Imaging Corneal Biomechanical Responses to Ocular Pulse Using High-Frequency Ultrasound

被引:30
|
作者
Pavlatos, Elias [1 ]
Chen, Hong [1 ,2 ]
Clayson, Keyton [1 ,3 ]
Pan, Xueliang [4 ]
Liu, Jun [1 ,3 ,5 ]
机构
[1] Ohio State Univ, Dept Biomed Engn, Columbus, OH 43210 USA
[2] Vasomedical Inc, Westbury, NY 11590 USA
[3] Ohio State Univ, Biophys Program, Columbus, OH 43210 USA
[4] Ohio State Univ, Dept Biomed Informat, Ctr Biostat, Columbus, OH 43210 USA
[5] Ohio State Univ, Dept Ophthalmol & Visual Sci, Columbus, OH 43210 USA
基金
美国国家卫生研究院;
关键词
Eye; biomechanical modeling; ultrasound; validation; ELASTICITY; DEFORMATION; KERATOCONUS; HEALTHY; SCLERA; STRAIN;
D O I
10.1109/TMI.2017.2775146
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Imaging corneal biomechanical changes or abnormalities is important for better clinical diagnosis and treatment of corneal diseases. We propose a novel ultrasound-based method, called ocular pulse elastography (OPE), to image corneal deformation during the naturally occurring ocular pulse. Experiments on animal and human donor eyes, as well as synthetic radiofrequency (RF) data, were used to evaluate the efficacy of the OPE method. Using very high-frequency ultrasound (center frequency = 55 MHz), correlation-based speckle tracking yielded an accuracy of less than 10% error for axial tissue displacements of 0.5 mu m or above. Satisfactory speckle tracking was achieved for out-of-plane displacements up to 32 mu m. Using synthetic RF data with or without a pre-defined uniform strain, the OPE method detected strains down to 0.0001 axially and 0.00025 laterally with an error less than 10%. Experiments in human donor eyes showed excellent repeatability with an intraclass correlation of 0.98. The measurement outcome from OPE was also shown to be highly correlated with that of standard inflation. These results suggest the feasibility of OPE as a potential clinical tool for evaluating corneal biomechanics in vivo.
引用
收藏
页码:663 / 670
页数:8
相关论文
共 50 条
  • [1] Ocular Pulse Elastography: Imaging Corneal Biomechanical Responses to Simulated Ocular Pulse Using Ultrasound
    Clayson, Keyton
    Pavlatos, Elias
    Pan, Xueliang
    Sandwisch, Thomas
    Ma, Yanhui
    Liu, Jun
    [J]. TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2020, 9 (01):
  • [2] A Multifunctional, Reconfigurable Pulse Generator for High-Frequency Ultrasound Imaging
    Qiu, Weibao
    Yu, Yanyan
    Tsang, Fu Keung
    Sun, Lei
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2012, 59 (07) : 1558 - 1567
  • [3] Optimization of pulse transmission in a high-frequency ultrasound imaging system
    Chen, WH
    Cao, PJ
    Maione, E
    Ritter, T
    Shung, KK
    [J]. 2001 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2001, : 995 - 998
  • [4] Development of a Bipolar Pulse Generator for High-Frequency Ultrasound Imaging System
    Wu, Jian-Xing
    Chen, Pei-Jarn
    Du, Yi-Chun
    Wang, Chien-Lin
    Chen, Guan-Chun
    Chen, Tainsong
    [J]. INFORMATION, COMMUNICATION AND ENGINEERING, 2013, 311 : 217 - +
  • [5] HIGH-FREQUENCY PULSE-COMPRESSION ULTRASOUND IMAGING WITH AN ANNULAR ARRAY
    Mamou, J.
    Ketterling, J. A.
    Silverman, R. H.
    [J]. ACOUSTICAL IMAGING, VOL 29, 2008, 29 : 81 - 86
  • [6] High-frequency ultrasonic arrays for ocular imaging
    Jaeger, M. D.
    Kline-Schoder, R. J.
    Douville, G. M.
    Gagne, J. R.
    Morrison, K. T.
    Audette, W. E.
    Kynor, D. B.
    [J]. MEDICAL IMAGING 2007: ULTRASONIC IMAGING AND SIGNAL PROCESSING, 2007, 6513
  • [7] Coded excitation using biphase-coded pulse with mismatched filters for high-frequency ultrasound imaging
    Hu, Chang-Hong
    Liu, Ruibin
    Zhou, Qifa
    Yen, Jesse
    Shung, K. Kirk
    [J]. ULTRASONICS, 2006, 44 (03) : 330 - 336
  • [8] Contrast Imaging in Mouse Embryos Using High-frequency Ultrasound
    Denbeigh, Janet M.
    Nixon, Brian A.
    Puri, Mira C.
    Foster, F. Stuart
    [J]. JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2015, (97):
  • [9] HIGH-FREQUENCY ULTRASOUND MEASUREMENT OF THE THICKNESS OF THE CORNEAL EPITHELIUM
    REINSTEIN, DZ
    SILVERMAN, RH
    COLEMAN, DJ
    [J]. REFRACTIVE AND CORNEAL SURGERY, 1993, 9 (05): : 385 - 387
  • [10] EXPERIMENTAL OCULAR LESIONS PRODUCED BY HIGH-FREQUENCY ULTRASOUND
    LIZZI, FL
    PACKER, A
    COLEMAN, DJ
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1976, 59 : S59 - S60