Objective Assessment System for Hearing Prediction Based on Stimulus-Frequency Otoacoustic Emissions

被引:5
|
作者
Gong, Qin [1 ,2 ]
Liu, Yin [1 ]
Xu, Runyi [1 ]
Liang, Dong [1 ]
Peng, Zewen [1 ]
Yang, Honghao [1 ]
机构
[1] Tsinghua Univ, Dept Biomed Engn, Beijing, Peoples R China
[2] Shanghai Univ, Sch Med, Shanghai, Peoples R China
来源
TRENDS IN HEARING | 2021年 / 25卷
基金
中国国家自然科学基金;
关键词
stimulus-frequency otoacoustic emissions; objective assessment system; threshold prediction; hearing screening; SUPPRESSION; PERIODICITY; THRESHOLDS; LATENCY; HUMANS; LEVEL;
D O I
10.1177/23312165211059628
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Stimulus-frequency otoacoustic emissions (SFOAEs) can be useful tools for assessing cochlear function noninvasively. However, there is a lack of reports describing their utility in predicting hearing capabilities. Data for model training were collected from 245 and 839 ears with normal hearing and sensorineural hearing loss, respectively. Based on SFOAEs, this study developed an objective assessment system consisting of three mutually independent modules, with the routine test module and the fast test module used for threshold prediction and the hearing screening module for identifying hearing loss. Results evaluated via cross-validation show that the routine test module and the fast test module predict hearing thresholds with similar performance from 0.5 to 8 kHz, with mean absolute errors of 7.06-11.61 dB for the routine module and of 7.40-12.60 dB for the fast module. However, the fast module involves less test time than is needed in the routine module. The hearing screening module identifies hearing status with a large area under the receiver operating characteristic curve (0.912-0.985), high accuracy (88.4-95.9%), and low false negative rate (2.9-7.0%) at 0.5-8 kHz. The three modules are further validated on unknown data, and the results are similar to those obtained through cross-validation, indicating these modules can be well generalized to new data. Both the routine module and fast module are potential tools for predicting hearing thresholds. However, their prediction performance in ears with hearing loss requires further improvement to facilitate their clinical utility. The hearing screening module shows promise as a clinical tool for identifying hearing loss.
引用
收藏
页数:19
相关论文
共 50 条
  • [41] MEASUREMENT OF OTOACOUSTIC EMISSIONS FOR HEARING ASSESSMENT
    WHITEHEAD, ML
    STAGNER, BB
    LONSBURYMARTIN, BL
    MARTIN, GK
    [J]. IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1994, 13 (02): : 210 - 226
  • [42] Newborn Hearing Screening System Based on Otoacoustic Emissions
    Sanchez Galego, Juliet
    Rodriguez Davila, Ernesto
    Regueiro Gomez, Angel
    [J]. REVISTA CUBANA DE INGENIERIA, 2010, 1 (03): : 5 - 13
  • [43] Loss of the Tectorial Membrane Protein CEACAM16 Enhances Spontaneous, Stimulus-Frequency, and Transiently Evoked Otoacoustic Emissions
    Cheatham, Mary Ann
    Goodyear, Richard J.
    Homma, Kazuaki
    Legan, P. Kevin
    Korchagina, Julia
    Naskar, Souvik
    Siegel, Jonathan H.
    Dallos, Peter
    Zheng, Jing
    Richardson, Guy P.
    [J]. JOURNAL OF NEUROSCIENCE, 2014, 34 (31): : 10325 - 10338
  • [44] Stimulus-Frequency Otoacoustic Emission Suppression Tuning in Humans: Comparison to Behavioral Tuning
    Karolina K. Charaziak
    Pamela Souza
    Jonathan H. Siegel
    [J]. Journal of the Association for Research in Otolaryngology, 2013, 14 : 843 - 862
  • [45] Stimulus-frequency otoacoustic emission: Measurements in humans and simulations with an active cochlear model
    Choi, Yong-Sun
    Lee, Soo-Young
    Parham, Kourosh
    Neely, Stephen T.
    Kim, Duck O.
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2008, 123 (05): : 2651 - 2669
  • [46] Nonlinear reflection as a cause of the short-latency component in stimulus-frequency otoacoustic emissions simulated by the methods of compression and suppression
    Vencovsky, Vaclav
    Vetesnik, Ales
    Gummer, Anthony W.
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2020, 147 (06): : 3992 - 4008
  • [47] Analyzing Stimulus-frequency Otoacoustic Emission Fine Structure Using an Additive Model
    Liu, Yin
    Ji, Fei
    Gong, Qin
    [J]. 42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 960 - 963
  • [48] Stimulus-Frequency Otoacoustic Emission Suppression Tuning in Humans: Comparison to Behavioral Tuning
    Charaziak, Karolina K.
    Souza, Pamela
    Siegel, Jonathan H.
    [J]. JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2013, 14 (06): : 843 - 862
  • [49] Time-frequency analyses of transient-evoked stimulus-frequency and distortion-product otoacoustic emissions: Testing cochlear model predictions
    Konrad-Martin, D
    Keefe, DH
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2003, 114 (04): : 2021 - 2043
  • [50] A direct method of estimation of stimulus frequency otoacoustic emissions