Matrix effects in plutonium isotope ratio measurements using thermal ionization mass spectrometry

被引:1
|
作者
Mathew, Kattathu Joseph [1 ]
Ottenfeld, Chelsea [1 ]
Levesque, Steven [1 ]
Keller, Russ [1 ]
机构
[1] Los Alamos Natl Lab, Actinide Analyt Chem, MS G740, Los Alamos, NM 87545 USA
关键词
Isotope ratio measurements; Thermal ionization mass spectrometry; Certified reference materials; Matrix effects; Nuclear forensics; Nuclear fuel cycle; NUCLEAR MATERIAL; URANIUM; SAMPLES; TIMS; UNCERTAINTY; ASSAY;
D O I
10.1007/s10967-021-08170-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Through high fidelity measurements of four different plutonium certified reference material standards from New Brunswick Laboratory, the mass fractionation correction factor for the Pu-240/Pu-239 major isotope ratio measurements by the total evaporation methodology using thermal ionization mass spectrometer instruments is shown to depend on the elapsed time since separation of the decay products from the analyzed plutonium fraction. Systematic biases arising from this matrix dependence of the mass fractionation correction factor are larger than the precision obtained in routine TE measurements of the Pu-240/Pu-239 major isotope ratio using thermal ionization mass spectrometry regarded as the gold standard for these measurements. LA-UR-21-21,941.
引用
收藏
页码:1005 / 1017
页数:13
相关论文
共 50 条
  • [21] TUNGSTEN ISOTOPE RATIO DETERMINATIONS BY NEGATIVE THERMAL IONIZATION MASS-SPECTROMETRY
    VOLKENING, J
    KOPPE, M
    HEUMANN, KG
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1991, 107 (02): : 361 - 368
  • [23] High-precision stable zirconium isotope ratio measurements by double spike thermal ionization mass spectrometry
    Feng, Lanping
    Hu, Wenfeng
    Jiao, Yu
    Zhou, Lian
    Zhang, Wen
    Hu, Zhaochu
    Liu, Yongsheng
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (04) : 736 - 745
  • [24] Isotope ratio measurements of trace Nd by the total evaporation normalization (TEN) method in thermal ionization mass spectrometry
    Wakaki, Shigeyuki
    Shibata, Shin-Nosuke
    Tanaka, Tsuyoshi
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2007, 264 (2-3) : 157 - 163
  • [25] IRIDIUM ISOTOPE RATIO MEASUREMENTS BY NEGATIVE THERMAL IONIZATION MASS-SPECTROMETRY AND ATOMIC-WEIGHT OF IRIDIUM
    WALCZYK, T
    HEUMANN, KG
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1993, 123 (02): : 139 - 147
  • [26] ISOTOPE FRACTIONATION OF PLUTONIUM IN THERMAL IONIZATION MASS-SPECTROMETRY - DIRECT DEPOSITION AND RESIN BEAD
    POUPARD, D
    JUERY, A
    [J]. RADIOCHIMICA ACTA, 1992, 57 (01) : 21 - 24
  • [27] High precision isotope ratio measurements on boron by thermal ionization mass spectrometry using Rb2BO2+ ion
    Rao, Radhika M.
    Parab, Ankush R.
    Bhushan, K. Sasi
    Aggarwal, Suresh Kumar
    [J]. ANALYTICAL METHODS, 2011, 3 (02) : 322 - 327
  • [28] Precise Re isotope ratio measurements by negative thermal ionization mass spectrometry (NTI-MS) using total evaporation technique
    Suzuki, K
    Miyata, Y
    Kanazawa, N
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2004, 235 (01) : 97 - 101
  • [29] HIGH-FIDELITY IN ISOTOPE RATIO MEASUREMENTS BY RESONANCE IONIZATION MASS-SPECTROMETRY
    MILLER, CM
    FEAREY, BL
    PALMER, BA
    NOGAR, NS
    [J]. RESONANCE IONIZATION SPECTROSCOPY 1988, 1989, 94 : 297 - 300
  • [30] Plutonium determination in bioassay samples using radiochemical thermal ionization mass spectrometry
    LaMont, SP
    Shick, CR
    Cable-Dunlap, P
    Fauth, DJ
    LaBone, TR
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2005, 263 (02) : 477 - 481