Performance of a novel collimator for high-sensitivity brain SPECT

被引:8
|
作者
El Fakhri, G
Ouyang, J
Zimmerman, RE
Fischman, AJ
Kijewski, MF
机构
[1] Harvard Univ, Sch Med, Dept Radiol, Boston, MA 02115 USA
[2] Brigham & Womens Hosp, Boston, MA 02115 USA
[3] Massachusetts Gen Hosp, Boston, MA 02115 USA
关键词
high-sensitivity brain SPECT; variable focusing collimator; estimation; and detection SNR;
D O I
10.1118/1.2143140
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We assessed improvements in performance in detection and estimation tasks due to a novel brain single photon computed tomography collimator. Data were acquired on the CeraSPECT (TM) scanner using both new and standard collimators. The new variable focusing collimator SensOgrade (TM) samples the projections unequally, with central regions more heavily represented, to compensate for attenuation of counts from central brain structures. Furthermore, it utilizes more of the cylindrical crystal surface. Two phantom studies were performed. The first phantom was a 21-cm-diameter cylindrical background containing nine spheres ranging from 0.5 to 5 cm(3) in volume. Tc-99m sphere to background activity ratio was 10:1. Twenty-nine 10-min datasets were acquired with each collimator. The second phantom was the Radiology Support Devices (Long Beach, CA) striatal phantom with striatal-background ratios of 10:1 on the left and 5:1 on the right. Twenty-nine 4-min datasets were acquired with each collimator. Perfusion imaging using Tc-99m-HMPAO was also performed in three healthy volunteers using both collimators under identical simulations. Projections were reconstructed by filtered backprojection with an unwindowed ramp filter. The nonprewhitening matched filter signal-to-noise ratio (NPW-SNR) was computed as a surrogate for human performance in detecting spherical lesions. Sphere activity concentration, radius, and location coordinates were simultaneously estimated by fitting images to an assumed model using an iterative nonlinear algorithm. Resolution recovery was implicit in the estimation procedure, as the point spread function was incorporated into the model. NPW-SNR for sphere detection was 1.5 to 2 times greater with the new collimator; for the striatal phantom the improvement in SNR was 54%. The SNR for estimating sphere activity concentration improved by 46 to 89% for spheres located more than 5 ern from the phantom center. Images acquired with the standard collimator were too noisy in the central regions to allow estimation of sphere activity. In Tc-99m-HMPAO human studies, SNR was improved by 21 to 41 % in the cortex, 66% in the basal ganglia, and 74% in the thalamus. The new collimator leads to substantially improved detection and estimation performance throughout the brain. The higher sensitivity will be particularly important for dynamic imaging. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:209 / 215
页数:7
相关论文
共 50 条
  • [41] Performance analysis of a high-sensitivity multi-pinhole cardiac SPECT system with hemi-ellipsoid detectors
    Bhusal, Narayan
    Dey, Joyoni
    Xu, Jingzhu
    Kalluri, Kesava
    Konik, Arda
    Mukherjee, Joyeeta M.
    Pretorius, P. Hendrik
    MEDICAL PHYSICS, 2019, 46 (01) : 116 - 126
  • [42] Compared Performance of High-Sensitivity Cameras Dedicated to Myocardial Perfusion SPECT: A Comprehensive Analysis of Phantom and Human Images
    Imbert, Laetitia
    Poussier, Sylvain
    Franken, Philippe R.
    Songy, Bernard
    Verger, Antoine
    Morel, Olivier
    Wolf, Didier
    Noel, Alain
    Karcher, Gilles
    Marie, Pierre-Yves
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (12) : 1897 - 1903
  • [43] Diagnostic Performance of a Novel High-sensitivity Cardiac Troponin I Assay: The Westcor Study
    Tjora, Hilde Lunde L.
    Steiro, Ole-Thomas
    Langorgen, Jorund
    Nygard, Ottar K.
    Bjorneklett, Rune O.
    Renate, Renstrom
    Skadberg, Oyvind
    Bonarjee, Vernon V. S. V.
    Hunt, Clare
    Collinson, Paul
    Omland, Torbjorn
    Vikenes, Kjell
    Aakre, Kristin Moberg
    CIRCULATION, 2019, 140
  • [44] Collimator design for a clinical brain SPECT/MRI insert
    Debora Salvado
    Kjell Erlandsson
    Alexandre Bousse
    Michele Occhipinti
    Paolo Busca
    Carlo Fiorini
    Brian Hutton
    EJNMMI Physics, 1 (Suppl 1)
  • [45] Multipinhole collimator with 20 apertures for a brain SPECT application
    Lee, Tzu-Cheng
    Ellin, Justin R.
    Huang, Qiu
    Shrestha, Uttam
    Gullberg, Grant T.
    Seo, Youngho
    MEDICAL PHYSICS, 2014, 41 (11)
  • [46] Collimator design for a multipinhole brain SPECT insert for MRI
    Van Audenhaege, Karen
    Van Holen, Roel
    Vanhove, Christian
    Vandenberghe, Stefaan
    MEDICAL PHYSICS, 2015, 42 (11) : 6679 - 6689
  • [47] Visualization of the scatter radiation in real time using a high-sensitivity CMOS camera with pinhole collimator
    Lee, Hyojin
    Fujibuchi, Toshioh
    Han, Donghee
    Arakawa, Hiroyuki
    HEALTH AND TECHNOLOGY, 2025,
  • [48] In vivo biodistribution of stable spherical and filamentous micelles probed by high-sensitivity SPECT
    Jennings, L.
    Ivashchenko, O.
    Marsman, I. J. C.
    Laan, A. C.
    Denkova, A. G.
    Waton, G.
    Beekman, F. J.
    Schosseler, F.
    Mendes, E.
    BIOMATERIALS SCIENCE, 2016, 4 (08) : 1202 - 1211
  • [49] High-Sensitivity Force Sensors Based on Novel Materials
    Zhang, Zijian
    Xu, Nan
    Huang, Zuan
    Lai, Jianchun
    Liu, Jianing
    Deng, Guangwei
    Wang, Xinhe
    Zhao, Weisheng
    Advanced Devices and Instrumentation, 2023, 4
  • [50] A novel, high-sensitivity, small specimen creep test
    Hyde, T. H.
    Sun, W.
    JOURNAL OF STRAIN ANALYSIS FOR ENGINEERING DESIGN, 2009, 44 (03): : 171 - 185