Numerical Analysis for Non-Uniformity of Balloon-Expandable Stent Deployment Driven by Dogboning and Foreshortening

被引:6
|
作者
Rahinj, Ganesh B. [1 ]
Chauhan, Harshit S. [1 ]
Sirivella, Martin L. [1 ]
Satyanarayana, Menta, V [1 ]
Ramanan, Laxminarayanan [1 ]
机构
[1] Sahajanand Med Technol SMT Ltd, Res & Dev Dept, Surat, India
关键词
Dogboning; Foreshortening; Asymmetry; Stent; Balloon; Finite element analysis; FINITE-ELEMENT-ANALYSIS; CORONARY STENTS; DESIGN; ARTERY; RESTENOSIS; EXPANSION; SIMULATION; IMPLICIT; BEHAVIOR; VESSEL;
D O I
10.1007/s13239-021-00573-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose Stenting is the most common intervention for arteriosclerosis treatment; however, the success of the treatment depends on the incidence of in-stent restenosis (ISR). Stent deployment characteristics are major influencers of ISR and can be measured in terms of dogboning, asymmetry, and foreshortening. This study aimed to analyse the implications of balloon and stent-catheter assembly parameters on the stent deployment characteristics. Methods Experimental approach to analyse the impact of the balloon and stent-catheter assembly parameters on stent deployment characteristics is a time-consuming and complex task, whereas numerical methods prove to be quick, efficient, and reliable. In this study, eleven finite element models were employed to analyse non-uniform balloon stent expansion pattern, comprised of variation in, stent axial position on balloon, balloon length, balloon folding pattern, and balloon wall thickness. Results Obtained results suggest that the axially noncentral position of the stent on balloon and variable balloon thickness lead to non-uniform stent deployment pattern. Also, it was proved that variation in balloon length and balloon folding pattern influence deployment process. Conclusion Improved positional accuracies, uniform balloon wall thickness, and selection of the appropriate length of a balloon for selected stent configuration will help to minimize dogboning, asymmetry, and foreshortening during non-uniform stent expansion, thereby reducing the risk of restenosis. The stated numerical approach will be helpful to optimize stent catheter assembly parameters thus minimizing in-vitro tests and product development time.
引用
收藏
页码:247 / 264
页数:18
相关论文
共 35 条
  • [1] Numerical Analysis for Non-Uniformity of Balloon-Expandable Stent Deployment Driven by Dogboning and Foreshortening
    Ganesh B. Rahinj
    Harshit S. Chauhan
    Martin L. Sirivella
    Menta V. Satyanarayana
    Laxminarayanan Ramanan
    Cardiovascular Engineering and Technology, 2022, 13 : 247 - 264
  • [2] Numerical study of the uniformity of balloon-expandable stent deployment
    Mortier, P.
    De Beule, M.
    Carlier, S. G.
    Van Impe, R.
    Verhegghe, B.
    Verdonck, P.
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2008, 130 (02):
  • [3] Finite element analysis of balloon-expandable coronary stent deployment: Influence of angioplasty balloon configuration
    Martin, David
    Boyle, Fergal
    INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2013, 29 (11) : 1161 - 1175
  • [4] Suggestion of Potential Stent Design Parameters to Reduce Restenosis Risk driven by Foreshortening or Dogboning due to Non-uniform Balloon-Stent Expansion
    Dohyung Lim
    Seung-Kwan Cho
    Won-Pil Park
    Anders Kristensson
    Jai-Young Ko
    S. T. S. Al-Hassani
    Han-Sung Kim
    Annals of Biomedical Engineering, 2008, 36 : 1118 - 1129
  • [5] Suggestion of potential stent design parameters to reduce restenosis risk driven by foreshortening or dogboning due to non-uniform balloon-stent expansion
    Lim, Dohyung
    Cho, Seung-Kwan
    Park, Won-Pil
    Kristensson, Anders
    Ko, Jai-Young
    Al-Hassani, S. T. S.
    Kim, Han-Sung
    ANNALS OF BIOMEDICAL ENGINEERING, 2008, 36 (07) : 1118 - 1129
  • [6] Successful Deployment of Two Balloon-Expandable Stents Using an In-Stent Bifurcating Approach
    Cho, Roy J.
    Zamora, Felix D.
    Gibson, Heidi
    Dincer, H. Erhan
    ANNALS OF THE AMERICAN THORACIC SOCIETY, 2018, 15 (07) : 884 - 886
  • [7] Experimental and numerical analysis of the mechanics of balloon-expandable stents
    Martin, H
    Lootz, D
    Behrens, P
    Schmidt, W
    Schmitz, KP
    GESA-SYMPOSIUM 2001: SICHERHEIT UND ZUVERLASSIGKEIT DURCH EXPERIMENTELLE STRUKTUR-UND BEANSPRUCHUNGSANALYSE, 2001, 1599 : 7 - 12
  • [8] Numerical analysis of crimping and inflation process of balloon-expandable coronary stent using implicit solution
    Bukala, Jakub
    Kwiatkowski, Piotr
    Malachowski, Jerzy
    INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2017, 33 (12)
  • [9] Self-expandable stent versus balloon-expandable stent: A serial volumetric analysis by intravascular ultrasound
    Kobayashi, Y
    Teirstein, PS
    Bailey, SR
    Brown, CL
    Christie, LG
    Matthews, RV
    Schwartz, RS
    De Franco, AC
    Handen, CE
    Yock, PG
    Fitzgerald, PJ
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) : 396A - 396A
  • [10] On the numerical investigation of cardiovascular balloon-expandable stent using finite element method
    Azaouzi, M.
    Makradi, A.
    Petit, J.
    Belouettar, S.
    Polit, O.
    COMPUTATIONAL MATERIALS SCIENCE, 2013, 79 : 326 - 335