Flow rate-independent electrical aerosol sensor

被引:2
|
作者
Rostedt, A. [1 ]
Keskinen, J. [1 ]
机构
[1] Tampere Univ Technol, Phys Lab, Aerosol Phys, Tampere, Finland
关键词
LOW-PRESSURE IMPACTOR; SURFACE-AREA; DIFFUSION; PARTICLES; LUNG; PERFORMANCE; ELPI;
D O I
10.1080/02786826.2018.1498586
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A new diffusion charging-based aerosol instrument design is presented, intended to be a starting point for a low-cost particle concentration sensor. The aim in the design is to minimize instrument response dependence on the sample flow rate. The operation principle, response functions of the components, and performance of a prototype instrument are reported. Based on the performance evaluation, the instrument response remained constant within +/- 15% over a wide sample flow rate range of 3 to 10 lpm. While in the design some sensitivity is sacrificed to minimize flow rate dependency, the detection limit is still sufficient for a practical sensor application. (c) 2018 American Association for Aerosol Research