Catheters and dose optimization using a modified CVT algorithm and multi-criteria optimization in prostate HDR brachytherapy

被引:5
|
作者
Chatigny, Philippe Y. [1 ,2 ,3 ,4 ]
Belanger, Cedric [1 ,2 ,3 ,4 ]
Poulin, Eric [3 ,4 ]
Beaulieu, Luc [1 ,2 ,3 ,4 ]
机构
[1] Univ Laval, Dept Phys Genie Phys & Opt, Quebec City, PQ, Canada
[2] Univ Laval, Ctr Rech Canc, Quebec City, PQ, Canada
[3] Univ Laval, Serv Phys Med & Radioprotect, Ctr Integre Cancerol, CHU Quebec, Quebec City, PQ, Canada
[4] CHU Quebec, Ctr Rech, Quebec City, PQ, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
catheter's optimization; dose optimization; prostate cancer; robustness evaluation; UNCERTAINTIES; PLAN; RECOMMENDATIONS; CONSEQUENCES; BOOST;
D O I
10.1002/mp.15878
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Currently, in high-dose rate (HDR) brachytherapy planning, the catheter's positions are often selected by the planner, which involves the planner's experience. The catheters are then inserted using a template that helps to guide the catheters. For certain applications, it is of interest to choose the optimal location and number of catheters needed for dose coverage and potential decrease of the treatment's toxicity. Hence, it is of great importance to develop patient-specific algorithms for catheters and dose optimization. Methods A modified Centroidal Voronoi tessellation (CVT) algorithm is implemented and merged with a graphics processing unit (GPU)-based multi-criteria optimization algorithm (gMCO). The CVT algorithm optimizes the catheters' positions, and the gMCO algorithm optimizes the dwell times and dwell positions. The CVT algorithm can be used simultaneously for insertion with or without a template. Some improvements to the CVT algorithm are presented such as a new way of considering the area that needs to be covered. One hundred eight previously treated prostates HDR cases using real-time ultrasound are used to evaluate the different optimization procedures. The plan robustness is evaluated using two types of errors: deviations (random) in the insertion and deviation (systematic) in the reconstruction of the catheters. Results Using gMCO on clinically inserted catheter increases the acceptance rate by 37% for Radiation Therapy Oncology Group (RTOG) criteria. Our results show that all the patients respect RTOG criteria with 11 catheters using CVT+gMCO with a template of 5 mm. The number of catheters needed for all patients to respect RTOG criteria with the freehand technique is 10 catheters using CVT+gMCO. When deviations are introduced, using a template, the acceptance rate goes to 85% with 3 mm deviations using 11 catheters. This decrease is less significant when the number of catheters is higher, decreasing by less than 5% with a 3 mm deviation using 13 catheters or more. In conclusion, it is feasible to decrease the number of catheters needed to treat most patients. Conclusions Some cases still need a high number of catheters to reach the plan's criteria. Using gMCO allows an increase in the plan quality, while using CVT reduces the number of catheters. A higher number of catheters equates to plans that are more robust to deviations.
引用
收藏
页码:6575 / 6587
页数:13
相关论文
共 50 条
  • [21] Multi-criteria Optimization of IMRT Plans in Prostate Cancer
    Wala, J. A.
    Craft, D.
    Paly, J. J.
    Efstathiou, J. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S419 - S420
  • [22] Dose-volume objectives in multi-criteria optimization
    Halabi, Tarek
    Craft, David
    Bortfeld, Thomas
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (15): : 3809 - 3818
  • [23] Evaluation of dose distribution parameters in HDR prostate brachytherapy - Influence of dose distribution optimization algorithms
    Kabacinska, R.
    Zuchora, B.
    Wronczewska, A.
    Lebioda, A.
    Makarewicz, R.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S257 - S257
  • [24] Impact of using linear optimization models in dose planning for HDR brachytherapy
    Holm, Asa
    Larsson, Torbjorn
    Tedgren, Asa Carlsson
    MEDICAL PHYSICS, 2012, 39 (02) : 1021 - 1028
  • [25] Genetic algorithm behaviour in the multi-criteria optimization task
    Takahashi, Arita
    International Conference on Knowledge-Based Intelligent Electronic Systems, Proceedings, KES, 2000, 2 : 514 - 519
  • [26] Dual-Source Optimization for HDR Prostate Brachytherapy
    Frank, H.
    Park, S.
    Sheng, K.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [27] Multi-criteria Optimization of neural networks using multi-objective genetic algorithm
    Senhaji, Kaoutar
    Ettaouil, Mohamed
    2017 INTELLIGENT SYSTEMS AND COMPUTER VISION (ISCV), 2017,
  • [28] Temporal optimization of interstitial HDR-brachytherapy of the prostate
    Lamprecht, U
    Kaulich, TW
    Budach, W
    Nüsslin, F
    Bamberg, M
    STRAHLENTHERAPIE UND ONKOLOGIE, 2002, 178 : 54 - 54
  • [29] A multi-criteria optimization for a CCHP with the fuel cell as primary mover using modified Harris Hawks optimization
    Zhao, Li
    Li, Zhiqian
    Chen, Hong
    Li, Jiang
    Xiao, Jie
    Yousefi, Nasser
    ENERGY SOURCES PART A-RECOVERY UTILIZATION AND ENVIRONMENTAL EFFECTS, 2024, 46 (01) : 10630 - 10645
  • [30] Evaluation of hybrid inverse planning and optimization (HIPO) algorithm for optimization in real-time, high-dose-rate (HDR) brachytherapy for prostate
    Pokharel, Shyam
    Rana, Suresh
    Blikenstaff, Joseph
    Sadeghi, Amir
    Prestidge, Bradley
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (04): : 96 - 107