Shock Due to an Obstructed Endotracheal Tube

被引:1
|
作者
Chua, Stefanie Foong Ling [1 ]
Chan, Chi Ho [1 ]
Thangavelautham, Suhitharan [1 ]
机构
[1] Singapore Gen Hosp, Singapore, Singapore
来源
JOURNAL OF CRITICAL CARE MEDICINE | 2021年 / 7卷 / 04期
关键词
obstruction; endotracheal tube; mucus plug; bronchoscopy; ball-valve;
D O I
10.2478/jccm-2021-0027
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Endotracheal tube obstruction by a mucus plug causing a ball-valve effect is a rare but significant complication. The inability to pass a suction catheter through the endotracheal tube with high peak and plateau pressure differences are classical features of an endotracheal tube obstruction. A case is described of endotracheal tube obstruction from a mucus plug that compounded severe respiratory acidosis and hypotension in a patient who simultaneously had abdominal compartment syndrome. The mucus plug was not identified until a bronchoscopic assessment of the airway was performed. Due to the absence of classical signs, the delayed identification of the obstructing mucus plug exacerbated diagnostic confusion. It resulted in various treatments being trialed whilst the patient continued to deteriorate from the evasive offending culprit. We suggest that earlier and more routine use of bronchoscopy should be employed in an intensive care unit, especially as a definitive way to rule out endotracheal obstruction.
引用
收藏
页码:308 / 311
页数:4
相关论文
共 50 条
  • [31] Large air leak from an endotracheal tube due to a manufacturing defect
    Lewer, BMF
    Karim, Z
    Henderson, RS
    [J]. ANESTHESIA AND ANALGESIA, 1997, 85 (04): : 944 - 945
  • [32] Right Tracheobronchial Injury Due to Endotracheal Tube Malpositioning: A Case Presentation
    Uckun, Serkan
    Kuzucuoglu, Tamer
    [J]. JOURNAL OF THE TURKISH SOCIETY OF INTENSIVE CARE-TURK YOGUN BAKM DERNEGI DERGISI, 2014, 12 (03): : 106 - 109
  • [33] ATTENUATION OF SHOCK IN A SHOCK TUBE DUE TO EFFECT OF WALL BOUNDARY LAYER
    MURTHY, VS
    ANDERSON, GF
    [J]. BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1968, 13 (05): : 795 - &
  • [34] Shock wave modulation due to discharged plasma using a shock tube
    Matsuda, Atsushi
    Kondo, Yoshiaki
    Aoyama, Naoki
    [J]. MECHANICAL ENGINEERING JOURNAL, 2016, 3 (06):
  • [35] SHOCK CURVATURE DUE TO BOUNDARY-LAYER EFFECTS IN A SHOCK TUBE
    HARTUNIAN, RA
    [J]. PHYSICS OF FLUIDS, 1961, 4 (09) : 1059 - 1063
  • [36] Conversion of oral endotracheal tube to nasal endotracheal tube without laryngoscopy
    Gholami, Amir S.
    Mourra, Ahmad A.
    [J]. JOURNAL OF CLINICAL ANESTHESIA, 2018, 51 : 118 - 120
  • [37] Endotracheal tube as a nasogastric tube introducer
    Kopterides, Petros
    Papageorgiou, Christos
    Dimopoulos, Georgios
    Armaganidis, Apostolos
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2009, 56 (11): : 874 - 875
  • [38] Endotracheal tube as a nasogastric tube introducer
    Petros Kopterides
    Christos Papageorgiou
    Georgios Dimopoulos
    Apostolos Armaganidis
    [J]. Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 2009, 56 : 874 - 875
  • [39] THE ENDOTRACHEAL-TUBE AS A TRACHEOTOMY TUBE
    BETTEZ, M
    MAVES, MD
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1991, 105 (03) : 480 - 482
  • [40] Airway Obstruction Due to Endotracheal Tube's Lumen Collapse Secondary to Cuff
    Davis, Dan
    Murphy, Joseph
    Pop, Radu B.
    Szmuk, Peter
    [J]. ANESTHESIA AND ANALGESIA, 2011, 112 (06): : 1511 - 1512