Flow-Diverter Stent for the Endovascular Treatment of Intracranial Aneurysms A Prospective Study in 29 Patients With 34 Aneurysms

被引:253
|
作者
Lubicz, Boris [1 ]
Collignon, Laurent [3 ]
Raphaeli, Gai [1 ]
Pruvo, Jean-Pierre [4 ]
Bruneau, Michael [2 ]
De Witte, Olivier [2 ]
Leclerc, Xavier [4 ]
机构
[1] Erasme Univ Hosp, Dept Neuroradiol, B-1070 Brussels, Belgium
[2] Erasme Univ Hosp, Dept Neurosurg, B-1070 Brussels, Belgium
[3] CHR Citadelle, Dept Radiol, Liege, Belgium
[4] Roger Salengro Univ Hosp, Dept Neuroradiol, Lille, France
关键词
detachable coils; intracranial aneurysms; stents; RECONSTRUCTION;
D O I
10.1161/STROKEAHA.110.589911
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background and Purpose-The purpose of this study is to report our preliminary experience with the flow-diverter Silk stent for the endovascular treatment of intracranial aneurysms. Methods-This prospective study was approved by the authors' ethical committees. Twenty-nine patients with 34 fusiform or wide-necked unruptured aneurysms were included and treated by Silk stent placement alone by 2 physicians in 3 different centers. Technical issues, immediate findings, delayed complications, clinical follow-up, and imaging follow-up at 3 and 6 months were assessed. Results-Endovascular treatment was successfully performed in 26 patients (90%). In 3 patients, the stent could not be delivered. Mortality and morbidity rates were of 4% (1 of 26) and 15% (4 of 26), respectively; 1 patient died from a delayed aneurysm rupture related to stent migration, 3 experienced a thromboembolic event, and 1 patient developed progressive visual disturbances related to an increased mass effect. Clinical outcome in 25 patients was unchanged (n=19), improved (n=2), or worsened (=4). Angiographic follow-up in 24 patients (29 aneurysms) showed 20 complete occlusions (69%), 1 neck remnant (3.5%), and 8 incomplete occlusions (27.5%). Significant parent artery stenosis at 6 months occurred in 8 cases (33%). Conclusions-Despite the potential interest of the Silk flow-diverter stent to treat complex intracranial aneurysms without coils, the delayed complication rate is quite high and leads to use this technique only in selective cases. (Stroke. 2010;41:2247-2253.)
引用
收藏
页码:2247 / 2253
页数:7
相关论文
共 50 条
  • [31] Usefulness of Silent MR Angiography for Intracranial Aneurysms Treated with a Flow-Diverter Device
    Oishi, H.
    Fujii, T.
    Suzuki, M.
    Takano, N.
    Teranishi, K.
    Yatomi, K.
    Kitamura, T.
    Yamamoto, M.
    Aoki, S.
    Arai, H.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (05) : 808 - 814
  • [32] Flow-Diverter Devices for Intracranial Aneurysms: Systematic Review and Meta-analysis
    Arrese, Ignacio
    Sarabia, Rosario
    Pintado, Rebeca
    Delgado-Rodriguez, Miguel
    NEUROSURGERY, 2013, 73 (02) : 193 - 199
  • [33] Flow diverter stents in the treatment of recanalized intracranial aneurysms
    Akgul, Erol
    Onan, Hasan Bilen
    Islek, Irem
    Tonge, Mehmet
    Durmus, Yavuz
    Barburoglu, Mehmet
    Azizova, Aynur
    Erol, Cengiz
    Hakyemez, Bahattin
    Sencer, Serra
    Aydin, Kubilay
    Arat, Anil
    INTERVENTIONAL NEURORADIOLOGY, 2021, 27 (04) : 481 - 489
  • [34] Treatment of intracranial aneurysms using the Tubridge flow diverter
    Yuan, Dun
    Nibu, Zhenmei
    Guo, Yuxin
    Cao, Fang
    Liu, Junyu
    Jiang, Weixi
    Li, Yifeng
    Yan, Junxia
    POSTGRADUATE MEDICAL JOURNAL, 2024,
  • [35] Flow diverter for endovascular treatment of intracranial mirror segment internal carotid artery aneurysms
    Kuhn, Anna Luisa
    Kan, Peter
    Srinivasan, Visish
    Rex, David E.
    Rodrigues, Katyucia de Macedo
    Howk, Mary C.
    Wakhloo, Ajay K.
    Puri, Ajit S.
    INTERVENTIONAL NEURORADIOLOGY, 2019, 25 (01) : 4 - 11
  • [36] Endovascular treatment of wide-necked paraophthalmic aneurysms with flow diverter stent implantation
    Guzeldag, Hamit
    Onan, Bilen
    Piskin, Ferhat Can
    Sozutok, Sinan
    CUKUROVA MEDICAL JOURNAL, 2021, 46 (04): : 1666 - 1674
  • [37] Endovascular Exclusion of the Anterior Communicating Artery with Flow-Diverter Stents as an Emergency Treatment for Blister-Like Intracranial Aneurysms A Case Report
    Rouchaud, Aymeric
    Saleme, Suzana
    Gory, Benjamin
    Ayoub, David
    Mounayer, Charbel
    INTERVENTIONAL NEURORADIOLOGY, 2013, 19 (04) : 471 - 478
  • [38] Flow-Diverter Devices for Intracranial Aneurysms: Systematic Review and Meta-analysis COMMENTS
    Briganti, Francesco
    Raymond, Jean
    Darsaut, Tim E.
    Byrne, James
    NEUROSURGERY, 2013, 73 (02) : 199 - 200
  • [39] Cerebral Revascularization for Aneurysms in the Flow-Diverter Era COMMENT
    Welch, Babu G.
    NEUROSURGERY, 2017, 80 (05) : 768 - 768
  • [40] Safety and efficacy of flow-diverter stents in endovascular treatment of intracranial aneurysm: Interest of the prospective DIVERSION observational study
    Gory, B.
    Bonafe, A.
    Pierot, L.
    Spelle, L.
    Berge, J.
    Piotin, M.
    Mounayer, C.
    Biondi, A.
    Courtheoux, P.
    Cognard, C.
    Desal, H.
    Herbreteaux, D.
    Gabrillargues, J.
    Ricolfi, F.
    Sourour, N.
    Sedat, J.
    Gallas, S.
    Boubagra, K.
    Huot, L.
    Embarek, S.
    Kulcsar, Z.
    Taschner, C.
    Chapuis, F.
    Turjman, F.
    JOURNAL OF NEURORADIOLOGY, 2014, 41 (02) : 93 - 96