Difference of optimal dose of contrast agent between gray-scale and power Doppler imaging in assessing graded coronary stenosis by myocardial contrast echocardiography

被引:1
|
作者
Masugata, H
DeMaria, AN
Peters, B
Lafitte, S
Strachan, GM
Kwan, OL
Ohmori, K
Mizushige, K
Kohno, M
机构
[1] Kagawa Med Univ, Dept Internal Med 2, Kagawa 7610793, Japan
[2] Univ Calif San Diego, Div Cardiovasc, San Diego, CA 92103 USA
关键词
myocardial contrast echocardiography; myocardial blood flow; perfusion;
D O I
10.1097/01.RLI.0000073641.93901.b9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: In myocardial contrast echocardiography (MCE), power Doppler imaging is more sensitive to contrast agent (microbubble) than gray-scale B-mode imaging; however, no data exist regarding the optimal contrast dose in power Doppler imaging. This study examined the optimal dose of contrast agent for power Doppler in assessing coronary stenosis. Methods: Three grades of coronary stenosis were produced in 6 open-chest dogs. NICE was performed with gray-scale and power Doppler during continuous infusion of 0.2 mL/min FS-069. Thereafter, NICE was repeated with power Doppler during continuous infusion of 0.1 mL/min FS-069. Results: Although the videointensity in the stenosed bed with power Doppler (214+/-14) was greater than gray scale (35+/-17) during 0.2 mL/min FS-069 infusion (P<0.0001), power Doppler failed to identify milder coronary stenoses because videointensity in stenosed bed was quickly saturated with contrast agent. The videointensity in the stenosed bed with power Doppler (127+/-49) during 0.1 mL/min FS-069 infusion was greater than gray scale (35+/-17) during 0.2 mL/min FS-069 infusion (P<0.0001), and all levels of stenosis were identified with power Doppler, even though the dose of contrast agent was half of that of gray scale imaging. The correlation between videointensity and myocardial blood flow was better in the case of power Doppler at 0.1 mL/min FS-069 infusion (r=0.77, P<0.0001) than in the case of gray scale imaging at 0.2 mL/min FS-069 infusion (r=0.66, P<0.01). Conclusions: These data support the need for a lower dose of contrast agent for power Doppler than for gray scale to detect milder coronary stenosis and avoid saturation of imaging fields.
引用
收藏
页码:550 / 558
页数:9
相关论文
共 50 条
  • [1] Assessment of coronary stenosis severity and transmural perfusion gradient by myocardial contrast echocardiography - Comparison of gray-scale B-mode with power Doppler imaging
    Masugata, H
    Cotter, B
    Peters, B
    Ohmori, K
    Mizushige, K
    DeMaria, AN
    CIRCULATION, 2000, 102 (12) : 1427 - 1433
  • [2] Myocardial contrast echocardiography with power Doppler imaging
    Spencer, KT
    Grayburn, PA
    Mor-Avi, V
    Bednarz, J
    Grimm, RA
    Furlong, K
    Farnum, RF
    Floer, SD
    Widner, PJ
    Lang, RM
    AMERICAN JOURNAL OF CARDIOLOGY, 2000, 86 (04): : 479 - +
  • [3] Radiofrequency dual-spectra compared to gray-scale myocardial contrast echocardiography
    Asanuma, T
    Belohlavek, M
    Khandheria, BK
    Seward, JB
    CIRCULATION, 2001, 104 (17) : 732 - 732
  • [4] Potential pitfalls of visualization of myocardial perfusion by myocardial contrast echocardiography with harmonic gray scale B-mode and power Doppler imaging
    Hisashi Masugata
    Kazushi Yukiiri
    Yuichiro Takagi
    Koji Ohmori
    Katsufumi Mizushige
    Masakazu Kohno
    The International Journal of Cardiovascular Imaging, 2004, 20 : 117 - 125
  • [5] Potential pitfalls of visualization of myocardial perfusion by myocardial contrast echocardiography with harmonic gray scale B-mode and power Doppler imaging
    Masugata, H
    Yukiiri, K
    Takagi, Y
    Ohmori, K
    Mizushige, K
    Kohno, M
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2004, 20 (02): : 117 - 125
  • [6] Comparative analysis of renal flow using contrast power Doppler and gray-scale ultrasound
    Sehgal, CM
    Arger, PH
    Bovee, KC
    Pugh, CR
    Kirchofer, JI
    PHYSIOLOGY AND FUNCTION FROM MULTIDIMENSIONAL IMAGES - MEDICAL IMAGING 1997, 1997, 3033 : 209 - 214
  • [7] Low power myocardial contrast echocardiography accurately discriminates between grades of coronary artery stenosis
    Jeetley, P
    Hickman, M
    Senior, R
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (05) : 316A - 316A
  • [8] Ultrasound angiography with intravenous contrast agent using wideband gray-scale harmonic imaging for evaluation of tips stenosis.
    Kono, Y
    Mattrey, RF
    Pinnell, SP
    McKlendin, K
    Rose, S
    Hassanein, T
    GASTROENTEROLOGY, 2000, 118 (04) : A964 - A964
  • [9] Quantitaive assessment of myocardial perfusion by myocardial contrast echocardiography with harmonic power Doppler imaging
    Tanabe, K
    Tani, T
    Katayama, M
    Yagi, T
    Tamita, K
    Yamabe, K
    Yamamuro, A
    Takagi, T
    Nagai, J
    Morioka, S
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) : 427A - 427A
  • [10] Contrast-enhanced ultrasonography compared to gray-scale and power doppler in the diagnosis of peripheral lymphadenopathy
    Slaisova, Radka
    Benda, Karel
    Jarkovsky, Jiri
    Petrasova, Hana
    Szturz, Petr
    Valek, Vlastimil
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (04) : 693 - 698