Plan complexity quantification of dual-layer multi-leaf collimator for volumetric modulated arc therapy with Halcyon linac

被引:21
|
作者
Tamura, Mikoto [1 ]
Matsumoto, Kenji [2 ]
Otsuka, Masakazu [1 ,2 ]
Monzen, Hajime [1 ]
机构
[1] Kindai Univ, Grad Sch Med Sci, Dept Med Phys, 377-2 Onohigashi, Osakasayama, Osaka 5898511, Japan
[2] Kindai Univ Hosp, Dept Radiol, 377-2 Onohigashi, Osakasayama, Osaka 5898511, Japan
关键词
Plan complexity; Dual-layer MLC; Halcyon; VMAT; GAMMA INDEX ANALYSIS; QUALITY-ASSURANCE; IMRT; VMAT; METRICS;
D O I
10.1007/s13246-020-00891-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study characterized the plan complexity of the dual-layer multi-leaf collimator (MLC) performance for volumetric modulated arc therapy (VMAT) in Halcyon. Plan complexity metrics were computed independently for each MLC layer (proximal and distal): modulation complexity score (MCS), plan averaged beam area, irregularity, and modulation (PA, PI, and PM) were obtained. These were investigated when weighted by proportions of the effective proximal MLC and distal MLC (MCSw, PA(w), PIw, and PMw). Conventional plan complexity metrics for the effective 5.0 mm resolution MLC (MCS5, PA(5), PI5, PM5), small aperture score (SAS), effective distal MLC score (EDS), and MU were also evaluated. Forty-five consecutive VMAT plans were analyzed retrospectively for treatment sites of the prostate, head and neck (HN), and other parts. The mean values of the metrics for proximal MLC and distal MLC were 0.310 and 0.245 (MCS), 71.09 cm(2)and 55.92 cm(2)(PA), 6.24 and 10.19 (PI), and 0.593 and 0.645 (PM), respectively. The degree of plan complexity in the distal MLC was larger than in the proximal MLC. The percentage of the effective distal MLC was higher than that of the proximal MLC (mean EDS = 0.555). The MCSw, PA(w), PIw, and PM(w)showed significant differences between the prostate and HN groups like the PA(5), PI5, and PM5; however, the MCS(5)did not. Additionally, the MCS(5)correlated with MU and PM5, while the MCS(w)correlated with SAS and PA(5). The plan complexity for each MLC layer and subsequently combining these complexities might be useful to better understand the dual-layer MLC performance.
引用
收藏
页码:947 / 957
页数:11
相关论文
共 50 条
  • [31] Dosimetric comparison between volumetric modulated arc therapy on truebeam and halcyon of boost plan for total body irradiation
    Jaihow, Wanwanut
    Tawonwong, Tanawat
    Sanghangthum, Taweap
    HEALTH AND TECHNOLOGY, 2024,
  • [32] Direct impact analysis of multi-leaf collimator leaf position errors on dose distributions in volumetric modulated arc therapy: a pass rate calculation between measured planar doses with and without the position errors
    Tatsumi, D.
    Hosono, M. N.
    Nakada, R.
    Ishii, K.
    Tsutsumi, S.
    Inoue, M.
    Ichida, T.
    Miki, Y.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (20): : N237 - N246
  • [33] Description and evaluation of a new volumetric-modulated arc therapy plan complexity metric *
    Li, Guangjun
    Jiang, Wei
    Li, Yanlong
    Wang, Qiang
    Xiao, Jianghong
    Zhong, Renming
    Bai, Sen
    MEDICAL DOSIMETRY, 2021, 46 (02) : 188 - 194
  • [34] A Collimator Setting Optimization Algorithm for Dual-Arc Volumetric Modulated Arc Therapy in Pancreas Stereotactic Body Radiation Therapy
    Li, X.
    Wu, Q.
    Palta, M.
    Zhang, Y.
    Sheng, Y.
    Zhang, J.
    Wang, C.
    MEDICAL PHYSICS, 2019, 46 (06) : E472 - E472
  • [35] A Collimator Setting Optimization Algorithm for Dual-Arc Volumetric Modulated Arc Therapy in Pancreas Stereotactic Body Radiation Therapy
    Li, Xinyi
    Wu, Jackie
    Palta, Manisha
    Zhang, You
    Sheng, Yang
    Zhang, Jiahan
    Wang, Chunhao
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2019, 18
  • [36] Comparison Between TrueBeam Millennium And High Definition MLC With Halcyon Dual-Layer MLC: Impact of Collimator Leaf Width On Plan Quality for Stereotactic Body Radiotherapy for Spine Tumors
    Petroccia, H.
    Dong, L.
    Malajovich, I.
    Teo, B.
    Li, T.
    MEDICAL PHYSICS, 2018, 45 (06) : E589 - E589
  • [37] Development and verification of a time delivery model for prostate intensity modulated radiotherapy using a Siemens® Artiste™ 160 Multi-leaf Collimator Linac
    Nicola Fourie
    Omer A. Ali
    William I. D. Rae
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 51 - 56
  • [38] The Impact of Plan Complexity Parameters On the Plan Quality and Deliverability of Volumetric Modulated Arc Therapy with Canonical Correlation Analysis
    Jin, X.
    Yi, J.
    Xie, C.
    MEDICAL PHYSICS, 2016, 43 (06) : 3372 - 3372
  • [39] Impact of Plan Complexity On the Delivery Accuracy and Dosimetric Accuracy of Volumetric-Modulated Arc Therapy
    Li, G.
    Li, Y.
    Wang, Q.
    Xiao, J.
    Bai, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2892 - 2892
  • [40] Development and verification of a time delivery model for prostate intensity modulated radiotherapy using a Siemens® Artiste™ 160 Multi-leaf Collimator Linac
    Fourie, Nicola
    Ali, Omer A.
    Rae, William I. D.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 51 - 56