4D dose-position verification in radiation therapy using the RADPOS system in a deformable lung phantom

被引:21
|
作者
Cherpak, Amanda [1 ,2 ]
Serban, Monica [3 ]
Seuntjens, Jan [4 ]
Cygler, Joanna E. [1 ,2 ]
机构
[1] Ottawa Hosp Canc Ctr, Ottawa, ON K1H 8L6, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
[3] Hop Maisonneuve Rosemont Hosp, Serv Radiophys, Montreal, PQ H1T 2M4, Canada
[4] McGill Univ, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
4D in vivo dosimetry; electromagnetic positioning system; MOSFET; deformable phantom; RADPOS; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; VOLUME HISTOGRAM ANALYSIS; IN-VIVO DOSIMETRY; RESPIRATORY MOTION; CANCER PATIENTS; TECHNICAL NOTE; PNEUMONITIS; RADIOTHERAPY; PHOTON; REGISTRATION;
D O I
10.1118/1.3515461
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A novel 4D in vivo dosimetry system (RADPOS), in conjunction with a deformable lung phantom, has been evaluated as a potential quality assurance tool for 4D radiotherapy. Methods: RADPOS detectors, which consist of a MOSFET dosimeter combined with an electromagnetic positioning probe, were placed inside the deformable lung phantom. One detector was positioned directly inside a tumor embedded in the lung phantom and another was positioned inside the lung portion of the phantom, outside the tumor. CT scans were taken with the phantom at three breathing phases, and for each phase, the detector position inside the phantom was read with the RADPOS software and compared to the position as determined from the CT data. These values were also compared to RADPOS measurements taken with the phantom on the couch of a Varian Clinac 6EX linac. The deformable phantom and the RADPOS system were also used in two radiation delivery scenarios: (1) A simulation of a free-breathing delivery and (2) a simulation of an adaptive treatment. Results: Compared to CT imaging, the RADPOS positional accuracy was found to be better than 2.5 mm. The radial displacement measurements taken in the CT and linac rooms agreed to within an average of (0.7 +/- 0.3) mm. Hence, the system can provide relative displacement measurements in the treatment room, consistent with measurements made in the CT room. For the free-breathing delivery, the total dose reported by RADPOS agreed to within 4% and 5% of the treatment planning doses in the tumor and the lung portion of the phantom, respectively. The RADPOS-measured dose values for the adaptive delivery were within 1.5% of the treatment plan values, which was well within the estimated experimental uncertainties. Conclusions: This work has shown that the deformable lung phantom-RADPOS system can be an efficient quality assurance tool for 4D radiation therapy. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3515461]
引用
收藏
页码:179 / 187
页数:9
相关论文
共 50 条
  • [1] PERFORMANCE OF THE RADPOS DOSE-POSITION VERIFICATION SYSTEM IN 4D RADIOTHERAPY USING A DEFORMABLE LUNG PHANTOM
    Cygler, J.
    Cherpak, A.
    Monica, S.
    Jan, S.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S35 - S35
  • [2] Dose - Position Verification of 4D Radiotherapy Using the RADPOS System in a Deformable Lung Phantom
    Cherpak, A.
    Serban, M.
    Seuntjens, J.
    Cygler, J.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [3] Deformable Lung Phantom for Verification of 4D Dose Calculations
    Vinogradskiy, Y.
    Starkschall, G.
    Balter, P.
    Followill, D.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [4] Dose and Position Quality Assurance using the RADPOS System for 4D Radiotherapy with CyberKnife
    Marants, R.
    Vandervoort, E.
    Cygler, J. E.
    MEDICAL PHYSICS, 2014, 41 (08) : 11 - 11
  • [5] A novel deformable lung phantom for 4D radiotherapy verification
    Margeanu, M.
    Stroian, G.
    Heath, E.
    Seuntjens, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2649 - 2649
  • [6] Dose and Position Quality Assurance Using the RADPOS System for 4D Radiotherapy with CyberKnife
    Marants, R.
    Vandervoort, E.
    Cygler, J. E.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 599 - 602
  • [7] Development of a Deformable Dosimetric Phantom for Verification of 4D Dose Calculation Algorithms
    Zhong, H.
    Glide-Hurst, C.
    Li, H.
    Nurushev, T.
    Movsas, B.
    Chetty, I.
    MEDICAL PHYSICS, 2012, 39 (06) : 3956 - 3956
  • [8] Evaluation of the 4D RADPOS dosimetry system for dose and position quality assurance of CyberKnife
    Marants, Raanan
    Vandervoort, Eric
    Cygler, Joanna E.
    MEDICAL PHYSICS, 2018, 45 (09) : 4030 - 4044
  • [9] Verification of 4D Dose Delivery Using 4D Digital Human Phantom
    Zhang, F.
    Qin, Y.
    Segars, W.
    Yin, F.
    Cai, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3859 - 3859
  • [10] Validation of 4D Monte Carlo dose calculations using a programmable deformable lung phantom
    Gholampourkashi, Sara
    Cygler, Joanna E.
    Lavigne, Bernie
    Heath, Emily
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 76 : 16 - 27