摘要
Most QCM-based devices devoted to PM measurement use substrates of various kinds to collect and retain particles. The novelty of this work is the possibility to use an integrated microheater actuator to change the substrate collecting features during the particulate adsorption. In fact, this paper presents preliminary results about the exploitation of a HQCM (Heated Quartz Crystal Microbalance) in particulate matter (PM) measurements. HQCM is a 10 MHz quartz crystal microbalance with a microheater and a resistive temperature detector (RTD) implemented on its surface. The effects of the sensor surface temperature (T-1 = 20 degrees C and T-2 = 80 degrees C) on the PM detection have been studied by a suitable particulate measurement system, where the HQCM coated by grease, worked both as impaction plate and mass sensor. HQCM was tested providing to it a sequence of impulses of microsphere of Cerium (IV) oxide dispersed in synthetic air (715 mu g/m(3) of PM2.5 and 43 mu g/m(3) of PM1, lasting 1 min) and recording the consequent resonant frequency shift. The results have shown that the HQCM was able to detect PM both at T = 20 degrees C and at 80 degrees C. Grease coated HQCM exhibited a good frequency stability at T = 80 degrees C (+/- 5 Hz) with a negligible grease mass loss during several heating cycles. Furthermore, the results have shown that the surface temperature changes the collection capacity. In particular at higher temperature (80 degrees C) the frequency change was 2.5 times greater than that at lower temperature (20 degrees C). (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:205 / 211
页数:7
相关论文
共 50 条
  • [31] In-stack condensible particulate matter measurements and issues
    Corio, LA
    Sherwell, J
    JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 2000, 50 (02): : 207 - 218
  • [32] An optimally designed virtual impactor integrated with a quartz crystal microbalance sensor for submicron particulate matter classification and detection
    Wang, Yong
    Meng, Hui
    Hu, Luoke
    Leng, Longlong
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2025, 96 (02):
  • [33] Low-Cost Sensor Node for Air Quality Monitoring: Field Tests and Validation of Particulate Matter Measurements
    Schilt, Ueli
    Barahona, Braulio
    Buck, Roger
    Meyer, Patrick
    Kappani, Prince
    Mockli, Yannis
    Meyer, Markus
    Schuetz, Philipp
    SENSORS, 2023, 23 (02)
  • [34] Effects of Road Traffic on the Accuracy and Bias of Low-Cost Particulate Matter Sensor Measurements in Houston, Texas
    Oluwadairo, Temitope
    Whitehead, Lawrence
    Symanski, Elaine
    Bauer, Cici
    Carson, Arch
    Han, Inkyu
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2022, 19 (03)
  • [35] Unmanned Aerial Vehicle-Borne Sensor System for Atmosphere-Particulate-Matter Measurements: Design and Experiments
    Wang, Tonghua
    Han, Wenting
    Zhang, Mengfei
    Yao, Xiaomin
    Zhang, Liyuan
    Peng, Xingshuo
    Li, Chaoqun
    Dan, Xvjia
    SENSORS, 2020, 20 (01)
  • [36] Integrated and microheater embedded gas sensor array based on the polymer composites dispensed in micromachined wells
    Ha, SC
    Kim, YS
    Yang, Y
    Kim, YJ
    Cho, SM
    Yang, H
    Kim, YT
    SENSORS AND ACTUATORS B-CHEMICAL, 2005, 105 (02) : 549 - 555
  • [37] An Integrated ECT and Electrostatic Sensor for Particulate Flow Measurement
    Long, Jun
    Frias, Marco A. Rodriguez
    Mokhtar, Khursiah Zainal
    Yang, Wuqiang
    2018 IEEE INTERNATIONAL CONFERENCE ON IMAGING SYSTEMS AND TECHNIQUES (IST), 2018, : 386 - 390
  • [38] The simulation and experimental research of particulate matter sensor on diesel engine with diesel particulate filter
    Tang, Dong
    Zhao, Ruonan
    Wang, Songhua
    Wang, Jiawei
    Ni, Li
    Chen, Lie
    SENSORS AND ACTUATORS A-PHYSICAL, 2017, 259 : 160 - 170
  • [39] QCM Sensor System Based on a Phase Detector Circuit for Measurements of Density and Viscosity of Liquids
    Costas, J.
    Rodriguez-Pardo, L.
    Cao-Paz, A.
    Farina, J.
    Perrot, H.
    Rose, D.
    2021 JOINT CONFERENCE OF THE EUROPEAN FREQUENCY AND TIME FORUM AND IEEE INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM (IEEE EFTF-IFCS 2021), 2021,
  • [40] Mobile measurements for distribution and attribution of particulate matter in urban environments
    Harr, Lorenz
    Sinsel, Tim
    Simon, Helge
    Torbenson, Max Carl Arne
    Jan, Esper
    ATMOSPHERIC ENVIRONMENT, 2023, 315