Entrance skin dose estimates derived from dose-area product measurements in interventional radiological procedures

被引:98
|
作者
McParland, BJ [1 ]
机构
[1] King Fahad Natl Guard Hosp, Dept Med Imaging, Riyadh 11426, Saudi Arabia
来源
BRITISH JOURNAL OF RADIOLOGY | 1998年 / 71卷 / 852期
关键词
D O I
10.1259/bjr.71.852.10319003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Patient skin doses resulting from interventional radiological procedures have the potential to exceed threshold doses for deterministic effects such as erythema and epilation. If the irradiation geometry is known, the entrance skin dose can be estimated fl om the measured dose-area product. For each of 10 non-coronary interventional procedures, a nominal geometry was identified. From a previous survey of patient dose-area products, the entrance skin doses were estimated under the assumption that all procedures were performed with the nominal geometry specific to it. An analysis of the uncertainties in these doses caused by realistic deviations from the nominal geometry was also performed and it was shown that the estimated entrance skin dose values are at least to within 40%, and generally to within about 30%, of those actually received. For example, the median estimated entrance skin doses for the posteroanterior and lateral projections of cerebral angiography were 100 and 110 mGy, respectively, and for hepatic angiography 425 mGy. The largest entrance skin dose estimate for a single projection was for the angiography component of a CT arterial portography procedure at 670 mGy. Comparisons between entrance skin dose estimates obtained from this study are made with data from other interventional radiology patient dose surveys.
引用
收藏
页码:1288 / 1295
页数:8
相关论文
共 50 条
  • [11] Correlation of patient skin doses in cardiac interventional radiology with dose-area product
    van de Putte, S
    Verhaegen, F
    Taeymans, Y
    Thierens, H
    BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (869): : 504 - 513
  • [12] Dose-area product measurements in paediatric radiography
    McDonald, S
    Martin, CJ
    Darragh, CL
    Graham, DT
    BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (820): : 318 - 325
  • [13] Estimate Eye Lens Dose by Using Personal Dose Equivalent (HP (10)) and Dose-Area Product (DAP) in Interventional Procedures
    Hu, P.
    Wang, J.
    Hu, W.
    Liu, H.
    MEDICAL PHYSICS, 2018, 45 (06) : E276 - E276
  • [14] Dose-area product in pediatric barium meal procedures
    Machado, S. O.
    Schelin, H.
    Denyak, V.
    Filipov, D.
    Bunick, A.
    Ledesma, J.
    Paschuk, S. A.
    RADIATION PHYSICS AND CHEMISTRY, 2019, 155 : 53 - 55
  • [15] Comparison of effective doses obtained from dose-area product and air kerma measurements in interventional radiology
    Bor, D
    Sancak, T
    Olgar, T
    Elcim, Y
    Adanali, A
    Sanlidilek, U
    Akyar, S
    BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916): : 315 - 322
  • [16] Dose-area product measurements in panoramic dental radiology
    Poppe, B.
    Looe, H. K.
    Pfaffenberger, A.
    Chofor, N.
    Eenboom, F.
    Sering, M.
    Ruehmann, A.
    Poplawski, A.
    Willborn, K.
    RADIATION PROTECTION DOSIMETRY, 2007, 123 (01) : 131 - 134
  • [17] Skin dose and dose-area product values in patients undergoing intracoronary brachytherapy
    Vano, E
    Prieto, C
    Fernandez, JM
    Gonzalez, L
    Sabate, M
    Galvan, C
    BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (901): : 32 - 38
  • [18] RELATIONSHIPS BETWEEN ENTRANCE SKIN DOSE, EFFECTIVE DOSE AND DOSE AREA PRODUCT FOR PATIENTS IN DIAGNOSTIC AND INTERVENTIONAL CARDIAC PROCEDURES (vol 90, pg 141, 2000)
    Hansson, B.
    Karambatsakidou, A.
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (03) : 341 - 341
  • [19] ESTIMATION OF EFFECTIVE DOSE FROM DOSE-AREA PRODUCT MEASUREMENTS FOR BARIUM MEALS AND BARIUM ENEMAS
    HART, D
    WALL, BF
    BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (797): : 485 - 489
  • [20] DOSE-AREA PRODUCT AND BODY DOSES
    PETOUSSIHENSS, N
    PANZER, W
    ZANKL, M
    DREXLER, G