A comparison of 18 different x-ray detectors currently used in dentistry

被引:103
|
作者
Farman, AG [1 ]
Farman, TT [1 ]
机构
[1] Univ Louisville, Sch Dent, Dept Surg & Hosp Dent, Louisville, KY 40292 USA
关键词
D O I
10.1016/j.tripleo.2004.04.002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose. There has been a proliferation of available dental x-ray detectors over the recent past. The purpose of this short technical report is to provide a basic comparison of spatial resolution, contrast perceptibility, and relative exposure latitudes of 18 current dental x-ray detectors, including solid-state systems (CCD and CMOS), photostimulable phosphors, and analog film. Methods. Spatial resolution was measured using a 0.025 mm Pb phantom test grid with a measurement range from 1.5 to 20 lp/mm. For contrast perceptibility, a 7-mm thick aluminum perceptibility test device with wells of 0.1-0.9 mm depth at 0.1 mm intervals and I defect of 1.5 mm was used. The relative exposure latitude was determined by expert consensus using clear discrimination of the enamel-dentin junction as the lower limit and pixel blooming or unacceptable levels of cervical burnout as the upper limit. Results. The highest spatial resolution was found with Kodak InSight film, RVG-ui (CCD), and RVG 6000 (CMOS) detectors, each of which achieved 20 lp/mm, followed by the Planmeca Dixi2 v3 at >= 16 lp/mm. Contrast resolution was at least to 0.2 mm through 7 mm aluminum for all 18 detectors, with the best results being found for the Visualix HDI, RVG-ui, RVG 5000, and RVG 6000 detectors and the Schick CDR wired and wireless systems. The greatest exposure ranges were found with photostimulable phosphors and the Kodak RVG 6000 and RVG 5000 detectors. Conclusions. Most current x-ray detectors generally perform well in terms of spatial and contrast resolutions, and in terms of exposure latitude. These findings were independent of the modality applied.
引用
收藏
页码:485 / 489
页数:5
相关论文
共 50 条
  • [21] INEXPENSIVE X-RAY DETECTORS
    ARNDT, UW
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1954, 31 (05): : 192 - 193
  • [22] PROBLEMS OF X-RAY DETECTORS
    PASCAL, A
    JOURNAL DE MICROSCOPIE, 1967, 6 (07): : 881 - +
  • [23] Superconducting X-ray detectors
    Can Yang
    Mengting Si
    Lixing You
    Science China Information Sciences, 2020, 63
  • [24] INEXPENSIVE X-RAY DETECTORS
    YATES, EL
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1954, 31 (08): : 306 - 306
  • [25] X-RAY imaging detectors
    Gruner, Sol M.
    PHYSICS TODAY, 2012, 65 (12) : 29 - 34
  • [26] Plastic x-ray detectors
    Westgate, Tom
    CHEMISTRY WORLD, 2008, 5 (01): : 29 - 29
  • [27] Superconducting X-ray detectors
    Yang, Can
    Si, Mengting
    You, Lixing
    SCIENCE CHINA-INFORMATION SCIENCES, 2020, 63 (08)
  • [28] Innovations in X-Ray Detectors
    Li, Ke
    Zhao, W.
    Antonuk, L.
    Ullberg, C.
    MEDICAL PHYSICS, 2019, 46 (06) : E487 - E487
  • [29] Superconducting X-ray detectors
    Can YANG
    Mengting SI
    Lixing YOU
    ScienceChina(InformationSciences), 2020, 63 (08) : 91 - 109
  • [30] X-ray opaque waveguide for dentistry
    Dostálová, T
    Jelínková, H
    Nemec, M
    Sulc, J
    Miyagi, M
    Shi, YW
    Matsuura, Y
    PHOTOMEDICINE AND LASER SURGERY, 2004, 22 (03) : 221 - 225