Characterization of radiation dose from tube current modulated CT examinations with considerations of both patient size and variable tube current

被引:4
|
作者
Li, Xinhua [1 ]
Yang, Kai [1 ]
Liu, Bob [1 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Div Diagnost Imaging Phys, Boston, MA 02114 USA
关键词
characterization; computed tomography (CT); patient size; radiation dose; tube current modulation;
D O I
10.1002/mp.12460
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The volume CT dose index (CTDIvol) and the size-specific dose estimate (SSDE) are widely used for monitoring patient dose from CT examinations. Both metrics may represent the average dose over the central scan plane of the CTDI phantom or the patient under constant tube current (mA), but they are not intended for the tube current modulation (TCM)-enabled CT examinations, in which the peak dose across the scanned range may not be at the scan range center. To overcome the limitation, this paper illustrates an alternative approach, its implementation, and the relationship between longitudinal dose distribution D-L(z) in water cylinder and mA line shape, scan length, as well as phantom diameter. Methods: A dose calculation algorithm and the published data by Li etal. [Med. Phys. 40, 031903 (10pp.) (2013); 41, 111910 (5pp.) (2014)] were used to calculate D-L(z) for the central and peripheral axes of 10- to 50-cm diameter water phantoms undergoing CT scans of one constant and three variable mA distributions, each of which in three scan lengths of 10, 28.6, and 50cm. All scans had an identical average tube current over the scan ranges. The results in the scanned ranges were used to assess the D-L(z) to mA(z) ratios, and their coefficients of variation (CV=stdev/mean) were used to compare the line shapes of D-L(z) and mA(z) for congruence: identical line shapes would result in CV=0, but largely different line shapes would result in high CV. Results: In 30-cm diameter water phantom, the line shape of D-L(z) was largely different from that of mA(z). CV was higher in a variable mA scan than in a constant mA scan. As the scan length of variable mA scan increased, CV mostly decreased, and the line shape of D-L(z) more closely resembled that of mA(z). When two phantom axes were compared, CV was smaller and the line shape of D-L(z) more closely resembled that of mA(z) on the peripheral axis than on the central axis. In 41 water phantoms included in this study, CV mostly increased with phantom diameter, and approached the limiting levels on the peripheral axes of large phantoms. In constant mA scans, CV ranged from 5.5% to 14.0% on the phantom central axes and from 4.6% to 6.4% on the phantom peripheral axes. However, in variable tube current scans, CV ranged from 7.4% to 70.0% on the phantom central axes and from 5.1% to 35.9% on the phantom peripheral axes. Conclusion: D-L(z) (water) may be advantageous over current CT dose metrics in characterizing the dose dependences on both patient size and mA line shape from tube current modulated examinations. Evaluating D-L(z) (water) with the water equivalent diameter and tube current curve from clinical examinations has a potential to improve CT dose monitoring program. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:5413 / 5422
页数:10
相关论文
共 50 条
  • [1] Estimating fetal dose from tube current-modulated (TCM) and fixed tube current (FTC) abdominal/pelvis CT examinations
    Hardy, Anthony J.
    Angel, Erin
    Bostani, Maryam
    Cagnon, Chris
    McNitt-Gray, Michael
    MEDICAL PHYSICS, 2019, 46 (06) : 2729 - 2743
  • [2] Fetal Dose Estimates for Tube Current Modulated (TCM) and Fixed Tube Current (FTC) Abdominal/Pelvic CT Examinations of Pregnant Patients
    Hardy, A.
    Bostani, M.
    Angel, E.
    Cagnon, C.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2019, 46 (06) : E519 - E519
  • [3] The relationship between organ dose and patient size in tube current modulated adult thoracic CT scans
    Khatonabadi, Maryam
    Zhang, Di
    Yang, Jeffrey
    DeMarco, John J.
    Cagnon, Chris C.
    McNitt-Gray, Michael F.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [4] The Effects of Patient Size on Organ Doses in Adaptive Tube Current Modulated CT
    Fisher, R.
    Hintenlang, D.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [5] Radiation Dose in Abdominal Computed Tomography: The Role of Patient Size and the Selection of Tube Current
    Aldrich, John E.
    Chang, Silvia D.
    Bilawich, Ana-Marie
    Mayo, John R.
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2006, 57 (03): : 152 - 158
  • [6] Patient Specific Tube Current Modulation for CT Dose Reduction
    Jin, Yannan
    Yin, Zhye
    Yao, Yangyang
    Wang, Hui
    Wu, Mingye
    Kalra, Mannudeep
    De Man, Bruno
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [7] Estimating Organ Dose From CT Scans Performed with Tube Current Modulated Scans
    Khatonabadi, M.
    Chu, G.
    Oria, D.
    Cagnon, C.
    DeMarco, J.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Dose conversion coefficients for CT examinations of adults with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (20): : 6243 - 6261
  • [9] Dose conversion coefficients for paediatric CT examinations with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (20): : 6309 - 6326
  • [10] Dose reduction in body CT examinations by anatomically adapted tube current modulation
    Bautz, WA
    Heim, I
    Wolf, H
    Kalender, WA
    RADIOLOGY, 1997, 205 : 1399 - 1399