Three-dimensional endoscopic OCT using sparse sampling with a miniature magnetic-driven scanning probe

被引:6
|
作者
Wang, Jie [1 ]
Hu, Yang [1 ]
Wu, Jigang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Univ Michigan Shanghai Jiao Tong Univ Joint Inst, Biophoton Lab, Shanghai 200240, Peoples R China
基金
国家高技术研究发展计划(863计划);
关键词
OPTICAL COHERENCE TOMOGRAPHY; HIGH-RESOLUTION;
D O I
10.1364/AO.57.010056
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We propose to apply sparse sampling and compressive sensing (CS) reconstruction in three-dimensional (3D) endoscopic optical coherence tomography (OCT) to reduce the amount of data required in the imaging process. We used a homemade miniature side-imaging magnetic-driven scanning probe with an outer diameter of 1.4 mm in a 1310 nm swept-source OCT system to acquire two-dimensional (2D) circumferential cross-sectional images of an ex vivo pigeon trachea sample. 3D imaging is then achieved by reconstruction from the multiple 2D images acquired while pulling the sample with a translation stage. Given a total translation distance, we achieved sparse sampling by randomizing the step sizes of the translation stage such that the total number of the acquired 2D frames was reduced compared with conventional 3D imaging with equally spaced step positions. We tested the CS reconstruction with reduced 2D frame numbers of 40%, 60%, and 80% compared with the case of equally spaced step positions. The results show that it is possible to recover reasonable OCT images using sparse sampling with CS reconstruction. Compared with the conventional equally spaced sampling method, our method provides a novel way for image acquisition and reconstruction that could significantly reduce the amount of 3D OCT imaging data, and thus the acquisition time. (C) 2018 Optical Society of America
引用
收藏
页码:10056 / 10061
页数:6
相关论文
共 50 条
  • [1] Miniaturized magnetic-driven scanning probe for endoscopic optical coherence tomography
    Pang, Ziwei
    Wu, Jigang
    [J]. BIOMEDICAL OPTICS EXPRESS, 2015, 6 (06): : 2231 - 2236
  • [2] High-speed submillimeter magnetic-driven rotational scanning side-imaging OCT probe
    Zhang, Chi
    Yang, Fan
    Wang, Fu
    Tan, Qingyue
    He, Hao
    Zhang, Su
    Wu, Jigang
    Tu, Shengxian
    [J]. JOURNAL OF BIOPHOTONICS, 2023, 16 (10)
  • [3] Three-dimensional time and Fourier domain endoscopic OCT using 2 axis scanning MEMS mirror
    Jung, Woonggyu
    McCormick, Daniel T.
    Ahn, Yeh-Chan
    Zhang, Jun
    Tien, Norman C.
    Chen, Zhongping
    [J]. COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE X, 2006, 6079
  • [4] High speed three-dimensional endoscopic OCT using MEMS technology
    Chen, Zhongping
    [J]. 2007 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, VOLS 1-4, 2007, : 170 - 170
  • [5] High speed three-dimensional endoscopic OCT using MEMS technology
    Chen, Zhongping
    Jung, Woonggyu
    Ahn, Yeh-Chan
    Sepehr, Ali
    Armstrong, William B.
    Brenner, Matt
    McCormick, Daniel T.
    Tien, Norman C.
    [J]. MOEMS AND MINIATURIZED SYSTEMS VI, 2007, 6466
  • [6] Automated In Vivo Navigation of Magnetic-Driven Microrobots Using OCT Imaging Feedback
    Li, Dongfang
    Dong, Dingran
    Lam, Wahshing
    Xing, Liuxi
    Wei, Tanyong
    Sun, Dong
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2020, 67 (08) : 2349 - 2358
  • [7] Three-dimensional endoscopic ultrasonography of gastrointestinal lesions using an ultrasound probe
    Nishimura, K
    Niwa, Y
    Goto, H
    Hase, S
    Arisawa, T
    Hayakawa, T
    [J]. SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1997, 32 (09) : 862 - 868
  • [8] Development of a three-dimensional vibrating tactile probe for miniature CMMs
    Claverley, James D.
    Leach, Richard K.
    [J]. PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2013, 37 (02): : 491 - 499
  • [9] Automatic three-dimensional segmentation of endoscopic airway OCT images
    Qi, Li
    Zheng, Kaibin
    Li, Xipan
    Feng, Qianjin
    Chen, Zhongping
    Chen, Wufan
    [J]. BIOMEDICAL OPTICS EXPRESS, 2019, 10 (02): : 642 - 656
  • [10] Three-dimensional Analysis of Nanomaterials by Scanning Probe Nanotomography
    Efimov, Anton E.
    Agapova, Olga I.
    Mochalov, Konstantin E.
    Agapov, Igor I.
    [J]. 4TH INTERNATIONAL CONFERENCE OF PHOTONICS AND INFORMATION OPTICS, PHIO 2015, 2015, 73 : 173 - 176