Coherence Tomography Imaging System
    Lee, Sang Hyuck
    Jian, Yifan
    Zawadzki, Robert
    Werner, John
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [3] Ultrahigh-resolution optical coherence tomography with monochromatic and chromatic aberration correction
    Zawadzki, Robert J.
    Cense, Barry
    Zhang, Yan
    Choi, Stacey S.
    Miller, Donald T.
    Werner, John S.
    OPTICS EXPRESS, 2008, 16 (11) : 8126 - 8143
  • [4] Adaptive-optics ultrahigh-resolution optical coherence tomography
    Hermann, B
    Fernández, EJ
    Unterhuber, A
    Sattmann, H
    Fercher, AF
    Drexler, W
    Prieto, PM
    Artal, P
    OPTICS LETTERS, 2004, 29 (18) : 2142 - 2144
  • [5] Three-dimensional adaptive optics ultrahigh resolution optical coherence tomography
    Fernandez, EJ
    Povazav, B
    Hermann, B
    Sattmann, H
    Unterhuber, A
    Leilgeb, R
    Prieto, PM
    Artal, P
    Dexler, W
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [6] Adaptive optics optical coherence tomography in glaucoma
    Dong, Zachary M.
    Wollstein, Gadi
    Wang, Bo
    Schuman, Joel S.
    PROGRESS IN RETINAL AND EYE RESEARCH, 2017, 57 : 76 - 88
  • [7] Ultrahigh-resolution ophthalmic optical coherence tomography
    Drexler, W
    Morgner, U
    Ghanta, RK
    Kärtner, FX
    Schuman, JS
    Fujimoto, JG
    NATURE MEDICINE, 2001, 7 (04) : 502 - 507
  • [8] Ultrahigh-resolution ophthalmic optical coherence tomography
    Wolfgang Drexler
    Uwe Morgner
    Ravi K. Ghanta
    Franz X. Kärtner
    Joel S. Schuman
    James G. Fujimoto
    Nature Medicine, 2001, 7 : 502 - 507
  • [9] Ultrahigh resolution ophthalmic imaging with optical coherence tomography
    Ghanta, RK
    Drexler, W
    Morgner, U
    Kärtner, F
    Ippen, EP
    Fujimoto, JG
    Schuman, JS
    Bursell, S
    BIOMEDICAL TOPICAL MEETINGS, TECHNICAL DIGEST, 2000, 38 : 47 - 49
  • [10] Retinal imaging with adaptive optics high speed and high resolution optical coherence tomography
    Zawadzki, RJ
    Laut, S
    Zhao, M
    Jones, S
    Olivier, S
    Izatt, JA
    Werner, JS
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46