Ultrasensitive detection of NDM-1 resistant bacteria based on signal amplification with sandwich-type LNA electrochemical biochips

被引:9
|
作者
Zhang, Liqun [1 ]
Liang, Wenbin [2 ]
Ran, Qiaosheng [1 ]
Liu, Fei [1 ]
Chen, Dong [3 ]
Xiong, Yu [4 ]
Jiang, Dongneng [1 ]
Li, Yi [1 ]
Tu, Dianji [1 ]
Luo, Xiaoge [1 ]
Pu, Xiaoyun [1 ]
Wang, Yunxia [4 ]
机构
[1] Army Med Univ, Xinqiao Hosp, Dept Clin Lab, Chongqing 400037, Peoples R China
[2] Southwest Univ, Key Lab Luminescent & Real Time Analyt Chem, Coll Chem & Chem Engn, Minist Educ, Chongqing 400715, Peoples R China
[3] 302 Hosp PLA, Beijing 100039, Peoples R China
[4] Army Med Univ, Southwest Hosp, Dept Lab Med, Chongqing 400038, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
Electrochemical; NDM-1; Locked nucleic acids (LNAs); Zinc oxide (ZnO); Reduced graphene oxide (rGO); Au nanocages (Au NCs); GRAPHENE;
D O I
10.1016/j.snb.2019.127556
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
New Delhi metallo-beta-lactamase-1 (NDM-1) is an identified metallo-beta-lactamase that confers resistance to carbapenems and all other beta-lactam antibiotics, with the exception of aztreonam. Herein, a novel electrochemical biochip was constructed based on a zinc oxide (ZnO) and reduced graphene oxide (rGO) complex (ZnO@rGO) supported by Au nanocages (Au NCs) for signal amplification for ultrasensitive detection of NDM-1 in early clinical diagnosis. A Au NC and a sulfhydryl (SH) modified locked nucleic acid-1 (Au NC@LNA-1) probe was synthesized with stable Au-S bonds to on the biochip surface. In hybridization solution, Au NCs and LNA-2 probes modified with SH groups formed stable Au NC@LNA-2 complexes. Au NC@LNA-1 and Au NC@LNA-2 were linked with the NDM-1 to form a Au NC@LNA-1/NDM-1/Au NC@LNA-2 "sandwich-like" complex through complementary base pairing, and this complex can bind additional copies of the target sequence to further amplify it's the detection signal. Under optimized conditions, the system showed a linear range from 1 pg L-1 to 100 mu g L-1, and the detection limit was 0.042 pg L-1. The time of biochip probe-modification was 45 min, incubation was 30 min, and electrochemical scanning detection was 1 min. In addition, the results indicate that the specific LNA probe identified single base mismatches and exhibited affinity for the target DNA molecule, resulting in the direct detection of the NDM-1 in complex clinical bacteria samples without PCR amplification. In conclusion, this novel electrochemical biochip method allows for the rapid, specific, and sensitive detection of the NDM-1, indicating its potential application in disease diagnostics.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] rhEPO/EPO discrimination with ultrasensitive electrochemical biosensor based on sandwich-type nano-Au/ZnO sol-gel/nano-Au signal amplification
    Zhang, Liqun
    Wang, Yunxia
    Wang, Jingjing
    Shi, Jianfeng
    Deng, Kun
    Fu, Weiling
    BIOSENSORS & BIOELECTRONICS, 2013, 50 : 217 - 223
  • [22] An ultrasensitive sandwich-type electrochemical immunosensor for carcino embryonie antigen based on supermolecular labeling strategy
    Yan, Zhaoqing
    Ma, Hongmin
    Fan, Dawei
    Hu, Lihua
    Pang, Xuehui
    Gao, Jian
    Wei, Qin
    Wang, Qi
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2016, 781 : 289 - 295
  • [23] A sandwich-type electrochemical immunosensor using polythionine/AuNPs nanocomposites as label for ultrasensitive detection of carcinoembryonic antigen
    Lai, Yuxuan
    Huang, Hao
    Xia, Zhihuan
    Li, Song
    Deng, Yan
    Liu, Xueying
    MATERIALS EXPRESS, 2019, 9 (05) : 444 - 450
  • [24] A sandwich-type electrochemical immunosensor for ultrasensitive detection of CEA based on core-shell Cu2O@Cu-MOF@Au NPs nanostructure attached with HRP for triple signal amplification
    Li, Wenjun
    Yang, Yue
    Ma, Chaoyun
    Song, Yiju
    Hong, Chenglin
    Qiao, Xiuwen
    JOURNAL OF MATERIALS SCIENCE, 2020, 55 (28) : 13980 - 13994
  • [25] An ultrasensitive sandwich-type electrochemical immunosensor based on the signal amplification strategy of mesoporous core-shell Pd@Pt nanoparticles/amino group functionalized graphene nanocomposite
    Li, Mingdang
    Wang, Ping
    Li, Faying
    Chu, Qingyan
    Li, Yueyun
    Dong, Yunhui
    BIOSENSORS & BIOELECTRONICS, 2017, 87 : 752 - 759
  • [26] A sandwich-type electrochemical immunosensor based on spherical nucleic acids-templated Ag nanoclusters for ultrasensitive detection of tumor biomarker
    Chen, Huinan
    Li, Yuanyuan
    Song, Yuchen
    Liu, Fujing
    Deng, Dongmei
    Zhu, Xiaoli
    He, Haibo
    Yan, Xiaoxia
    Luo, Liqiang
    BIOSENSORS & BIOELECTRONICS, 2023, 223
  • [27] Molecular Detection of New Delhi Metallo-Beta-Lactamase-1 (NDM-1) Positive Bacteria from Environmental and Drinking Water Samples by Loop Mediated Isothermal Amplification of bla NDM-1
    Rathinasabapathi, P.
    Hiremath, Deepak S.
    Arunraj, Rex
    Parani, M.
    INDIAN JOURNAL OF MICROBIOLOGY, 2015, 55 (04) : 400 - 405
  • [28] Sandwich-type electrochemical aptasensor for sensitive detection of myoglobin based on Pt@CuCo-oxide nanoparticles as a signal marker
    Yu, Yueying
    Zhao, Hongli
    Chen, Kaicha
    Cao, Shida
    Lan, Minbo
    TALANTA, 2024, 272
  • [29] A sensitive sandwich-type electrochemical aptasensor for thrombin detection based on platinum nanoparticles decorated carbon nanocages as signal labels
    Gao, Fenglei
    Du, Lili
    Zhang, Yu
    Zhou, Fuyi
    Tang, Daoquan
    BIOSENSORS & BIOELECTRONICS, 2016, 86 : 185 - 193
  • [30] Electrochemical sandwich-type thrombin aptasensor based on dual signal amplification strategy of silver nanowires and hollow Au-CeO2
    Zhang, Qiaoxia
    Fan, Gaochao
    Chen, Wei
    Liu, Qing
    Zhang, Xiao
    Zhang, Xianxi
    Liu, Qingyun
    BIOSENSORS & BIOELECTRONICS, 2020, 150 (150):