DOI-PET image reconstruction with accurate system modeling that reduces redundancy of the imaging system

被引:39
|
作者
Yamaya, T [1 ]
Hagiwara, N
Obi, T
Yamaguchi, M
Kita, K
Ohyama, N
Kitamura, K
Hasegawa, T
Haneishi, H
Murayama, H
机构
[1] Tokyo Inst Technol, Yokohama, Kanagawa 2268503, Japan
[2] Shimadzu Co Ltd, Kyoto 6048511, Japan
[3] Kitasato Univ, Sagamihara, Kanagawa 2288555, Japan
[4] Chiba Univ, Chiba 2638522, Japan
[5] Natl Inst Radiol Sci, Chiba 2638555, Japan
关键词
depth-of-interaction (DOI); imaging system model; maximum likelihood expectation maximization (ML-EM); positron emission tomography; statistical image reconstruction;
D O I
10.1109/TNS.2003.817307
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A high-performance positron emission tomography (PET) scanner, which measures depth-of-interaction (DOI) information, is under development at the National Institute of Radiological Sciences in Japan. Image reconstruction methods with accurate modeling of the system response functions have been successfully used to improve PET image quality. It is, however, difficult to apply these methods to the DOI-PET scanner because the dimension of DOI-PET data increases in proportion to the square of the number of DOI layers. In this paper, we propose a compressed imaging system model for DOI-PET image reconstruction, in order to reduce computational cost while keeping image quality. The basic idea of the proposed method is that the DOI-PET imaging system is highly redundant. First, DOI-PET data is transformed into compact data so that data bins with highly correlating sensitivity functions are combined. Then image reconstruction methods based on accurate system modeling, such as the maximum likelihood expectation maximization (ML-EM), are applied. The proposed method was applied to simulated data for the DOI-PET scanner operated in 2-D mode. Then the tradeoff between the background noise and the spatial resolution was investigated. Numerical simulation results show that the proposed method followed by ML-EM reduces computational cost effectively while keeping the advantages of the accurate system modeling and DOI information.
引用
收藏
页码:1404 / 1409
页数:6
相关论文
共 50 条
  • [1] DOI-PET image reconstruction with accurate system model reducing redundancy of imaging system
    Yamaya, T
    Hagiwara, N
    Obi, T
    Yamaguchi, M
    Kita, K
    Ohyama, N
    Kitamura, K
    Hasegawa, T
    Haneishi, H
    Murayama, H
    [J]. 2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1226 - 1230
  • [2] System modeling of small bore DOI-PET scanners for fast and accurate 3D image reconstruction
    Takahashi, Hisashi
    Yamaya, Taiga
    Kobayashi, Tetsuya
    Kitamura, Keishi
    Hasegawa, Tornoyuki
    Murayarna, Hideo
    Suga, Mikio
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3478 - +
  • [3] Investigation of a compact DOI-PET system for small animal imaging
    Kao, CM
    Schmand, M
    Gives, G
    Pan, XC
    Nutt, R
    Chen, CT
    [J]. 2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1292 - 1297
  • [4] Accurate image reconstruction for DOI-PET systems and its implications for the development of economic, compact PET (ezPET) systems
    Kao, CM
    Pan, XC
    Chen, CT
    [J]. 1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1363 - 1367
  • [5] Investigation of a whole-body DOI-PET system
    Ohi, Junichi
    Tonami, Hiromichi
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 223 - 226
  • [6] One-Pair Measurement System for Efficient Imaging Performance Evaluation of Prototype DOI-PET Detectors
    Tashima, Hideaki
    Yoshida, Eiji
    Hirano, Yoshiyuki
    Nishikido, Fumihiko
    Inadama, Naoko
    Murayama, Hideo
    Ito, Hiroshi
    Yamaya, Taiga
    [J]. 2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2523 - 2525
  • [7] A direct sinogram-restoration method for fast image reconstruction in compact DOI-PET systems
    Kao, CM
    Chen, CT
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) : 208 - 214
  • [8] Preliminary study of a DOI-PET detector with optical imaging capability
    Takahashi, Kei
    Inadama, Naoko
    Murayama, Hideo
    Yamaya, Taiga
    Yoshida, Eiji
    Nishikido, Fmihiko
    Shibuya, Kengo
    Lam, Chih Fung
    Oda, Ichiro
    Kitamura, Keishi
    Kawai, Hideyuki
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3318 - +
  • [9] Prototype integrated system of DOI-PET and the RF-coil specialized for simultaneous PET-MRI measurements
    Nishikido, Fumihiko
    Obata, Takayuki
    Inadama, Naoko
    Yoshida, Eiji
    Tashima, Hideaki
    Suga, Mikio
    Murayama, Hideo
    Ito, Hiroshi
    Yamaya, Taiga
    [J]. 2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2750 - 2752
  • [10] Preliminary performance evaluation of the prototype system for a brain DOI-PET scanner: jPET-D4
    Yamaya, T
    Hagiwara, N
    Obi, T
    Tsuda, TI
    Kitamura, K
    Hasegawa, T
    Haneishi, H
    Inadama, N
    Yoshida, E
    Murayama, H
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2610 - 2614