The ArcCHECK diode array for dosimetric verification of HybridArc

被引:26
|
作者
Petoukhova, A. L. [1 ]
van Egmond, J. [1 ]
Eenink, M. G. C. [1 ]
Wiggenraad, R. G. J. [1 ]
van Santvoort, J. P. C. [1 ]
机构
[1] Radiotherapy Ctr W, NL-2501 CK The Hague, Netherlands
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 16期
关键词
MODULATED ARC THERAPY; QUALITY-ASSURANCE; DOSE CALCULATION; CLINICAL-EVALUATION; TREATMENT DELIVERY; IMRT; PHOTON; PHANTOM; TOMOTHERAPY; ALGORITHM;
D O I
10.1088/0031-9155/56/16/021
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this work is to evaluate dosimetric accuracy of a new treatment modality, HybridArc, in iPlan RT Dose 4.5 (BrainLAB, Feldkirchen, Germany) using a four-dimensional diode array (ArcCHECK, Sun Nuclear Corporation, Melbourne, USA). HybridArc is able to enhance dynamic conformal arcs with inversely planned elements. HybridArc plans for various sites (intracranial and extracranial) were constructed and after that these plans were recalculated for the ArcCHECK diode array with Monte Carlo (MC) and Pencil Beam (PB) dose algorithms in iPlan RT Dose. All measurements of these HybridArc plans were performed with 6 MV photon beams of a Novalis accelerator (BrainLAB, Feldkirchen, Germany) using the ArcCHECK device without and with an insert containing an ionization chamber. Comparison of the absolute dose distributions measured and calculated in iPlan RT Dose with the MC algorithm at the cylinder of the ArcCHECK diode array for HybridArc plans gives good agreement, even for the 2% dose difference and 2 mm distance-to-agreement criteria. The PB calculations significantly differ from the ArcCHECK measurements so that the MC algorithm is found to be superior to the PB algorithm in the calculation of the HybridArc plans. One of the drawbacks of the PB calculations in iPlan RT Dose is the too large arc step size of 10 degrees. Use of a finer angular resolution may improve the PB results significantly.
引用
下载
收藏
页码:5411 / 5428
页数:18
相关论文
共 50 条
  • [21] Picking a high precision measuring diode array with two-dimensional dosimetric verification of intenstity modulation and atom-triggered radiation fields
    Block, A.
    Bauer, R.
    Mewes, A.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2009, 185 : 168 - 168
  • [22] Evaluation of the ArcCHECK QA system for IMRT and VMAT verification
    Li, Guangjun
    Zhang, Yingjie
    Jiang, Xiaoqin
    Bai, Sen
    Peng, Guang
    Wu, Kui
    Jiang, Qingfeng
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (03): : 295 - 303
  • [23] Dosimetric Uncertainties Associated with a Two-Dimensional Multiple Diode Array Detector Used in Dose Verification Quality Assurance Due to CT Image Artifacts
    Almatouq, F.
    Ahmad, S.
    Ali, I.
    MEDICAL PHYSICS, 2018, 45 (06) : E305 - E305
  • [24] Dosimetric verification for intensity-modulated arc therapy plans by use of 2D diode array, radiochromic film and radiosensitive polymer gel
    Hayashi, Naoki
    Malmin, Ryan L.
    Watanabe, Yoichi
    JOURNAL OF RADIATION RESEARCH, 2014, 55 (03) : 541 - 552
  • [25] A phantom for brachytherapy treatment planning systems verification with the ArcCHECK® device
    Chelminski, K.
    Sobotka, P.
    Buczek, B.
    Gruszczynska, E.
    Bulski, W.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S707 - S707
  • [26] Experimental evaluation of high density ArcCHECK mode for SBRT verification
    Montes, E.
    Modolell, I.
    Sancho, I.
    Puxeu, J.
    Reynes, G.
    Picon, C.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S746 - S747
  • [27] Dosimetric verification of enhanced dynamic wedges by a 2D ion chamber array
    Se An Oh
    Sung Kyu Kim
    Min Kyu Kang
    Ji Woon Yea
    Eng Chan Kim
    Journal of the Korean Physical Society, 2013, 63 : 2215 - 2219
  • [28] Dosimetric verification of enhanced dynamic wedges by a 2D ion chamber array
    Oh, Se An
    Kim, Sung Kyu
    Kang, Min Kyu
    Yea, Ji Woon
    Kim, Eng Chan
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 63 (11) : 2215 - 2219
  • [29] DOSIMETRIC VERIFICATION OF IMRT TREATMENTS USING A 2D ION CHAMBER ARRAY
    Marcantonini, M.
    Raymondi, C.
    Zucchetti, C.
    Leogrande, M. P.
    Aristei, C.
    Gobbi, G.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S511 - S511
  • [30] ARCCHECK IMRT-QA: EXPLORING ITS SENSITIVITY TO INTRODUCED DOSIMETRIC ERRORS
    Rietveld, P.
    de Goede, M.
    Wiggenraad, R.
    Eenink, M.
    van Egmond, J.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S508 - S508