Ionization quenching correction for a 3D scintillator detector exposed to scanning proton beams

被引:8
|
作者
Alsanea, Fahed [1 ,2 ]
Darne, Chinmay [1 ]
Robertson, Daniel [3 ]
Beddar, Sam [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson, UTHlth, Grad Sch Biomed Sci, Houston, TX 77030 USA
[3] Mayo Clin Scottsdale Phoenix, Dept Radiat Oncol, Scottsdale, AZ USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2020年 / 65卷 / 07期
基金
美国国家卫生研究院;
关键词
dosimetry; proton therapy; scintillators; ionization quenching; scintillation detectors; PLASTIC SCINTILLATOR; LIQUID SCINTILLATOR; QUALITY-ASSURANCE; DOSIMETRY; INTENSITY; RANGE; MODEL;
D O I
10.1088/1361-6560/ab7876
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The ionization quenching phenomenon in scintillators must be corrected to obtain accurate dosimetry in particle therapy. The purpose of this study was to develop a methodology for correcting camera projection measurements of a 3D scintillator detector exposed to proton pencil beams. Birks' ionization quenching model and the energy deposition by secondary electrons (EDSE) model were used to correct the light captured by a prototype 3D scintillator detector. The detector was made of a 20 cm x 20 cm x 20 cm tank filled with liquid scintillator, and three cameras. The detector was exposed to four proton-beam energies (84.6, 100.9, 144.9, and 161.6 MeV) at The University of Texas MD Anderson Cancer Center's Proton Therapy Center. The dose and track averaged linear energy transfer (LET) were obtained using validated Monte Carlo (MC) simulations. The corrected light output was compared to the dose calculated by the MC simulation. Optical artefact corrections were used to correct for refraction at the air-scintillator interface, and image perspective. These corrections did not account for the non-orthogonal integration of data off the central axis of the image. Therefore, we compared the light output to an integrated MC dose and LET along the non-orthogonal path. After accounting for the non-orthogonal integration of the data, the corrected light output reduced the dose error at the Bragg peak region from 15% to 3% for low proton-beam energies. Overall, the doses at the Bragg peak region using the Birks' model and EDSE model were less than +/- 3% and +/- 7% of the MC dose, respectively. We have improved the application of Birks' model quenching corrections in 3D scintillators by numerically projecting the dose and LET 3D grid to camera projections. This study shows that scintillator projections can be corrected using average LET values at the central axes.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Ionization quenching correction for 3D scintillator detectors for spot scanning proton therapy
    Alsanea, F.
    Beddar, S.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S701 - S701
  • [2] Light yield and ionization quenching measurements of the 3D scintillator detector for proton therapy
    Alsanea, F.
    Darne, C.
    Robertson, D.
    Beddar, S.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S41 - S42
  • [3] Evaluating analytical ionization quenching correction models for 3D liquid organic scintillator detector
    Alsanea, F.
    Beddar, S.
    9TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, 2017, 847
  • [4] Voxelized Quenching Correction for Proton Pencil Beams in a Liquid Scintillator Detector
    Robertson, D.
    Mirkovic, D.
    Mohan, R.
    Beddar, S.
    MEDICAL PHYSICS, 2011, 38 (06) : 3816 - +
  • [5] 3D Scintillator Detector System for Proton Scanning Beam Therapy
    Darne, C. D.
    Alsanea, F.
    Robertson, D. G.
    Hojo, Y.
    Sahoo, N.
    Beddar, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : S89 - S90
  • [6] Measurement of proton quenching in a plastic scintillator detector
    Awe, C.
    Barbeau, P.
    Haghighat, A.
    Hedges, S.
    Johnson, T.
    Li, S.
    Link, J. M.
    Mascolino, V
    Runge, J.
    Steenis, J.
    Subedi, T.
    Walkup, K.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (02):
  • [7] Development of a novel proton CT system using a 3D scintillator detector
    Beddar, S.
    Darne, C. D.
    Robertson, D. G.
    11TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, IC3DDOSE, 2022, 2167
  • [8] Development of a liquid scintillator-based 3D detector for range measurements of spot-scanned proton beams
    Darne, C. D.
    Robertson, D.
    Beddar, S.
    9TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, 2017, 847
  • [9] Development of a 3D plastic Scintillator detector for a fast verification of ocular proton beam
    Ziri, H.
    Robertson, D.
    Beddar, S.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S29 - S29
  • [10] Fast range measurement of spot scanning proton beams using a volumetric liquid scintillator detector
    Hui, Cheuk Kai
    Robertson, Daniel
    Alsanea, Fahed
    Beddar, Sam
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (02):