Intra-aortic balloon pumping: effects on left ventricular diastolic function

被引:27
|
作者
Khir, AW
Price, S
Henein, MY
Parker, KH
Pepper, JR
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Bioengn, London SW7 2AZ, England
[2] Royal Brompton & Natl Heart Hosp, Echocardiog Dept, London SW3 6NP, England
关键词
intra-aortic balloon pump; diastolic function; coronary flow; mitral flow;
D O I
10.1016/S1010-7940(03)00268-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The intra-aortic balloon pump is the most widely used form of temporary cardiac assist and often utilised in patients before and after cardiac surgery. Several effects of balloon counter-pulsation have been reported previously, but its effect on left ventricular diastolic function has not been thoroughly investigated. The aim of this study is to examine the effect of the intra-aortic balloon pump on left ventricular wall motion and transmittal flow. Methods: We studied 20 patients in the intensive care unit, less than 36 It following cardiac surgery. We recorded left anterior descending coronary artery and transmitral E-wave flow velocities using transesophageal echocardiography pulsed Doppler. We also recorded left ventricular long axis free-wall movement using M-mode. The intra-aortic balloon pump was set to full augmentation and recordings were made at pumping cycles 1: 1, 1:2, 1:3, and when the pump was on stand-by, leaving a minimum of 5 min between the pumping modes to allow the return to control conditions. In order to eliminate time effects, the sequence of recording was varied between patients using a 4 by 4 Latin-square. Results: The peak diastolic left anterior descending coronary artery and transmitral E-wave flow velocities, and left ventricular free-wall early diastolic lengthening velocity increased significantly with intra-aortic balloon pumping cycles 1: 1, 1:2 and 1:3 compared to their value with the pump on stand-by, all P < 0.001. The increase in peak transmitral E-wave flow velocity correlated with the increase in peak left anterior descending coronary artery diastolic flow velocity (r = 0.74, P = 0.02), and with the increase in left ventricular free-wall early diastolic lengthening velocity (r = 0.80, P < 0.001). Conclusion: Using the intra-aortic balloon pump post-cardiac surgery significantly increases peak diastolic left anterior descending coronary artery flow velocities and left ventricular free-wall early diastolic lengthening velocity, whose increase explains the increase in peak transmitral E-wave velocity. Although coronary flow is epicardial and mitral flow is intracardial, their close relationship suggests an improvement in left ventricular diastolic function with intra-aortic balloon pump. (C) 2003 Elsevier Science B.V. All fights reserved.
引用
收藏
页码:277 / 282
页数:6
相关论文
共 50 条
  • [1] COMPARISON BETWEEN THE EFFECTS OF INTRA-AORTIC BALLOON PUMPING ON SYSTOLIC AND DIASTOLIC LEFT-VENTRICULAR FUNCTIONS
    ISKANDRIAN, AS
    HAKKI, AH
    BEMIS, CE
    MANDLER, J
    KANE, S
    CLINICAL RESEARCH, 1985, 33 (02): : A196 - A196
  • [2] INTRA-AORTIC BALLOON PUMPING - FOR MECHANICAL LEFT VENTRICULAR ASSISTANCE
    SCHOONMAKER, FW
    GROW, JB
    PREVEDEL, AE
    HOPEMAN, AR
    SMITH, D
    BRANTIGAN, C
    ROCKY MOUNTAIN MEDICAL JOURNAL, 1974, 71 (06): : 339 - 343
  • [3] EFFECTS OF INTRA-AORTIC BALLOON DIASTOLIC AUGMENTATION AND NITROPRUSSIDE ON POSTOPERATIVE REGIONAL LEFT-VENTRICULAR FUNCTION
    HILL, RC
    SINK, JD
    CHITWOOD, WR
    OLSEN, CO
    JONES, RN
    COX, JL
    WECHSLER, AS
    AMERICAN JOURNAL OF CARDIOLOGY, 1980, 45 (02): : 432 - 432
  • [4] EFFECT OF INTRA-AORTIC BALLOON PUMPING ON LEFT-VENTRICULAR FUNCTION - EVALUATION BY RADIONUCLIDE VENTRICULOGRAPHY
    ISKANDRIAN, AS
    COLBY, J
    HAKKI, AH
    BOSTON, B
    KANE, SA
    CLINICAL CARDIOLOGY, 1984, 7 (04) : 211 - 216
  • [5] EFFECTS OF PROLONGED INTRA-AORTIC BALLOON PUMPING
    MADRAS, PN
    LAIRD, JD
    IATRIDIS, E
    KANTROWITZ, AR
    BUCKLEY, MJ
    AUSTEN, WG
    TRANSACTIONS AMERICAN SOCIETY FOR ARTIFICIAL INTERNAL ORGANS, 1969, 15 : 400 - +
  • [6] INTRA-AORTIC BALLOON PUMPING
    WEBB, WR
    ANNALS OF THORACIC SURGERY, 1976, 21 (06): : 571 - 572
  • [7] INTRA-AORTIC BALLOON PUMPING
    MICHELS, R
    HUGENHOLTZ, PG
    CIRCULATION, 1979, 59 (04) : 846 - 846
  • [8] INTRA-AORTIC BALLOON PUMPING
    KAYSER, KL
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1977, 73 (03): : 487 - 487
  • [9] INTRA-AORTIC BALLOON PUMPING
    SPODICK, DH
    CIRCULATION, 1978, 58 (06) : 1212 - 1212
  • [10] INTRA-AORTIC BALLOON PUMPING
    不详
    LANCET, 1972, 2 (7789): : 1238 - +