Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy

被引:8
|
作者
Mackeprang, P-H [1 ,2 ,3 ]
Vuong, D. [1 ,2 ,3 ]
Volken, W. [1 ,2 ,3 ]
Henzen, D. [1 ,2 ,3 ]
Schmidhalter, D. [1 ,2 ,3 ]
Malthaner, M. [1 ,2 ,3 ]
Mueller, S. [1 ,2 ,3 ]
Frei, D. [1 ,2 ,3 ]
Stampanoni, M. F. M. [4 ,5 ,6 ]
Dal Pra, A. [1 ,2 ,3 ]
Aebersold, D. M. [1 ,2 ,3 ]
Fix, M. K. [1 ,2 ,3 ]
Manser, P. [1 ,2 ,3 ]
机构
[1] Bern Univ Hosp, Inselspital, Div Med Radiat Phys, Bern, Switzerland
[2] Bern Univ Hosp, Inselspital, Dept Radiat Oncol, Bern, Switzerland
[3] Univ Bern, Bern, Switzerland
[4] Univ Zurich, Inst Biomed Engn, Zurich, Switzerland
[5] Swiss Fed Inst Technol, Zurich, Switzerland
[6] Paul Scherrer Inst, Swiss Light Source, Villigen, Switzerland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 01期
关键词
Monte Carlo; QA; CyberKnife; STEREOTACTIC RADIOSURGERY SYSTEM; DELIVERY QUALITY-ASSURANCE; MV PHOTON BEAMS; CLINICAL IMPLEMENTATION; FILM; SIMULATIONS; TOMOTHERAPY; TISSUE; VMAT;
D O I
10.1088/1361-6560/aa97f8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work aims to develop, implement and validate a Monte Carlo (MC)-based independent dose calculation (IDC) framework to perform patient-specific quality assurance (QA) for multi-leaf collimator (MLC)-based CyberKnife (R) (Accuray Inc., Sunnyvale, CA) treatment plans. The IDC framework uses an XML-format treatment plan as exported from the treatment planning system (TPS) and DICOM format patient CT data, an MC beam model using phase spaces, CyberKnife MLC beam modifier transport using the EGS++ class library, a beam sampling and coordinate transformation engine and dose scoring using DOSXYZnrc. The framework is validated against dose profiles and depth dose curves of single beams with varying field sizes in a water tank in units of cGy/Monitor Unit and against a 2D dose distribution of a full prostate treatment plan measured with Gafchromic EBT3 (Ashland Advanced Materials, Bridgewater, NJ) film in a homogeneous water-equivalent slab phantom. The film measurement is compared to IDC results by gamma analysis using 2% (global)/2 mm criteria. Further, the dose distribution of the clinical treatment plan in the patient CT is compared to TPS calculation by gamma analysis using the same criteria. Dose profiles from IDC calculation in a homogeneous water phantom agree within 2.3% of the global max dose or 1 mm distance to agreement to measurements for all except the smallest field size. Comparing the film measurement to calculated dose, 99.9% of all voxels pass gamma analysis, comparing dose calculated by the IDC framework to TPS calculated dose for the clinical prostate plan shows 99.0% passing rate. IDC calculated dose is found to be up to 5.6% lower than dose calculated by the TPS in this case near metal fiducial markers. An MC-based modular IDC framework was successfully developed, implemented and validated against measurements and is now available to perform patient-specific QA by IDC.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Benchmarking of Monte Carlo dose calculation for MLC based CyberKnife Radiotherapy
    Mackeprang, P. H.
    Vuong, D.
    Volken, W.
    Henzen, D.
    Schmidhalter, D.
    Malthaner, M.
    Mueller, S.
    Frei, D.
    Aebersold, D. M.
    Fix, M. K.
    Manser, P.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S994 - S995
  • [2] Benchmarking Monte-Carlo dose calculation for MLC CyberKnife treatments
    Mackeprang P.-H.
    Vuong D.
    Volken W.
    Henzen D.
    Schmidhalter D.
    Malthaner M.
    Mueller S.
    Frei D.
    Kilby W.
    Aebersold D.M.
    Fix M.K.
    Manser P.
    Radiation Oncology, 14 (1)
  • [3] Assessing accuracy and efficiency of Monte Carlo based dose calculation for MLC CyberKnife treatments
    Mackeprang, P. -H.
    Volken, W.
    Henzen, D.
    Schmidhalter, D.
    Malthaner, M.
    Mueller, S.
    Frei, D.
    Kilby, W.
    Aebersold, D. M.
    Fix, M. K.
    Manser, P.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 (06) : 597 - 597
  • [4] Use of VirtuaLinac to Perform Monte-Carlo Based Dose Calculation for Verification of Radiotherapy Treatments On a Jawless Ring Gantry Linac with Double Stack MLC
    Sethi, A.
    Ji, Z.
    Liu, R.
    Laugeman, E.
    Sawkey, D.
    Zhao, T.
    Mutic, S.
    Hugo, G.
    Cai, B.
    MEDICAL PHYSICS, 2020, 47 (06) : E872 - E873
  • [5] An independent dose calculation algorithm for MLC-based stereotactic radiotherapy
    Lorenz, Friedlieb
    Killoran, Joseph H.
    Wenz, Frederik
    Zygmanski, Piotr
    MEDICAL PHYSICS, 2007, 34 (05) : 1605 - 1614
  • [6] Implementation of Monte Carlo Dose Calculation for CyberKnife treatment planning
    Ma, C-M
    Li, J. S.
    Deng, J.
    Fan, J.
    INTERNATIONAL WORKSHOP ON MONTE CARLO TECHNIQUES IN RADIOTHERAPY DELIVERY AND VERIFICATION - THIRD MCGILL INTERNATIONAL WORKSHOP, 2008, 102
  • [7] MONTE-CARLO CALCULATION OF ABSORBED DOSE IN XEROMAMMOGRAPHY
    DANCE, DR
    BAKER, AM
    DAVIS, R
    STACEY, AJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (04): : 760 - 760
  • [8] A Monte-Carlo based dose engine for proton radiotherapy treatment planning
    Newhauser, W.
    Fontenot, J.
    Zheng, Y.
    Taddei, P.
    Mirkovic, D.
    Titt, U.
    Zhu, X.
    Sahoo, N.
    Schaffner, B.
    Langenegger, A.
    Koch, N.
    Zhang, X.
    Mohan, R.
    MEDICAL PHYSICS, 2007, 34 (06) : 2406 - 2406
  • [9] validation of Monte Carlo dose calculation algorithm for CyberKnife multileaf collimator
    Gondre, M.
    Marsolat, F.
    Bourhis, J.
    Bochud, F.
    Moeckli, R.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1362 - S1363
  • [10] Validation of Monte Carlo dose calculation algorithm for CyberKnife multileaf collimator
    Gondre, Maude
    Marsolat, Fanny
    Bourhis, Jean
    Bochud, Francois
    Moeckli, Raphael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (02):