4D dose-position verification in radiation therapy using the RADPOS system in a deformable lung phantom

被引:21
|
作者
Cherpak, Amanda [1 ,2 ]
Serban, Monica [3 ]
Seuntjens, Jan [4 ]
Cygler, Joanna E. [1 ,2 ]
机构
[1] Ottawa Hosp Canc Ctr, Ottawa, ON K1H 8L6, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
[3] Hop Maisonneuve Rosemont Hosp, Serv Radiophys, Montreal, PQ H1T 2M4, Canada
[4] McGill Univ, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
4D in vivo dosimetry; electromagnetic positioning system; MOSFET; deformable phantom; RADPOS; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; VOLUME HISTOGRAM ANALYSIS; IN-VIVO DOSIMETRY; RESPIRATORY MOTION; CANCER PATIENTS; TECHNICAL NOTE; PNEUMONITIS; RADIOTHERAPY; PHOTON; REGISTRATION;
D O I
10.1118/1.3515461
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A novel 4D in vivo dosimetry system (RADPOS), in conjunction with a deformable lung phantom, has been evaluated as a potential quality assurance tool for 4D radiotherapy. Methods: RADPOS detectors, which consist of a MOSFET dosimeter combined with an electromagnetic positioning probe, were placed inside the deformable lung phantom. One detector was positioned directly inside a tumor embedded in the lung phantom and another was positioned inside the lung portion of the phantom, outside the tumor. CT scans were taken with the phantom at three breathing phases, and for each phase, the detector position inside the phantom was read with the RADPOS software and compared to the position as determined from the CT data. These values were also compared to RADPOS measurements taken with the phantom on the couch of a Varian Clinac 6EX linac. The deformable phantom and the RADPOS system were also used in two radiation delivery scenarios: (1) A simulation of a free-breathing delivery and (2) a simulation of an adaptive treatment. Results: Compared to CT imaging, the RADPOS positional accuracy was found to be better than 2.5 mm. The radial displacement measurements taken in the CT and linac rooms agreed to within an average of (0.7 +/- 0.3) mm. Hence, the system can provide relative displacement measurements in the treatment room, consistent with measurements made in the CT room. For the free-breathing delivery, the total dose reported by RADPOS agreed to within 4% and 5% of the treatment planning doses in the tumor and the lung portion of the phantom, respectively. The RADPOS-measured dose values for the adaptive delivery were within 1.5% of the treatment plan values, which was well within the estimated experimental uncertainties. Conclusions: This work has shown that the deformable lung phantom-RADPOS system can be an efficient quality assurance tool for 4D radiation therapy. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3515461]
引用
收藏
页码:179 / 187
页数:9
相关论文
共 50 条
  • [41] Validation of 4D CT Image Reconstruction Using a 3D Lung Phantom
    Zheng, Y.
    Girbino, M.
    Oshinsky, R.
    Vroege, L.
    Jesseph, F.
    Lee, S.
    Yuan, J.
    Machtay, M.
    Sohn, J.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [42] Monte Carlo Verification of Jaws-Only IMRT Dose Distributions From Prowess Panther System On Octavius 4D Phantom
    Tai, T.
    Son, N.
    Loan, T.
    Oanh, L.
    Chow, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E292 - E292
  • [43] 4D imaging for radiation therapy using MRI and PET
    Thorwarth, D.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S277 - S277
  • [44] SU-GG-J-69: Development of a Programmable Moving Phantom for the Verification of 4D Dose Delivery
    Yoon, J.
    Chung, J.
    Lee, J.
    Jung, W.
    Suh, T.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [45] A 4D IMRT Planning Method Using Deformable Image Registration for Lung Cancer
    Li, X.
    Wang, X.
    Li, Y.
    Pan, X.
    Mohan, R.
    Zhang, X.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [46] A Patient-Specific 4D Digital Phantom for Dynamic Dose Estimation of External Ionizing Radiation
    Deng, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3693 - 3693
  • [47] EVALUATION OF THE ARCCHECK 4D QA SYSTEM FOR THE VERIFICATION OF ARC THERAPY TREATMENT
    Edwards, C.
    Noble, R.
    Handley, J. C.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S534 - S535
  • [48] Accuracy of 4D Cone Beam CT and 4D Fan Beam CT for Thoracic Radiation Therapy Simulation and Verification Imaging
    Nyflot, M.
    Bowen, S. R.
    Lee, T.
    Meyer, J.
    Sandison, G. A.
    Kinahan, P. E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S640 - S640
  • [49] Compact System for 4D Treatment Planning and Dose Verification Based On Flexible Movement
    Nunes, C.
    Jimenez-Ortega, E.
    Linares-Doblado, R.
    Leal Plaza, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E947 - E947
  • [50] Validation of proton dose calculation on scatter corrected 4D cone beam computed tomography using a porcine lung phantom
    Schmitz, Henning
    Rabe, Moritz
    Janssens, Guillaume
    Bondesson, David
    Rit, Simon
    Parodi, Katia
    Belka, Claus
    Dinkel, Julien
    Kurz, Christopher
    Kamp, Florian
    Landry, Guillaume
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (17):