A wearable real-time telemonitoring electrocardiogram device compared with traditional Holter monitoring

被引:4
|
作者
Shen, Qin [1 ]
Li, Jianqing [2 ,3 ]
Cui, Chang [1 ]
Wang, Xingyao [3 ]
Gao, Hongxiang [3 ]
Liu, Chengyu [3 ]
Chen, Minglong [1 ]
机构
[1] Nanjing Med Univ, Dept Cardiol, Affiliated Hosp 1, Nanjing 210029, Jiangsu, Peoples R China
[2] Nanjing Med Univ, Sch Biomed Engn & Informat, Nanjing 211166, Jiangsu, Peoples R China
[3] Southeast Univ, Sch Instrument Sci & Engn, Nanjing 210096, Jiangsu, Peoples R China
来源
JOURNAL OF BIOMEDICAL RESEARCH | 2021年 / 35卷 / 03期
关键词
wearable ECG device; Holter; real-time; remote ECG monitoring; CARDIAC OUTPATIENT TELEMETRY; SILENT ATRIAL-FIBRILLATION; IPHONE ECG; DIAGNOSIS; ACCURACY; PALPITATIONS; PREVENTION; SYSTEM; DETECT;
D O I
10.7555/JBR.34.20200074
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Arrhythmias are very common in the healthy populations as well as patients with cardiovascular diseases. Among them, atrial fibrillation (AF) and malignant ventricular arrhythmias are usually associated with some clinical events. Early diagnosis of arrhythmias, particularly AF and ventricular arrhythmias, is very important for the treatment and prognosis of patients. Holter is a gold standard commonly recommended for noninvasive detection of paroxysmal arrhythmia. However, it has some shortcomings such as fixed detection timings, delayed report and inability of remote real-time detection. To deal with such problems, we designed and applied anew wearable 72-hour triple-lead H3-electrocardiogram (ECG) device with a remote cloud-based ECG platform and an expert-supporting system. In this study, 31 patients were recruited and 24-hour synchronous ECG data by H3-ECG and Holter were recorded. In the H3-ECG group, ECG signals were transmitted using remote real-time modes, and confirmed reports were made by doctors in the remote expert-supporting system, while the traditional modes and detection systems were used in the Holter group. The results showed no significant differences between the two groups in 24-hour total heart rate (HR), averaged HR, maximum HR, minimum HR, premature atrial complexes (PACs) and premature ventricular complexes (PVCs) (P>0.05). The sensitivity and specificity of capture and remote automatic cardiac events detection of PACs, PVCs, and AF by H3-ECG were 93% and 99%, 98% and 99%, 94% and 98%, respectively. Therefore, the long-term limb triple-lead H3-ECG device can be utilized for domiciliary ECG self-monitoring and remote management of patients with common arrhythmia under medical supervision.
引用
收藏
页码:238 / 246
页数:9
相关论文
共 50 条
  • [1] A wearable real-time telemonitoring electrocardiogram device compared with traditional Holter monitoring
    Qin Shen
    Jianqing Li
    Chang Cui
    Xingyao Wang
    Hongxiang Gao
    Chengyu Liu
    Minglong Chen
    [J]. The Journal of Biomedical Research, 2021, 35 (03) : 238 - 246
  • [2] Wearable device for real-time monitoring of human falls
    Lin, Chern-Sheng
    Hsu, Hung Chun
    Lay, Yun-Long
    Chiu, Chuang-Chien
    Chao, Chi-Shih
    [J]. MEASUREMENT, 2007, 40 (9-10) : 831 - 840
  • [3] Wearable Biosignal Monitoring Nodes for Real-time Electrocardiogram and Motion Measurement
    Wong, Kiing-Ing
    Ho, Mel M. S.
    [J]. 2008 5TH INTERNATIONAL SUMMER SCHOOL AND SYMPOSIUM ON MEDICAL DEVICES AND BIOSENSORS, 2008, : 89 - +
  • [4] WEARABLE OPTICAL DEVICE FOR REAL-TIME MONITORING OF NEWBORN JAUNDICE
    Inamori, Go
    Isoda, Yutaka
    Song, Zihao
    Uozumi, Azusa
    Ito, Shuichi
    Ota, Hiroki
    [J]. 2019 IEEE 32ND INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2019, : 541 - 543
  • [5] A Low Power Wearable Device for Real-Time Electrocardiogram Monitoring and Cardiovascular Arrhythmia Detection for Resource Constrained Regions
    Arora, Neha
    Mishra, Biswajit
    Vora, Yash
    [J]. JOURNAL OF LOW POWER ELECTRONICS, 2019, 15 (02) : 234 - 245
  • [6] Real-Time DWT-Based Compression for Wearable Electrocardiogram Monitoring System
    Al-Busaidi, Asiya M.
    Khriji, Lazhar
    Touati, Farid
    Rasid, Mohd Fadlee A.
    Ben Mnaouer, Adel
    [J]. 2015 IEEE 8TH GCC CONFERENCE AND EXHIBITION (GCCCE), 2015,
  • [7] Real-Time Bluetooth Low Energy (BLE) Electrocardiogram Monitoring Device
    Sivanathan, Sivagunalan
    Oleon, Alexandre
    [J]. 2021 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC 2021), 2021,
  • [8] Wearable Device for Yogic Breathing with Real-Time Heart Rate and Posture Monitoring
    Puranik, Anmol
    Kanthi, M.
    Nayak, Anupama V.
    [J]. JOURNAL OF MEDICAL SIGNALS & SENSORS, 2021, 11 (04): : 253 - 261
  • [9] Tracking Tidal Volume From Holter and Wearable Armband Electrocardiogram Monitoring
    Lazaro, Jesus
    Reljin, Natasa
    Bailon, Raquel
    Gil, Eduardo
    Noh, Yeonsik
    Laguna, Pablo
    Chon, Ki H.
    [J]. IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2024, 28 (06) : 3457 - 3465
  • [10] Design of a Novel Wearable Sensor Device for Real-Time Bolted Joints Health Monitoring
    Wang, Chenyu
    Wang, Ning
    Ho, Siu-Chun
    Chen, Xuemin
    Pan, Miao
    Song, Gangbing
    [J]. IEEE INTERNET OF THINGS JOURNAL, 2018, 5 (06): : 5307 - 5316