Wear analysis of the Bryan Cervical Disc prosthesis

被引:51
|
作者
Anderson, PA
Rouleau, JP
Bryan, VE
Carlson, CS
机构
[1] Univ Wisconsin, Dept Orthoped Surg & Rehabil, Univ Wisconsin Hosp, Madison, WI 53792 USA
[2] Medtron Sofamor Danek, Memphis, TN USA
[3] Univ Minnesota, Coll Vet Med, St Paul, MN 55108 USA
关键词
arthroplasty; cervical spine; Bryan cervical disc prosthesis; prosthetic wear; biologic response; treatment of cervical spine;
D O I
10.1097/01.BRS.0000092212.42388.79
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. In vitro wear testing of the Bryan(R) Cervical Disc prosthesis was performed in a cervical spine 1 simulator. The biologic response was assessed in chimpanzee and goat animal models. Objective. Determine the wear characteristics of the Bryan disc. Summary of Background Data.,Large joint arthroplasties fail most-commonly by wear and consequent formation of particulate material, which-induces an inflammatory response. Therefore measuring the wear characteristics of the new spinal disc replacements is important. Methods. Six prosthetic assembles were tested to,10 or 40 million cycles by load and-motion and 3 additional assemblies were tested by load only in a cervical spine simulator. Any debris was examined using ASTM standards. The local biologic response to the prosthesis was examined in two chimpanzees. Nine goats were used to assess the biologic response in both local and distant tissues. Arthrodesis was performed on three additional, control-goats that received an allograft and an anterior cervical plate. Results. Wear results: cervical spine simulators that applied the loads and motions associated with activities of daily living, produced wear particulate at a rate of 1.2 mg per million cycles. Device height decreased 0.02 mm per million cycles with approximately 77% of this decrease due to gradual creep of the nucleus under the constant compressive load. Particles generated were granular in shape with a mean feret diameter of 3.9 mum. All animals tolerated placement of the Bryan disc. Wear debris was present, in the periprosthetic and epidural spaces in some animals. However, no significant inflammatory response was observed. No wear material was found distant from the implant in draining lymph tissue, the liver, or the spleen. Conclusions. The Bryan disc has satisfactory wear characteristics and does not produce a significant inflammatory response.
引用
收藏
页码:S186 / S194
页数:9
相关论文
共 50 条
  • [21] Long-term kinematic analysis of cervical spine after single-level implantation of Bryan cervical disc prosthesis: a review
    Resnick, Daniel K.
    SPINE JOURNAL, 2013, 13 (06): : 635 - 636
  • [22] 10-year follow-up after implantation of the Bryan Cervical Disc Prosthesis
    Dejaegher, Joost
    Walraevens, Joris
    van Loon, Johannes
    Van Calenbergh, Frank
    Demaerel, Philippe
    Goffin, Jan
    EUROPEAN SPINE JOURNAL, 2017, 26 (04) : 1191 - 1198
  • [23] 10-year follow-up after implantation of the Bryan Cervical Disc Prosthesis
    Joost Dejaegher
    Joris Walraevens
    Johannes van Loon
    Frank Van Calenbergh
    Philippe Demaerel
    Jan Goffin
    European Spine Journal, 2017, 26 : 1191 - 1198
  • [24] Cervical Arthroplasty Using the Bryan Cervical Disc System
    Weber, Friedrich
    Detzner, Michael
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2010, 22 (5-6): : 468 - 479
  • [25] Follow-up study on the motion range after treatment of degenerative disc disease with the Bryan cervical disc prosthesis
    Yang Shuhua
    Hu Yong
    Zhao Jijun
    He Xianfeng
    Liu Yong
    Xu Weihua
    Du Jingyuan
    Fu Dehao
    JOURNAL OF HUAZHONG UNIVERSITY OF SCIENCE AND TECHNOLOGY-MEDICAL SCIENCES, 2007, 27 (02) : 176 - 178
  • [26] Follow-up Study on the Motion Range after Treatment of Degenerative Disc Disease with the Bryan Cervical Disc Prosthesis
    杨述华
    胡勇
    赵继军
    何贤峰
    刘勇
    许伟华
    杜靖远
    付德皓
    Journal of Huazhong University of Science and Technology(Medical Sciences), 2007, (02) : 176 - 178
  • [27] Follow-up study on the motion range after treatment of degenerative disc disease with the Bryan cervical disc prosthesis
    Shuhua Yang
    Yong Hu
    Jijun Zhao
    Xianfeng He
    Yong Liu
    Weihua Xu
    Jingyuan Du
    Dehao Fu
    Journal of Huazhong University of Science and Technology, 2007, 27 : 176 - 178
  • [28] Intervertebral disc replacement for cervical degenerative disease -: clinical results and functional outcome at two years in patients implanted with the Bryan® cervical disc prosthesis
    Heidecke, V.
    Burkert, W.
    Brucke, M.
    Rainov, N. G.
    ACTA NEUROCHIRURGICA, 2008, 150 (05) : 453 - 459
  • [29] Intervertebral disc replacement for cervical degenerative disease – clinical results and functional outcome at two years in patients implanted with the Bryan® cervical disc prosthesis
    V. Heidecke
    W. Burkert
    M. Brucke
    N. G. Rainov
    Acta Neurochirurgica, 2008, 150
  • [30] Bryan total disc arthroplasty: a replacement disc for cervical disc disease
    Wenger, Markus
    Markwalder, Thomas-Marc
    MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2010, 3 : 11 - 24