Technical errors in the application of Graf's hip sonography method.

被引:7
|
作者
Müller, W
Lercher, K
DeVaney, TTJ
Giner, F
Graf, R
机构
[1] Inst Med Phys & Biophys, A-8010 Graz, Austria
[2] Allgemeines & Orthopad Landeskrankenhaus, Stolzalpe, Austria
来源
ULTRASCHALL IN DER MEDIZIN | 2001年 / 22卷 / 01期
关键词
infant hip sonography; hip screening; developmental dysplasia of the hip (DDH); CDH; artefacts;
D O I
10.1055/s-2001-11250
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Aim: The aim of the study is to point out possible examination errors because of a titled transducer, and to analyse their sources. Methods: The results obtained by experienced examiners using the correct examination technique were compared to results obtained under defined deviations from the standard. With the use of a computer model the course taken by the ultrasound waves in the neonatal hip was simulated and the sonographic image reconstructed. Results: Twisted as well as tilted positions of the transducer lead to significant errors in ct, the acetabular roof angle. This is in accordance with the computer model predictions. Caudo-cranially tilted probes lead to false-positive, therapeutically relevant incorrect diagnoses. Conclusion: The computer simulation predicts a high degree of accuracy of hip sonography in Graf's technique under standard conditions. The effectiveness of sonographic infant hip screening largely depends on the ability to adhere to the standardized approach, minimizing possible errors. A mechanical transducer guide prevents tilting errors by reducing the degrees of variability in positioning the transducer.
引用
收藏
页码:48 / 54
页数:7
相关论文
共 50 条
  • [41] The genesis of Lachmann's method.
    Caruso, Carlo
    MODERN LANGUAGE REVIEW, 2007, 102 : 469 - 470
  • [42] Meaning in Spinoza's method.
    Bagley, P
    INTERNATIONAL PHILOSOPHICAL QUARTERLY, 2005, 45 (01) : 133 - 136
  • [43] Meaning in Spinoza's method.
    DeBrabander, F
    REVIEW OF METAPHYSICS, 2004, 58 (02): : 434 - 436
  • [44] The genesis of Lachmann's method.
    Courtney, E.
    CLASSICAL JOURNAL, 2007, 102 (03): : 312 - 315
  • [45] Steinach's regeneration method.
    Roux, W
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1920, 46 : 891 - 891
  • [46] PROBLEMS ENCOUNTERED IN SONOGRAPHY OF THE HIP (STANDARDIZED RECORDING TECHNIQUE, ERRORS OF MEASUREMENT, CONSEQUENCES FOR THERAPY)
    GRAF, R
    HEUBERER, I
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1985, 123 (02): : 127 - 135
  • [47] Reproducibility of acquiring ultrasonographic infant hip images by the Graf method after an infant hip ultrasound training course
    Yasuhiro Shirai
    Kenjiro Wakabayashi
    Ikuo Wada
    Hideyuki Goto
    Yoshino Ueki
    Atsushi Tsuchiya
    Yoshiaki Tsuboi
    Myongsu Ha
    Takanobu Otsuka
    Journal of Medical Ultrasonics, 2018, 45 : 583 - 589
  • [48] Reproducibility of acquiring ultrasonographic infant hip images by the Graf method after an infant hip ultrasound training course
    Shirai, Yasuhiro
    Wakabayashi, Kenjiro
    Wada, Ikuo
    Goto, Hideyuki
    Ueki, Yoshino
    Tsuchiya, Atsushi
    Tsuboi, Yoshiaki
    Ha, Myongsu
    Otsuka, Takanobu
    JOURNAL OF MEDICAL ULTRASONICS, 2018, 45 (04) : 583 - 589
  • [49] SONOGRAPHY - A NEW METHOD FOR OBJECTIVE MEASUREMENT OF HIP INSTABILITY IN INFANTS
    OBERG, L
    DAHLSTROM, H
    FRIBERG, S
    ACTA ORTHOPAEDICA SCANDINAVICA, 1985, 56 (04): : 353 - 353
  • [50] Experiences in diagnosis and treatment of hip dislocation and dysplasia in populations screened by the ultrasound method of Graf
    Kobler, G
    Hell, AK
    SWISS MEDICAL WEEKLY, 2003, 133 (35-36) : 484 - 487