Diagnostic accuracy of stress first-pass contrast-enhanced myocardial perfusion MRI compared with stress myocardial perfusion scintigraphy

被引:83
|
作者
Sakuma, H
Suzawa, N
Ichikawa, Y
Makino, K
Hirano, T
Kitagawa, K
Takeda, K
机构
[1] Mie Univ Hosp, Dept Radiol, Tsu, Mie 5148507, Japan
[2] Matsusaka Cent Hosp, Dept Radiol, Matsusaka, Mie, Japan
[3] Matsusaka Cent Hosp, Dept Internal Med, Matsusaka, Mie, Japan
关键词
D O I
10.2214/ajr.185.1.01850095
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to determine the diagnostic accuracy of stress perfusion MRI acquired with saturation-recovery prepared turbo fast low-angle shot (turbo FLASH) compared with stress myocardial perfusion scintigraphy. Recent studies show that first-pass contrast-enhanced myocardial perfusion MRI can provide noninvasive detection of low-limiting stenosis in the coronary artery. MATERIALS AND METHODS. First-pass contrast-enhanced MR images were acquired at rest and during stress in 40 patients with suspected coronary artery disease. All patients underwent thallium-201 SPECT without attenuation correction and coronary angiography. Two reviewers independently assigned one of five confidence grades without knowing the results of coronary angiography for receiver operating characteristic (ROC) analysis. Luminal stenosis > 70% on coronary angiography was used as a reference standard. RESULTS. On coronary angiography, 70% or greater diameter stenosis of the coronary artery was observed in 21 (52.5%) of 40 patients. The areas under the ROC curve for detection of significant stenosis in the individual coronary artery were 0.86 (observer 1) and 0.84 (observer 2) for MRI. These values were 0.79 (observer 1, p = not significant) and 0.72 (observer 2, p = not significant) for (201)T1 SPECT. CONCLUSION. The diagnostic accuracy of stress perfusion MRI acquired with saturation-recovery-prepared turbo FLASH was comparable with that of stress 201Tl SPECT. Stress first-pass contrast-enhanced MRI is a noninvasive technique that can be used as an alternative to stress myocardial per-fusion scintigraphy.
引用
下载
收藏
页码:95 / 102
页数:8
相关论文
共 50 条
  • [41] T*2 measurement during first-pass contrast-enhanced cardiac perfusion imaging
    Kellman, Peter
    Aletras, Anthony H.
    Hsu, Li-yueh
    McVeigh, Elliot R.
    Arai, Andrew E.
    MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (05) : 1132 - 1134
  • [42] Myocardial first-pass perfusion assessment using rotational long-axis MRI
    Wang, Y
    Moin, K
    Mathew, ST
    Akinboboye, O
    Reichek, N
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 22 (01) : 53 - 58
  • [43] Quantitative first-pass MRI measures increased myocardial perfusion after vasodilation in mice
    Patrick Antkowiak
    Christopher M Kramer
    Craig H Meyer
    Brent A French
    Frederick H Epstein
    Journal of Cardiovascular Magnetic Resonance, 14 (Suppl 1)
  • [44] Double-bolus method for quantitation of myocardial perfusion by first-pass cardiac MRI
    Christian, TF
    Rettmann, D
    Liao, S
    Balaban, RS
    Arai, AE
    CIRCULATION, 2001, 104 (17) : 639 - 639
  • [45] Reproducibility of first-pass cardiovascular magnetic resonance myocardial perfusion
    Larghat, Abdulghani M.
    Maredia, Neil
    Biglands, John
    Greenwood, John P.
    Ball, Stephen G.
    Jerosch-Herold, Michael
    Radjenovic, Aleksandra
    Plein, Sven
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 37 (04) : 865 - 874
  • [46] Automatic realignment of myocardial first-pass MR perfusion images
    Comte, A
    Lalande, A
    Brunotte, F
    ISPA 2003: PROCEEDINGS OF THE 3RD INTERNATIONAL SYMPOSIUM ON IMAGE AND SIGNAL PROCESSING AND ANALYSIS, PTS 1 AND 2, 2003, : 706 - 710
  • [47] Comparison of Adenosine Stress Myocardial Perfusion Scintigraphy and Oral Dipyridamole Stress Myocardial Perfusion Scintigraphy for Hemodynamic Changes and Adverse Effects
    Yanarates, Ahmet
    Hatipoglu, Filiz
    Akgun, Aysegul
    Kumanlioglu, Kamil
    MEANDROS MEDICAL AND DENTAL JOURNAL, 2016, 17 (02): : 90 - 97
  • [48] Automatic registration of MR first-pass myocardial perfusion images
    Bracoud, L
    Vincent, F
    Pachai, C
    Canet, E
    Croisille, P
    Revel, D
    FUNCTIONAL IMAGING AND MODELING OF THE HEART, PROCEEDINGS, 2003, 2674 : 215 - 223
  • [49] CMR First-Pass Perfusion for Suspected Inducible Myocardial Ischemia
    Hendel, Robert C.
    Friedrich, Matthias G.
    Schulz-Menger, Jeanette
    Zemmrich, Claudia
    Bengel, Frank
    Berman, Daniel S.
    Camici, Paolo G.
    Flamm, Scott D.
    Le Guludec, Dominique
    Kim, Raymond
    Lombardi, Massimo
    Mahmarian, John
    Sechtem, Udo
    Nagel, Eike
    JACC-CARDIOVASCULAR IMAGING, 2016, 9 (11) : 1338 - 1348
  • [50] Quantification of pulmonary perfusion in idiopathic pulmonary fibrosis with first pass dynamic contrast-enhanced perfusion MRI
    Weatherley, Nicholas D.
    Eaden, James A.
    Hughes, Paul J. C.
    Austin, Matthew
    Smith, Laurie
    Bray, Jody
    Marshall, Helen
    Renshaw, Stephen
    Bianchi, Stephen M.
    Wild, Jim M.