Microwave-assisted wet digestion with H2O2 at high temperature and pressure using single reaction chamber for elemental determination in milk powder by ICP-OES and ICP-MS

被引:54
|
作者
Muller, Edson I. [1 ]
Souza, Juliana P. [1 ]
Muller, Cristiano C. [1 ]
Muller, Aline L. H. [1 ]
Mello, Paola A. [1 ]
Bizzi, Cezar A. [1 ]
机构
[1] Univ Fed Santa Maria, Dept Quim, BR-97105900 Santa Maria, RS, Brazil
关键词
Hydrogen peroxide; Milk powder; Single reaction chamber; Microwave-assisted wet digestion; Elemental analysis; Trace elements; ACTIVE PHARMACEUTICAL INGREDIENTS; HYDROGEN-PEROXIDE; METALS DETERMINATION; TRACE-ELEMENTS; TOXIC ELEMENTS; SAMPLES; SPECTROMETRY; OXIDATION; DECOMPOSITION; FEASIBILITY;
D O I
10.1016/j.talanta.2016.05.019
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work a green digestion method which only used H2O2 as an oxidant and high temperature and pressure in the single reaction chamber system (SRC-UltraWave (TM)) was applied for subsequent elemental determination by inductively coupled plasma-based techniques. Milk powder was chosen to demonstrate the feasibility and advantages of the proposed method. Samples masses up to 500 mg were efficiently digested, and the determination of Ca, Fe, K, Mg and Na was performed by inductively coupled plasma optical emission spectrometry (ICP-OES), while trace elements (B, Ba, Cd, Cu, Mn, Mo, Pb, Sr and Zn) were determined by inductively coupled plasma mass spectrometry (ICP-MS). Residual carbon (RC) lower than 918 mg L-1 of C was obtained for digests which contributed to minimizing interferences in determination by ICP-OES and ICP-MS. Accuracy was evaluated using certified reference materials NIST 1549 (non-fat milk powder certified reference material) and NIST 8435 (whole milk powder reference material). The results obtained by the proposed method were in agreement with the certified reference values (t-test, 95% confidence level). In addition, no significant difference was observed between results obtained by the proposed method and conventional wet digestion using concentrated HNO3. As digestion was performed without using any kind of acid, the characteristics of final digests were in agreement with green chemistry principles when compared to digests obtained using conventional wet digestion method with concentrated HNO3. Additionally, H2O2 digests were more suitable for subsequent analysis by ICP-based techniques due to of water being the main product of organic matrix oxidation. The proposed method was suitable for quality control of major components and trace elements present in milk powder in consonance with green sample preparation. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:232 / 238
页数:7
相关论文
共 47 条
  • [11] Determination of trace element concentrations in tomato samples at different stages of maturation by ICP OES and ICP-MS following microwave-assisted digestion
    Bressy, Fernanda C.
    Brito, Geysa B.
    Barbosa, Isa S.
    Teixeira, Leonardo S. G.
    Korn, Maria Gracas A.
    MICROCHEMICAL JOURNAL, 2013, 109 : 145 - 149
  • [12] Feasibility of nut digestion using single reaction chamber for further trace element determination by ICP-OES
    Muller, Cristiano C.
    Muller, Aline L. H.
    Pirola, Camillo
    Duarte, Fabio A.
    Flores, Erico M. M.
    Muller, Edson I.
    MICROCHEMICAL JOURNAL, 2014, 116 : 255 - 260
  • [13] Greening microwave-assisted digestion method using hydrogen peroxide for determination of elemental impurities by ICP-OES in antihypertensive active pharmaceutical ingredients
    Senger, Caroline M.
    Dornelles, Isadora S.
    Queiroz, Jennefer M.
    Mello, Paola A.
    Muller, Edson I.
    Muller, Aline L. H.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2024, 238
  • [14] ICP-OES ELEMENTAL IMPURITIES STUDY ON DIFFERENT PHARMACEUTICAL DOSAGE FORMS OF IBUPROFEN USING A MICROWAVE-ASSISTED DIGESTION PROCEDURE
    Jancevska, Katerina
    Petrushevski, Gjorgji
    Bogdanoska, Mirjana
    Stafilov, Trajce
    Ugarkovic, Sonja
    MACEDONIAN JOURNAL OF CHEMISTRY AND CHEMICAL ENGINEERING, 2021, 40 (01) : 43 - 50
  • [15] Screening of TiO2 and Au nanoparticles in cosmetics and determination of elemental impurities by multiple techniques (DLS, SP-ICP-MS, ICP-MS and ICP-OES)
    de la Calle, Inmaculada
    Menta, Mathieu
    Klein, Marlene
    Seby, Fabienne
    TALANTA, 2017, 171 : 291 - 306
  • [16] Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element Determination with ICP-MS
    Rantasa, Matjaz
    Majer, David
    Finsgar, Matjaz
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2023, (202):
  • [17] Determination of platinum and rhodium in dust and plant samples using microwave-assisted sample digestion and ICP-MS
    Niemelä, M
    Perämäki, P
    Piispanen, J
    Poikolainen, J
    ANALYTICA CHIMICA ACTA, 2004, 521 (02) : 137 - 142
  • [18] Single reaction chamber microwave digestion coupled with ICP-MS for the determination of ultra-trace mercury in rocks
    Liang, Minsi
    Liu, Hongtao
    ANALYTICAL METHODS, 2024, 16 (35) : 6049 - 6056
  • [19] Ultraviolet radiation combined with microwave-assisted wet digestion of Antarctic seaweeds for further determination of toxic elements by ICP-MS
    Mesko, Marcia F.
    Picoloto, Rochele S.
    Ferreira, Lizangela R.
    Costa, Vanize C.
    Pereira, Claudio M. P.
    Colepicolo, Pio
    Muller, Edson I.
    Flores, Erico M. M.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (01) : 260 - 266
  • [20] Study and determination of elemental impurities by ICP-MS in active pharmaceutical ingredients using single reaction chamber digestion in compliance with USP requirements
    Muller, Aline L. H.
    Oliveira, Jussiane S. S.
    Mello, Paola A.
    Muller, Edson I.
    Flores, Erico M. M.
    TALANTA, 2015, 136 : 161 - 169