Amperometric biosensor for the determination of histamine in fish samples

被引:68
|
作者
Perez, Sandra [1 ]
Bartroli, Jordi [1 ]
Fabregas, Esteve [1 ]
机构
[1] Univ Autonoma Barcelona, Dept Chem, Sensors & Biosensors Grp, E-08193 Barcelona, Spain
关键词
Amperometric biosensor; Histamine; Diamine oxidase; Fish; Polysulfone; Carbon nanotubes; CHROMATOGRAPHY-MASS SPECTROMETRY; BIOGENIC-AMINES; ELECTROCHEMICAL BIOSENSOR; DIAMINE OXIDASE; ENZYME SENSOR; FOOD SAMPLES; ELECTROPHORESIS; WINE; QUANTIFICATION; MEMBRANES;
D O I
10.1016/j.foodchem.2013.06.125
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
A bienzymatic biosensor employing diamine oxidase (DOx) and horseradish peroxidase (HRP) for the detection of histamine in fish samples has been developed and optimized in this work. These enzymes have been co-immobilized into a polysulfone/carbon nanotubes/ferrocene membrane by means of phase inversion technique onto screen-printed electrodes. The electrochemical measurements have been carried out in phosphate buffer solution at pH 8.0 in batch mode and low applied potential (-50 mV vs. Ag/AgCl, KCl 0.1 M) to minimize the interferences. Developed biosensor exhibits high sensitivity (1.9 x 10(7) nA(-1)), low limit of detection (1.7 x 10(-7) M), high storage stability and excellent reproducibility, obtaining a linear interval range from 3 x 10(-7) to 2 x 10(-5) M. Finally, applicability of the biosensor to the estimation of histamine content in different fish samples has been assessed; obtaining a good correlation between results obtained with the biosensor and those obtained with the reference method (ELISA) in case of sardines, mackerel and greater weever. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:4066 / 4072
页数:7
相关论文
共 50 条
  • [21] Amperometric Acetylcholinesterase Biosensor for Physostigmine Determination
    Davletshina, R. R.
    Ivanov, A. N.
    Evtugyn, G. A.
    [J]. UCHENYE ZAPISKI KAZANSKOGO UNIVERSITETA-SERIYA ESTESTVENNYE NAUKI, 2018, 160 (01): : 5 - 16
  • [22] Mediated Amperometric Biosensor for the Determination of Ammonium
    Abd Rahman, Samsulida
    Abdullah, Jaafar
    Sidek, Hamidah
    Azmi, Nur Ellina
    [J]. ANALYTICAL & BIOANALYTICAL ELECTROCHEMISTRY, 2012, 4 (03): : 262 - 276
  • [24] Validation Study of a HistaSure™ ELISAFast Track for the Determination of Histamine in Fish Samples
    Manz, Georg
    Booltink, Esther
    [J]. JOURNAL OF AOAC INTERNATIONAL, 2014, 97 (06) : 1601 - 1614
  • [25] Validation and comparison of analytical methods for the determination of histamine in tuna fish samples
    Cinquina, AL
    Longo, F
    Calì, A
    De Santis, L
    Baccelliere, R
    Cozzani, R
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2004, 1032 (1-2) : 79 - 85
  • [26] Comparison of analytical methods for the determination of histamine in reference canned fish samples
    Jaksic, S.
    Balos, M. Z.
    Mihaljev, Z.
    Radulovic, J. Prodanov
    Nesic, K.
    [J]. 59TH INTERNATIONAL MEAT INDUSTRY CONFERENCE MEATCON2017, 2017, 85
  • [27] Development of a Catalase-Based Amperometric Biosensor for the Determination of Increased Catalase Content in Milk Samples
    Futo, Peter
    Markus, Gabriella
    Kiss, Attila
    Adanyi, Nora
    [J]. ELECTROANALYSIS, 2012, 24 (01) : 107 - 113
  • [28] Novel Amperometric Tri-Enzyme Biosensor for Lysine Determination in Pharmaceutical Products and Food Samples
    Boka, Beata
    Korozs, Marietta
    Nanasi, Marta
    Adanyi, Nora
    [J]. ELECTROANALYSIS, 2015, 27 (03) : 817 - 824
  • [29] Enzyme Based Amperometric Biosensor for Adenine Determination
    Boka, Beata
    Adanyi, Nora
    Virag, Diana
    Frebort, Ivo
    Kiss, Attila
    [J]. ELECTROANALYSIS, 2013, 25 (01) : 237 - 242
  • [30] Amperometric determination of glucose with a carbon paste biosensor
    Schachl, K
    Turkusic, E
    Komersová, A
    Bartos, M
    Moderegger, H
    Svancara, I
    Alemu, H
    Vytras, K
    Jimenez-Castro, M
    Kalcher, K
    [J]. COLLECTION OF CZECHOSLOVAK CHEMICAL COMMUNICATIONS, 2002, 67 (03) : 302 - 313