Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial

被引:199
|
作者
Myers, Martin G. [1 ]
Godwin, Marshall [2 ]
Dawes, Martin [3 ]
Kiss, Alexander [4 ]
Tobe, Sheldon W.
Grant, F. Curry [5 ]
Kaczorowski, Janusz [6 ]
机构
[1] Univ Toronto, Sunnybrook Hlth Sci Ctr, Schulich Heart Ctr, Div Cardiol,Dept Med, Toronto, ON M4N 3M5, Canada
[2] Mem Univ Newfoundland, Dept Family Med, St John, NF, Canada
[3] McGill Univ, Dept Family Med, Montreal, PQ H3A 2T5, Canada
[4] Sunnybrook Hlth Sci Ctr, Dept Res Design & Biostat, Toronto, ON M4N 3M5, Canada
[5] Quinte Hlth Care, Stroke Prevent Clin, Belleville, ON, Canada
[6] Univ British Columbia, Dept Family Practice, Vancouver, BC V5Z 1M9, Canada
来源
关键词
MEASUREMENTS-WILL; OFFICE; DEVICE; VALIDATION; ALGORITHM; REDUCE;
D O I
10.1136/bmj.d286
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective To compare the quality and accuracy of manual office blood pressure and automated office blood pressure using the awake ambulatory blood pressure as a gold standard. Design Multi-site cluster randomised controlled trial. Setting Primary care practices in five cities in eastern Canada. Participants 555 patients with systolic hypertension and no serious comorbidities under the care of 88 primary care physicians in 67 practices in the community. Interventions Practices were randomly allocated to either ongoing use of manual office blood pressure (control group) or automated office blood pressure (intervention group) using the BpTRU device. The last routine manual office blood pressure (mm Hg) was obtained from each patient's medical record before enrolment. Office blood pressure readings were compared before and after enrolment in the intervention and control groups; all readings were also compared with the awake ambulatory blood pressure. Main outcome measure Difference in systolic blood pressure between awake ambulatory blood pressure minus automated office blood pressure and awake ambulatory blood pressure minus manual office blood pressure. Results Cluster randomisation allocated 31 practices (252 patients) to manual office blood pressure and 36 practices (303 patients) to automated office blood pressure measurement. The most recent routine manual office blood pressure (149.5 (SD 10.8)/81.4 (8.3)) was higher than automated office blood pressure (135.6 (17.3)/77.7 (10.9)) (P<0.001). In the control group, routine manual office blood pressure before enrolment (149.9 (10.7)/81.8 (8.5)) was reduced to 141.4 (14.6)/80.2 (9.5) after enrolment (P<0.001/P=0.01), but the reduction in the intervention group from manual office to automated office blood pressure was significantly greater (P<0.001/P=0.02). On the first study visit after enrolment, the estimated mean difference for the intervention group between the awake ambulatory systolic/diastolic blood pressure and automated office blood pressure (-2.3 (95% confidence interval -0.31 to -4.3)/-3.3 (-2.7 to -4.4)) was less (P=0.006/P=0.26) than the difference in the control group between the awake ambulatory blood pressure and the manual office blood pressure (-6.5 (-4.3 to -8.6)/-4.3 (-2.9 to -5.8)). Systolic/diastolic automated office blood pressure showed a stronger (P<0.001) within group correlation (r=0.34/r=0.56) with awake ambulatory blood pressure after enrolment compared with manual office blood pressure versus awake ambulatory blood pressure before enrolment (r=0.10/r=0.40); the mean difference in r was 0.24 (0.12 to 0.36)/0.16 (0.07 to 0.25)). The between group correlation comparing diastolic automated office blood pressure and awake ambulatory blood pressure (r=0.56) was stronger (P<0.001) than that for manual office blood pressure versus awake ambulatory blood pressure (r=0.30); the mean difference in r was 0.26 (0.09 to 0.41). Digit preference with readings ending in zero was substantially reduced by use of automated office blood pressure. Conclusion In compliant, otherwise healthy, primary care patients with systolic hypertension, introduction of automated office blood pressure measurement into routine primary care significantly reduced the white coat response compared with the ongoing use of manual office blood pressure measurement. The quality and accuracy of automated office blood pressure in relation to the awake ambulatory blood pressure was also significantly better when compared with manual office blood pressure.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Impact of yoga on blood pressure and quality of life in patients with hypertension - a controlled trial in primary care, matched for systolic blood pressure
    Wolff, Moa
    Sundquist, Kristina
    Lonn, Sara Larsson
    Midlov, Patrik
    [J]. BMC CARDIOVASCULAR DISORDERS, 2013, 13 : 1 - 9
  • [2] Impact of yoga on blood pressure and quality of life in patients with hypertension – a controlled trial in primary care, matched for systolic blood pressure
    Moa Wolff
    Kristina Sundquist
    Sara Larsson Lönn
    Patrik Midlöv
    [J]. BMC Cardiovascular Disorders, 13
  • [3] Conventional versus automated measurement of blood pressure in the office (CAMBO) trial
    Myers, Martin G.
    Godwin, Marshall
    Dawes, Martin
    Kiss, Alexander
    Tobe, Sheldon W.
    Kaczorowski, Janusz
    [J]. FAMILY PRACTICE, 2012, 29 (04) : 376 - 382
  • [4] VITAMIN D SUPPLEMENTATION TO REDUCE BLOOD PRESSURE IN OLDER PATIENTS WITH ISOLATED SYSTOLIC HYPERTENSION: A RANDOMISED CONTROLLED TRIAL
    Witham, M. D.
    Price, R. J. G.
    Struthers, A. D.
    Donnan, P. T.
    Ford, I.
    McMurdo, M. E. T.
    [J]. AGE AND AGEING, 2013, 42 : 22 - 22
  • [5] The conventional versus automated measurement of blood pressure in the office (CAMBO) trial: masked hypertension sub-study
    Myers, Martin G.
    Godwin, Marshall
    Dawes, Martin
    Kiss, Alexander
    Tobe, Sheldon W.
    Kaczorowski, Janusz
    [J]. JOURNAL OF HYPERTENSION, 2012, 30 (10) : 1937 - 1941
  • [6] A Cluster Randomised Controlled Trial of Oscillometric Versus Manual Sphygmomanometers for Blood Pressure Management in Primary Care (CRAB)
    Nelson, Mark R.
    Quinn, Stephen
    Bowers-Ingram, Linda
    Nelson, Jan M.
    Winzenberg, Tania M.
    [J]. CIRCULATION, 2008, 118 (18) : S1170 - S1170
  • [7] A cluster randomised controlled trial of an automated versus manual device for blood pressure management (CRAB)
    Nelson, M. R.
    Winzenberg, T. M.
    [J]. HYPERTENSION, 2007, 49 (06) : 1474 - 1474
  • [8] Automated office blood pressure measurement in primary care
    Myers, Martin G.
    Kaczorowski, Janusz
    Dawes, Martin
    Godwin, Marshall
    [J]. CANADIAN FAMILY PHYSICIAN, 2014, 60 (02) : 127 - 132
  • [9] Randomised controlled factorial trial of dietary advice for patients with a single high blood pressure reading in primary care
    Little, P
    Kelly, J
    Barnett, J
    Dorward, M
    Margetts, B
    Warm, D
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 2004, 328 (7447): : 1054 - 1057
  • [10] Automated recruitment and randomisation for an efficient randomised controlled trial in primary care
    Victoria R. Cornelius
    Lisa McDermott
    Alice S. Forster
    Mark Ashworth
    Alison J. Wright
    Martin C. Gulliford
    [J]. Trials, 19