Implementation of a novel pulsed ultrasonic Doppler velocity profiler

被引:0
|
作者
Luo, SN [1 ]
Liu, Y [1 ]
Feng, GP [1 ]
机构
[1] Tsing Hua Univ, Dept Precis Instruments & Mechanol, Beijing 100084, Peoples R China
来源
FIFTH INTERNATIONAL SYMPOSIUM ON INSTRUMENTATION AND CONTROL TECHNOLOGY | 2003年 / 5253卷
关键词
pulsed Doppler echography; modified covariance; velocity profiler;
D O I
10.1117/12.521372
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
The objective of the work is to apply advanced digital signal processing (DSP) and pulsed Doppler echography techniques to analyze velocity profile of the measured fluid in a closed full-pipe. In order to reliably measure velocity profile where non-intrusive ultrasonic Doppler velocity profilers (UDVPs) cannot properly transmit and receive ultrasonic signals, a single-crystal transducer was inserted directly into the fluid stream. Then ultrasonic signals were transmitted and received based on the pulsed Doppler echography theory. A modified covariance (MCOV) estimator and a TMS320VC5410 DSP chip with a high-speed up to 160 MIPS were employed to analyze spectrum of the software-gated echo signals. Finally, computer simulations were also presented to demonstrate effectiveness of the novel UDVP. Results indicate that this instrument is particularly effective in terms of velocity profiling and can be widely applied for flow measurement and fluid dynamics research.
引用
收藏
页码:75 / 78
页数:4
相关论文
共 50 条
  • [41] Effects of the number of pulse repetitions and noise on the velocity data from the ultrasonic pulsed Doppler method with different algorithms
    Murakawa, Hideki
    Sugimoto, Katsumi
    Takenaka, Nobuyuki
    FLOW MEASUREMENT AND INSTRUMENTATION, 2014, 40 : 9 - 18
  • [42] Simultaneous measurements of the solid particles velocity and concentration profiles in two phase flow by pulsed ultrasonic doppler velocimetry
    University of Ibn Khaldoun of Tiaret, Research Laboratory of Industrial Technologies, Algeria, Algeria
    J. Braz. Soc. Mech. Sci. Eng., 2009, 4 (333-343):
  • [43] PULSED ULTRASONIC DOPPLER BLOOD-FLOW SENSING
    BAKER, DW
    IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1970, SU17 (03): : 170 - &
  • [44] ANNULAR ARRAYS FOR QUANTITATIVE PULSED DOPPLER ULTRASONIC FLOWMETERS
    FU, CC
    GERZBERG, L
    ULTRASONIC IMAGING, 1983, 5 (01) : 1 - 16
  • [45] PULMONARY REGURGITATION STUDIED WITH THE ULTRASONIC PULSED DOPPLER TECHNIQUE
    MIYATAKE, K
    OKAMOTO, M
    KINOSHITA, N
    MATSUHISA, M
    NAGATA, S
    BEPPU, S
    PARK, YD
    SAKAKIBARA, H
    NIMURA, Y
    CIRCULATION, 1982, 65 (05) : 969 - 976
  • [46] AN IMPLANTABLE MULTICHANNEL ULTRASONIC PULSED DOPPLER BLOOD FLOWMETER
    YEUNG, KWW
    SEBEK, JJ
    MEINDL, JD
    IMAGES OF THE TWENTY-FIRST CENTURY, PTS 1-6, 1989, 11 : 73 - 74
  • [47] PULSED ULTRASONIC DOPPLER FLOWMETER BIOLOGICAL AND ENGINEERING APPLICATIONS
    BAKER, DW
    IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1970, SU17 (01): : 65 - &
  • [48] NON-ALIASING ULTRASONIC PULSED DOPPLER MEASUREMENT
    NISHIYAMA, H
    KATAKURA, K
    TOSHIO, O
    ULTRASONIC IMAGING, 1988, 10 (01) : 70 - 70
  • [49] Application of Short Time Fourier Transform in Ultrasonic Velocity Profiler on Bubbly Flow
    Wongsaroj, Wongsakorn
    Hamdani, Ari
    Thong-un, Natee
    Takahashi, Hideharu
    Treenuson, Weerachon
    Kikura, Hiroshige
    2018 6TH INTERNATIONAL ELECTRICAL ENGINEERING CONGRESS (IEECON), 2018,
  • [50] RESOLUTION OF FREQUENCY ALIASES IN ULTRASONIC PULSED DOPPLER VELOCIMETERS
    HARTLEY, CJ
    IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1981, 28 (02): : 69 - 75