On the uncertainty in the current waveform measurement of an ESD generator

被引:8
|
作者
Kang, TW [1 ]
Chung, YC
Won, SH
Kim, HT
机构
[1] Korea Res Inst Stand & Sci, Div Electromagnet Metrol, EMC Grp, Taejon 305600, South Korea
[2] Pohang Univ Sci & Technol, Dept Elect Engn, Pohang 790784, South Korea
关键词
current amplitude; electrostatic discharge; IEC standard; measurement uncertainty; output current waveform; rise time;
D O I
10.1109/15.902310
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The uncertainty in the current waveform measurement of an electrostatic discharge (ESD) generator is evaluated. The measurands are the current amplitude and the rise time of the output current waveform of the ESD generator. An intuitively simple model is proposed to evaluate the uncertainty in the current amplitude measurement. Type A and Type B evaluations for all contributions to the measurement uncertainty are performed to obtain the combined standard uncertainty, The evaluated expanded uncertainty (95.5% confidence level) of the current amplitude and the rise time at ESD voltages of 2, 4, 6, and 8 kV are within the specification of IEC 61000-4-2. The results show that the uncertainty in the current amplitude measurement stems from the voltage reading of the measuring equipment, the difference between the displayed and the actual voltages of the discharge tip of the ESD generator, and the inaccuracy of the delta time measurement of an oscilloscope, whereas the uncertainty in the rise time measurement mainly originates from the measuring equipment.
引用
收藏
页码:405 / 413
页数:9
相关论文
共 50 条
  • [31] ESD Current Spread Measurement Using Mesh Structure
    Lim, Jae-Deok
    Pornmerenke, David
    Lee, Jong-Sung
    Seol, Byong-Su
    2009 20TH INTERNATIONAL ZURICH SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY, 2009, : 229 - +
  • [32] Design and Realization of a Current Sensor for Impulse Current Waveform Measurement
    Yang, Xiaoguang
    Zhu, Bo
    Li, Congcong
    Jin, Shuangshuang
    Gao, Lijing
    Zhang, Shuai
    2016 IEEE CONFERENCE ON ELECTROMAGNETIC FIELD COMPUTATION (CEFC), 2016,
  • [33] Measurement of electric field waveform caused by micro gap ESD in a pair of spherical electrodes
    Ishigami, Shinobu
    Kawamata, Ken
    Minegishi, Shigeki
    Fujiwara, Osamu
    2017 ASIA-PACIFIC INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY (APEMC), 2017, : 63 - 65
  • [34] Sizing Optimization of the Synchronous Generator and the Measurement Uncertainty Analysis
    Kim, Youn-Hwan
    Kim, Hae-Joong
    Jung, Sang-Yong
    Moon, Jae-Won
    IEEE TRANSACTIONS ON MAGNETICS, 2017, 53 (11)
  • [35] Arbitrary light waveform generator for the validation of the light flicker measurement devices
    Bouroussis, Constantinos A.
    Tsirbas, Elias D.
    Canale, Laurent
    Dupuis, Pascal
    Topalis, Frangiskos, V
    IECON 2020: THE 46TH ANNUAL CONFERENCE OF THE IEEE INDUSTRIAL ELECTRONICS SOCIETY, 2020, : 3603 - 3608
  • [36] Digital waveform generator provides flexible frequency tuninfr for sensor measurement
    Slattery, C
    EDN, 2004, 49 (26) : 68 - +
  • [37] Design and Optimization of a High-Current Versatile Arbitrary Waveform Generator
    Alvarez-Gariburo, Ignacio
    Sarnago, Hector
    Lucia, Oscar
    Burdio, Jose M.
    2023 IEEE APPLIED POWER ELECTRONICS CONFERENCE AND EXPOSITION, APEC, 2023, : 710 - 714
  • [38] A multi-waveform pulsed current generator for slow kicker magnets
    Grawer, Gregor
    2019 21ST EUROPEAN CONFERENCE ON POWER ELECTRONICS AND APPLICATIONS (EPE '19 ECCE EUROPE), 2019,
  • [39] The measurement uncertainty of direct current Offset
    Liang, Zhi-Guo
    Journal of Test and Measurement Technology, 2004, 18 (03) : 253 - 258
  • [40] Measurement and analysis of impulse current due to corona on the ESD brush
    Hu, Xiao-Feng
    Liu, Shang-He
    Wu, Zhan-Cheng
    Wei, Ming
    Wang, Lei
    Gaodianya Jishu/High Voltage Engineering, 2007, 33 (09): : 171 - 174