Utilization of Adaptive-coefficient Estimation Method for Motion Artifacts Reduction from Photoplethysmographic Signals

被引:0
|
作者
Ram, M. Raghu [1 ]
Sivani, K. [2 ]
Reddy, K. Ashoka [3 ]
机构
[1] Kakatiya Inst Technol & Sci, Dept Elect & Instrumentat Engn, Warangal, Telangana, India
[2] Kakatiya Inst Technol & Sci, Elect & Instrumentat Engn, Warangal, Telangana, India
[3] Kakatiya Inst Technol & Sci, Dept Elect & Commun Engn, Warangal, Telangana, India
关键词
Pulse oximeter; PPG; MA noise; co-efficient estimation; SpO(2);
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Modern pulse oximeters are employed in critical care units to measure the vital medical parameters like heart rate and blood oxygen saturation levels. These devices work by acquiring the photoplethysmographic (PPG) data utilizing PPG sensors attached to finger/earlobe/forehead of the patient and then computing the ratio of amplitudes of the red and IR PPG signals. Further the ratio of amplitudes is used to estimate oxygen saturation levels with the help of calibration curve. Patient movements while recording the pulse oximeter data may result in erroneous estimation of medical data and may result in wrong diagnosis by the clinician. Data corruption in recorded PPG signals is referred as Motion artifact (MA) corruption and its detection and reduction is a crucial problem for most of the researchers. It is clear evident that detection and reduction of MA component from recorded PPG data by any signal processing technique may guarantee error-free estimation of oxygen saturation in arterial blood (SpO(2)). In this work, we present a adaptive coefficient estimation technique to detect MA components from quasi periodic natured PPG signal and then deduce MA reduced PPG signal. Fourier coefficients are estimated using basic least mean squares algorithm. The novelty of the proposed technique lies in detection and reduction of MA noise by estimating the Fourier coefficients and then based on randomness measures considering only the required number of Fourier coefficients dynamically to generate MA reduced PPG signal. SpO(2) is estimated from MA reduced PPGs by utilizing the calibration curve. The superiority of proposed technique is been proved by comparing the experimental results with results obtained using basic least mean squares (LMS) method. PPG data recorded with different MA (Vertical, Horizontal and Bending movements of patient's finger) is considered for experiment analysis. Obtained SpO(2) parameter calculations proved the efficacy of estimation technique in measurement of reliable and accurate SpO(2,) helpful for medical diagnosis.
引用
收藏
页码:818 / 822
页数:5
相关论文
共 50 条
  • [1] E2MD for Reduction of Motion Artifacts from Photoplethysmographic Signals
    Raghuram, M.
    Sivani, Kosaraju
    Reddy, K. Ashoka
    [J]. 2014 INTERNATIONAL CONFERENCE ON ELECTRONICS AND COMMUNICATION SYSTEMS (ICECS), 2014,
  • [2] Characterization and Reduction of Motion Artifacts in Photoplethysmographic Signals from a Wrist-worn Device
    Tautan, Alexandra-Maria
    Young, Alex
    Wentink, Eva
    Wieringa, Fokko
    [J]. 2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 6146 - 6149
  • [3] HHT based signal decomposition for reduction of motion artifacts in Photoplethysmographic signals
    Raghuram, M.
    Madhav, K. Venu
    Krishna, E. Hari
    Komalla, Nagarjuna Reddy
    Sivani, Kosaraju
    Reddy, K. Ashoka
    [J]. 2012 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2012, : 1730 - 1734
  • [4] On the performance of wavelets in reducing motion artifacts from photoplethysmographic signals
    Raghuram, M.
    Madhav, K. Venu
    Krishna, E. Hari
    Reddy, K. Ashoka
    [J]. 2010 4TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING (ICBBE 2010), 2010,
  • [5] AFODSS: Heart Rate Estimation Method From Photoplethysmographic Signals With Motion Artifacts Using Fourier-Sparse Dual Optimization
    Raj, Remya
    Selvakumar, J.
    Maik, Vivek
    [J]. IEEE SENSORS JOURNAL, 2019, 19 (21) : 9953 - 9963
  • [6] An Automatic Method for Motion Artifacts Detection in Photoplethysmographic Signals Referenced With Electrocardiography Data
    Vaz, Pedro
    Henriques, Jorge
    de Carvalho, Paulo
    Couceiro, Ricardo
    [J]. 2014 7TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2014), 2014, : 704 - 708
  • [7] A comb filter based signal processing method to effectively reduce motion artifacts from photoplethysmographic signals
    Peng, Fulai
    Liu, Hongyun
    Wang, Weidong
    [J]. PHYSIOLOGICAL MEASUREMENT, 2015, 36 (10) : 2159 - 2170
  • [8] Determination of Absolute Heart Beat from Photoplethysmographic Signals in the Presence of Motion Artifacts
    Karna, Viswavardhan Reddy
    Kumar, Navin
    [J]. 2018 SECOND INTERNATIONAL CONFERENCE ON ADVANCES IN ELECTRONICS, COMPUTERS AND COMMUNICATIONS (ICAECC), 2018,
  • [9] Influence of Motion Artifacts on Photoplethysmographic Signals for Measuring Pulse Rates
    Lee, Yunjoo
    Han, Hyonyoung
    Kim, Jung
    [J]. 2008 INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS, VOLS 1-4, 2008, : 854 - 857
  • [10] Reduction of motion artifacts from photoplethysmographic recordings using a wavelet denoising approach
    Lee, CM
    Zhang, YT
    [J]. IEEE EMBS APBME 2003, 2003, : 194 - 195