Efficient temperature corrections for gamma-ray energy reconstruction in 3-D position-sensitive CdZnTe detectors

被引:2
|
作者
Xia, Jiawei [1 ]
Zhu, Yuefeng [1 ]
He, Zhong [1 ]
机构
[1] 2355 Bonisteel Blvd, Ann Arbor, MI 48109 USA
关键词
CdZnTe; Temperature; 3-D reconstruction;
D O I
10.1016/j.nima.2018.10.018
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Maintaining good energy resolution in 3-D position-sensitive cadmium zinc telluride (CdZnTe) detectors without temperature regulation is desired in field operations to lower power consumption. This requires several calibration measurements at various ambient temperatures. However, required calibration times and storage space for calibration data increase linearly with working temperature range. In this experiment, gamma-ray interactions were measured by 3-D, position-sensitive CdZnTe detectors read out via VAD_UM v2.2 ASICs at various ambient temperatures. Physical causes of changes in detector behavior were studied. A proposed method to reconstruct events using only one complete calibration measurement, and several faster measurements at different ambient temperatures is discussed. Calibration data estimated using the proposed method showed almost identical performance compared to measured calibration data for measurements for various ambient temperatures at equilibrium. The paper also discusses experimental results that showed good resolution in 3-D CdZnTe detector systems can be achieved even if temperature regulation is fully discarded.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Intelligent gamma-ray spectroscopy using 3-D position-sensitive detectors
    Lehner, CE
    He, Z
    Knoll, GE
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (04) : 1090 - 1097
  • [2] 3-D position sensitive CdZnTe gamma-ray spectrometers
    He, Z
    Li, W
    Knoll, GF
    Wehe, DK
    Berry, J
    Stahle, CM
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3): : 173 - 178
  • [3] DOI correction for gamma ray energy reconstruction based on energy segment in 3D position-sensitive CdZnTe detectors
    Li, Y.
    Gong, P.
    Tang, X.
    Hu, Z.
    Wang, P.
    Tian, F.
    Wu, S.
    Ye, M.
    Zhou, C.
    Zhu, X.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (03)
  • [4] Performance of 3-D Position Sensitive CdZnTe Detectors for Gamma-Ray Energies above 1 MeV
    Zhang, Feng
    Kaye, William R.
    He, Zhong
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2012 - 2016
  • [5] Detection of gamma ray polarization using a 3-D position-sensitive CdZnTe detector
    Xu, D
    He, Z
    Zhang, F
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (04) : 1160 - 1164
  • [6] Time-Encoded Gamma-Ray Imaging Using a 3-D Position-Sensitive CdZnTe Detector Array
    Brown, Steven T.
    Goodman, David
    Chu, Jiyang
    Williams, Bennett
    Williamson, Martin R.
    He, Zhong
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2020, 67 (02) : 464 - 472
  • [7] Gamma-ray imaging with position-sensitive HPGe detectors
    Vetter, K
    Burks, M
    Mihailescu, L
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (1-2): : 322 - 327
  • [8] Intelligent gamma-ray spectroscopy using 3D position-sensitive detectors
    Lehner, CE
    He, Z
    Knoll, GF
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 411 - 415
  • [9] Low Background Measurements using 3-D Position-Sensitive CdZnTe Detectors
    Boucher, Yvan A.
    He, Zhong
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [10] Sensitivity Image Compensation in Pixelated, 3-D Position-Sensitive CdZnTe Detectors
    Williams, Bennett
    He, Zhong
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,