4D digitally reconstructed radiography for verifying a lung tumor position during volumetric modulated arc therapy

被引:8
|
作者
Nakagawa, Keiichi [1 ]
Kida, Satoshi [1 ]
Haga, Akihiro [1 ]
Masutani, Yoshitaka [1 ]
Yamashita, Hideomi [1 ]
Onoe, Tsuyoshi [1 ]
Imae, Toshikazu [1 ]
Tanaka, Kenichiro [1 ]
Ohtomo, Kuni [1 ]
Yoda, Kiyoshi [2 ]
机构
[1] Tokyo Univ Hosp, Dept Radiol, Bunkyo Ku, Tokyo 1138655, Japan
[2] Elekta KK, Minato Ku, Tokyo 1080023, Japan
关键词
DRR; VMAT; 4D; in-treatment; BEAM COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; CT;
D O I
10.1093/jrr/rrs013
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
We have proposed four dimensional (4D) digitally reconstructed radiography (DRR) for verifying a lung tumor position during volumetric modulated arc therapy (VMAT). An internal target volume (ITV) was defined based on two clinical target volumes (CTVs) delineated on maximum exhalation and maximum inhalation images acquired by 4D planning computed tomography (CT). A planning target volume (PTV) was defined by adding a margin of 5 mm to the ITV on the maximum exhalation 3D CT images. A single-arc VMAT plan was created on the same CT data using Pinnacle Smart Arc with a maximum multi-leaf collimator leaf speed of 1 mm/degree, thereby resulting in quasi-conformal field shapes while optimizing each beam intensity for each gantry angle. During VMAT delivery, cone-beam CT (CBCT) projection data were acquired by an on-board kilovoltage X-ray unit and a flat panel 2D detector. Four CBCT image sets with different respiratory phases were reconstructed using in-house software, where respiratory phases were extracted from the projection data. Subsequently a CTV was delineated on each of the 4D CBCT images by an oncologist. Using the resulting 4D CBCT data including the CTV contours, 4D DRRs during the VMAT delivery were calculated as a function of gantry angle. It was confirmed that the contoured CTV was within the radiation field during the four-fraction lung VMAT delivery. The proposed 4D DRR may facilitate the verification of the position of a respiratory moving lung tumor during VMAT delivery on each treatment day.
引用
收藏
页码:628 / 632
页数:5
相关论文
共 50 条
  • [1] 4D registration and 4D verification of lung tumor position for stereotactic volumetric modulated arc therapy using respiratory-correlated cone-beam CT
    Nakagawa, Keiichi
    Haga, Akihiro
    Kida, Satoshi
    Masutani, Yoshitaka
    Yamashita, Hideomi
    Takahashi, Wataru
    Sakumi, Akira
    Saotome, Naoya
    Shiraki, Takashi
    Ohtomo, Kuni
    Iwai, Yoshio
    Yoda, Kiyoshi
    [J]. JOURNAL OF RADIATION RESEARCH, 2013, 54 (01) : 152 - 156
  • [2] 3D Lung Tumor Position Verification During Breath-Hold Stereotactic Volumetric Modulated Arc Therapy
    Hazelaar, C.
    Dahele, M.
    Mostafavi, H.
    van der Weide, L.
    Keek, S.
    Slotman, B.
    Verbakel, W.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3224 - 3224
  • [3] A Formulation of 4D Treatment Planning for Tumour Tracking Volumetric Modulated Arc Therapy for Lung Cancer
    Cheung, Michael L. M.
    Lee, Louis K. Y.
    Chan, Anthony T. C.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3645 - 3645
  • [4] Inverse planning for four-dimensional (4D) volumetric modulated arc therapy
    Ma, Yunzhi
    Chang, Daniel
    Keall, Paul
    Xie, Yiaoqin
    Park, Jae-yoon
    Suh, Tae-Suk
    Xing, Lei
    [J]. MEDICAL PHYSICS, 2010, 37 (11) : 5627 - 5633
  • [5] Volumetric modulated arc therapy quality assurance: Characterisation and evaluation of Octavius 4D system
    Lafond, C.
    Irazola, L.
    Manens, J. P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S150 - S150
  • [6] Effect of MLC tracking latency on conformal volumetric modulated arc therapy (VMAT) plans in 4D stereotactic lung treatment
    Bedford, James L.
    Fast, Martin F.
    Nil, Simeon
    McDonald, Fiona M. A.
    Ahmed, Merina
    Hansen, Vibeke N.
    Oelfke, Uwe
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 117 (03) : 491 - 495
  • [7] Motion compensated volumetric modulated are therapy (4D VMAT)
    Keshet, O.
    Lee, L.
    Ma, Y.
    Xing, L.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S605 - S605
  • [8] IN-TREATMENT DOSE CALCULATION USING INTREATMENT 4D KILOVOLTAGE CONE-BEAM CT AND IN-TREATMENT LINAC PARAMETERS DURING VOLUMETRIC MODULATED ARC THERAPY FOR A LUNG TUMOR
    Sakumi, A.
    Haga, A.
    Satoshi, K.
    Naoya, S.
    Yoda, K.
    Nakagawa, K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S125 - S125
  • [9] Planning 4D intensity-modulated arc therapy for tumor tracking with a multileaf collimator
    Niu, Ying
    Betzel, Gregory T.
    Yang, Xiaocheng
    Gui, Minzhi
    Parke, William C.
    Yi, Byongyong
    Yu, Cedric X.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (04): : 1480 - 1500
  • [10] Evaluation of delivered dose to a moving target by 4D dose reconstruction in gated volumetric modulated arc therapy
    Chung, Hyekyun
    Jung, Jinhong
    Jeong, Chiyoung
    Kwak, Jungwon
    Park, Jin-hong
    Kim, Su Ssan
    Yoon, Sang Min
    Song, Si Yeol
    Kim, Jong Hoon
    Choi, Eun Kyung
    Cho, Seungryong
    Cho, Byungchul
    [J]. PLOS ONE, 2018, 13 (09):