Wear analysis of the Bryan Cervical Disc prosthesis

被引:51
|
作者
Anderson, PA
Rouleau, JP
Bryan, VE
Carlson, CS
机构
[1] Univ Wisconsin, Dept Orthoped Surg & Rehabil, Univ Wisconsin Hosp, Madison, WI 53792 USA
[2] Medtron Sofamor Danek, Memphis, TN USA
[3] Univ Minnesota, Coll Vet Med, St Paul, MN 55108 USA
关键词
arthroplasty; cervical spine; Bryan cervical disc prosthesis; prosthetic wear; biologic response; treatment of cervical spine;
D O I
10.1097/01.BRS.0000092212.42388.79
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. In vitro wear testing of the Bryan(R) Cervical Disc prosthesis was performed in a cervical spine 1 simulator. The biologic response was assessed in chimpanzee and goat animal models. Objective. Determine the wear characteristics of the Bryan disc. Summary of Background Data.,Large joint arthroplasties fail most-commonly by wear and consequent formation of particulate material, which-induces an inflammatory response. Therefore measuring the wear characteristics of the new spinal disc replacements is important. Methods. Six prosthetic assembles were tested to,10 or 40 million cycles by load and-motion and 3 additional assemblies were tested by load only in a cervical spine simulator. Any debris was examined using ASTM standards. The local biologic response to the prosthesis was examined in two chimpanzees. Nine goats were used to assess the biologic response in both local and distant tissues. Arthrodesis was performed on three additional, control-goats that received an allograft and an anterior cervical plate. Results. Wear results: cervical spine simulators that applied the loads and motions associated with activities of daily living, produced wear particulate at a rate of 1.2 mg per million cycles. Device height decreased 0.02 mm per million cycles with approximately 77% of this decrease due to gradual creep of the nucleus under the constant compressive load. Particles generated were granular in shape with a mean feret diameter of 3.9 mum. All animals tolerated placement of the Bryan disc. Wear debris was present, in the periprosthetic and epidural spaces in some animals. However, no significant inflammatory response was observed. No wear material was found distant from the implant in draining lymph tissue, the liver, or the spleen. Conclusions. The Bryan disc has satisfactory wear characteristics and does not produce a significant inflammatory response.
引用
收藏
页码:S186 / S194
页数:9
相关论文
共 50 条
  • [1] A radiostereometric analysis of the Bryan® cervical disc prosthesis
    Lind, Bengt
    Zoega, Bjorn
    Anderson, Paul A.
    SPINE, 2007, 32 (08) : 885 - 890
  • [2] A radiostereometric analysis of the Bryan® cervical disc prosthesis -: Point of view
    Monsey, Robert D.
    SPINE, 2007, 32 (08) : 891 - 891
  • [3] Preliminary clinical experience with the Bryan cervical disc prosthesis
    Botelho, RV
    NEUROSURGERY, 2003, 53 (03) : 785 - 785
  • [4] Preliminary clinical experience with the Bryan Cervical Disc Prosthesis
    Goffin, J
    Casey, A
    Kehr, P
    Liebig, K
    Lind, B
    Logroscino, C
    Pointillart, V
    Van Calenbergh, F
    van Loon, J
    NEUROSURGERY, 2002, 51 (03) : 840 - 845
  • [5] The Bryan cervical disc prosthesis as an alternative to arthrodesis in the treatment of cervical spondylosis
    Lafuente, J
    Casey, ATH
    Petzold, A
    Brew, S
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (04): : 508 - 512
  • [6] Traumatic loading of the Bryan cervical disc prosthesis: An in vitro study
    Duggal, Neil
    Rabin, Doron
    Chamberlain, Robert H.
    Baek, Seungwon
    Crawford, Neil R.
    NEUROSURGERY, 2007, 60 (04) : 388 - 392
  • [7] Segmental malalignment with the Bryan Cervical Disc prosthesis - Contributing factors
    Sears, William R.
    Duggal, Neil
    Sekhon, Lali H.
    Williamson, Owen D.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2007, 20 (02): : 111 - 117
  • [8] Preliminary clinical experience with the Bryan Cervical Disc Prosthesis - Comments
    Cooper, PR
    Benzel, EC
    Haid, RW
    Sonntag, VKH
    McCormick, PC
    Traynelis, VC
    NEUROSURGERY, 2002, 51 (03) : 845 - 847
  • [9] Segmental malalignment with the Bryan Cervical Disc prosthesis - Does it occur?
    Sears, William R.
    Sekhon, Lali H.
    Duggal, Neil
    Williamson, Owen D.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2007, 20 (01): : 1 - 6
  • [10] Early clinical and radiologic outcomes of cervical arthroplasty with Bryan Cervical Disc prosthesis
    Shim, Chan Shik
    Lee, Sang-Ho
    Park, Hyun-Ju
    Kang, Han-Sug
    Hwang, Ji-Hee
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (07): : 465 - 470