Longevity of implantable cardioverter defibrillators: a comparison among manufacturers and over time

被引:33
|
作者
von Gunten, Simon [1 ]
Schaer, Beat A. [1 ]
Yap, Sing-Chien [2 ]
Szili-Torok, Tamas [2 ]
Kuhne, Michael [1 ]
Sticherling, Christian [1 ]
Osswald, Stefan [1 ]
Theuns, Dominic A. M. J. [2 ]
机构
[1] Univ Basel Hosp, Dept Cardiol, CH-4031 Basel, Switzerland
[2] Erasmus MC, Dept Cardiol, Rotterdam, Netherlands
来源
EUROPACE | 2016年 / 18卷 / 05期
关键词
Longevity; Implantable cardioverter-defibrillator; Survival; Cardiac resynchronization therapy; CARDIAC RESYNCHRONIZATION THERAPY; CLINICAL CIRCUMSTANCES; LIFE; GENERATION; MANAGEMENT; MORTALITY; INFECTION; IMPACT; LEADS; RISK;
D O I
10.1093/europace/euv296
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Longevity of implantable cardioverter defibrillators (ICDs) is crucial for patients and healthcare systems as replacements impact on infection rates and cost-effectiveness. Aim was to determine longevity using very large databases of two teaching hospitals with a high number of replacements and a rather homogeneous distribution among manufacturers. The study population consists of all patients in whom an ICD was inserted in. All ICD manufacturers operating in Switzerland and the Netherlands and all implanted ICDs were included. Implantable cardioverter defibrillator replacements due to normal battery depletion were considered events, and other replacements were censored. Longevity was assessed depending on manufacturers, pacing mode, implant before/after 2006, and all parameters combined. We analysed data from 3436 patients in whom 4881 ICDs [44.2% VVI-ICDs, 27.4% DDD-ICDs, 26.3% cardiac resynchronization therapy (CRT)-ICDs, 2.0% subcutaneous ICDs] were implanted. The four major manufacturers had implant shares between 18.4 and 31.5%. Replacement due to battery depletion (27.4%) was performed for 1339 ICDs. Patient survival at 5 years was 80.1%. Longevity at 5 years improved in contemporary compared with elderly ICDs [63.9-80.6% across all ICDs, of 73.7-92.1% in VVIs, 58.2-76.1% in DDDs, and of 47.1-66.3% in CRT defibrillators, allP value < 0.05]. Remarkable differences were seen among manufacturers, and those with better performance in elderly ICDs were not those with better performance in contemporary ones. Implantable cardioverter defibrillator longevity increased in contemporary models independent of manufacturer and pacing mode. Still, significant differences exist among manufacturers. These results might impact on device selection.
引用
收藏
页码:710 / 717
页数:8
相关论文
共 50 条
  • [21] Significant differences in the expected versus observed longevity of implantable cardioverter defibrillators (ICDs)
    Shafat, Tali
    Baumfeld, Yael
    Novack, Victor
    Konstantino, Yuval
    Amit, Guy
    CLINICAL RESEARCH IN CARDIOLOGY, 2013, 102 (01) : 43 - 49
  • [22] Significant differences in the expected versus observed longevity of implantable cardioverter defibrillators (ICDs)
    Tali Shafat
    Yael Baumfeld
    Victor Novack
    Yuval Konstantino
    Guy Amit
    Clinical Research in Cardiology, 2013, 102 : 43 - 49
  • [23] Implantable cardioverter-defibrillators
    DiMarco, JP
    NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (19): : 1836 - 1847
  • [24] Deactivating implantable cardioverter defibrillators
    Lynn, J
    ANNALS OF INTERNAL MEDICINE, 2005, 143 (09) : 691 - 691
  • [25] Athletes with Implantable Cardioverter Defibrillators
    Ponamgi, Shiva P.
    DeSimone, Christopher V.
    Ackerman, Michael J.
    CLINICS IN SPORTS MEDICINE, 2015, 34 (03) : 473 - +
  • [26] IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS
    KUTALEK, SP
    DREIFUS, LS
    ADVANCES IN INTERNAL MEDICINE, VOLUME 38, 1993, 38 : 421 - 438
  • [27] Implantable cardioverter-defibrillators
    Pinski, SL
    Fahy, GJ
    AMERICAN JOURNAL OF MEDICINE, 1999, 106 (04): : 446 - 458
  • [28] Advances in implantable cardioverter defibrillators
    Swygman, C
    Wang, PJ
    Link, MS
    Homoud, MK
    Estes, NAM
    CURRENT OPINION IN CARDIOLOGY, 2002, 17 (01) : 24 - 28
  • [29] Deactivation of implantable cardioverter defibrillators
    Wan, Darryl
    Chakrabarti, Santabhanu
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 2021, 193 (33) : E1326 - E1327
  • [30] Implantable cardioverter-defibrillators
    Zaidan, JR
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1999, 13 (04) : 475 - 483