The Influence of Cochlear Implant Electrode Position on Performance

被引:50
|
作者
van der Marel, Kim S. [1 ]
Briaire, Jeroen J. [1 ]
Verbist, Berit M. [2 ,3 ]
Muurling, Tjeerd J. [1 ]
Frijns, Johan H. M. [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Otorhinolaryngol, NL-2300 RC Leiden, Netherlands
[2] Leiden Univ, Med Ctr, Dept Radiol, NL-2300 RC Leiden, Netherlands
[3] Radboud Univ Nijmegen, Med Ctr, Dept Radiol, NL-6525 ED Nijmegen, Netherlands
关键词
Cochlear implantation; Electrode position; Insertion depth; Modiolus proximity; Performance scores; Phoneme scores; Word scores; SPEECH-PERCEPTION; INSERTION DEPTH; DEEP INSERTION; ADULTS; MORPHOLOGY; HEARING; ARRAY; DEAFNESS; OUTCOMES; CII;
D O I
10.1159/000377616
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objectives: To study the relation between variables related to cochlear implant electrode position and speech perception performance scores in a large patient population. Design: The study sample consisted of 203 patients implanted with a CII or HiRes90K implant with a HiFocus 1 or 1J electrode of Advanced Bionics. Phoneme and word score averages for the 1- and 2-year follow-up were calculated for 41 prelingually deaf and 162 postlingually deaf patients. Analyses to reveal correlations between these performance outcomes and 6 position-related variables (angle of most basal electrode contact, surgical insertion angle, surgical insertion, wrapping factor, angular insertion depth, linear insertion depth) were executed. The scalar location, as an indication for the presence of intracochlear trauma, and modiolus proximity beyond the basal turn were not evaluated in this study. In addition, different patient-specific variables (age at implantation, age at onset of hearing loss, duration of deafness, preoperative phoneme and word scores) were tested for correlation with performance. Results: The performance scores of prelingual patients were correlated with age at onset of hearing loss, duration of deafness and preoperative scores. For the postlingual patients, performance showed correlations with all 5 patient-specific variables. None of the 6 position-related variables influenced speech perception in cochlear implant patients. Conclusions: Although several patient-specific variables showed correlations with speech perception outcomes, not one of the studied angular and linear position-related variables turned out to have a demonstrable influence on performance. (C) 2015 S. Karger AG, Basel
引用
收藏
页码:202 / 211
页数:10
相关论文
共 50 条
  • [1] Diversity in Cochlear Morphology and Its Influence on Cochlear Implant Electrode Position
    van der Marel, Kim S.
    Briaire, Jeroen J.
    Wolterbeek, Ron
    Snel-Bongers, Jorien
    Verbist, Berit M.
    Frijns, Johan H. M.
    EAR AND HEARING, 2014, 35 (01): : E9 - E20
  • [2] Intracochlear Electrocochleography: Influence of Scalar Position of the Cochlear Implant Electrode on Postinsertion Results
    Riggs, William J.
    Dwyer, Robert T.
    Holder, Jourdan T.
    Mattingly, Jameson K.
    Ortmann, Amanda
    Noble, Jack H.
    Dawant, Benoit M.
    Valenzuela, Carla V.
    O'Connell, Brendan P.
    Harris, Michael S.
    Litvak, Leonid M.
    Koka, Kanthaiah
    Buchman, Craig A.
    Labadie, Robert F.
    Adunka, Oliver F.
    OTOLOGY & NEUROTOLOGY, 2019, 40 (05) : E503 - E510
  • [3] Evaluation of Rigid Cochlear Models for Measuring Cochlear Implant Electrode Position
    Cakir, Ahmet
    Labadie, Robert F.
    Zuniga, M. Geraldine
    Dawant, Benoit M.
    Noble, Jack H.
    OTOLOGY & NEUROTOLOGY, 2016, 37 (10) : 1560 - 1564
  • [4] Development of Insertion Models Predicting Cochlear Implant Electrode Position
    van der Marel, Kim S.
    Briaire, Jeroen J.
    Wolterbeek, Ron
    Verbist, Berit M.
    Frijns, Johan H. M.
    EAR AND HEARING, 2016, 37 (04): : 473 - 482
  • [5] Word Recognition with a Cochlear Implant in Relation to Prediction and Electrode Position
    Franke-Trieger, Annett
    Lailach, Susen
    Shetty, Joshua
    Murrmann, Katrin
    Zahnert, Thomas
    Neudert, Marcus
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (01)
  • [6] COCHLEAR IMPLANT ELECTRODE
    CLARK, GM
    HALLWORTH, RJ
    ZDANIUS, K
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1975, 89 (08): : 787 - 792
  • [7] Psychophysical Tuning Curves as a Correlate of Electrode Position in Cochlear Implant Listeners
    Lindsay DeVries
    Julie G. Arenberg
    Journal of the Association for Research in Otolaryngology, 2018, 19 : 571 - 587
  • [8] Psychophysical Tuning Curves as a Correlate of Electrode Position in Cochlear Implant Listeners
    DeVries, Lindsay
    Arenberg, Julie G.
    JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2018, 19 (05): : 571 - 587
  • [9] Measurements of electrode position inside the cochlea for different cochlear implant systems
    Kós, MI
    Boëx, C
    Sigrist, A
    Guyot, JP
    Pelizzone, M
    ACTA OTO-LARYNGOLOGICA, 2005, 125 (05) : 474 - 480
  • [10] The influence of cochlear morphology on the final electrode array position
    Ketterer, M. C.
    Aschendorff, A.
    Arndt, S.
    Hassepass, F.
    Wesarg, T.
    Laszig, R.
    Beck, R.
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2018, 275 (02) : 385 - 394