Amperometric biosensor based on D-aminoacid oxidase for the R-perindopril assay

被引:21
|
作者
van Staden, AF [1 ]
Stefan, RI
Aboul-Enein, HY
机构
[1] Univ Pretoria, Dept Chem, ZA-0002 Pretoria, South Africa
[2] King Faisal Specialist Hosp & Res Ctr, Bioanalyt & Drug Dev Lab, Biol & Med Res Dept MBC03, Riyadh 11211, Saudi Arabia
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 2000年 / 367卷 / 02期
关键词
D O I
10.1007/s002160051620
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new amperometric biosensor based on D-aminoacid oxidase is described for the assay of R-perindopril. R-perindopril can be determined in the 400-20 nmol/L concentration range; the detection limit is 10 nmol/L. The selectivity was checked with S-perindopril, D- and L-proline, and polyvinylpyrrolidone. The main interfering species was D-proline. An automated system for the assay of R-perindopril based on the concept of flow injection with an amperometric biosensor (based on D-aminoacid oxidase) as detector is also described. The system is suitable for the on-line monitoring of R-perindopril at a sampling rate of 72 samples/h, in the Linear range: 100 nmol/L -20 nmol/L with an RSD better than 0.09% (n = 10).
引用
收藏
页码:178 / 180
页数:3