A novel method for simultaneous 3D B1 and T1 mapping: the method of slopes (MoS)

被引:21
|
作者
Chavez, Sofia [1 ]
Stanisz, Greg J. [2 ,3 ]
机构
[1] Ctr Addict & Mental Hlth, Res Imaging Ctr, Toronto, ON M5T 1R8, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
[3] Sunnybrook Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
基金
加拿大健康研究院;
关键词
B1; inhomogeneities; T1; mapping; SPGR; flip angle; spoiled-gradient-echo; RF-FIELD INHOMOGENEITIES; IN-VIVO; T-2; RELAXATION; STEADY-STATE; ACCURATE; RADIOFREQUENCY; MRI;
D O I
10.1002/nbm.2769
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A novel three-dimensional simultaneous B1 and T1 mapping method is introduced: the method of slopes (MoS). The linearity of the spoiled gradient recalled echo (SPGR) signal vs flip angle relation is exploited: B1 mapping is achieved by a two-point extrapolation to signal null with a correction scheme while T1 mapping uses the slopes of the SPGR signal vs flip angle curves near the origin and near the signal null. This new method improves upon the existing variable flip angle (VFA) T1-mapping method in that (i) consistency between B1 and T1 maps is ensured (ii) the sampling scheme is T1-independent (iii) the noise bias and singularity, associated with using a linear form for the SPGR signal equation, is eliminated by using the full equation. The method is shown to yield accurate and robust results via simulations. Initial estimates of B1 and T1 values are obtained from three data points via simple computations and straight line approximations. Initial estimates of B1 values, for a range of values, are shown to be accurate due to the proposed B1 correction scheme. The accuracy and robustness of T1 values is achieved via a non-linear fitting algorithm which includes a fourth data point sampled at high SNR. The MoS was validated by comparing resulting B1 and T1 maps with those obtained using other standard methods. Finally, the ability to obtain brain B1 and T1 maps using the MoS was demonstrated by in vivo experiments. The MoS is expected to perform well on other motion-free anatomical regions as well. Copyright (c) 2012 John Wiley & Sons, Ltd.
引用
收藏
页码:1043 / 1055
页数:13
相关论文
共 50 条
  • [41] True T1 mapping with SMART1Map (saturation method using adaptive recovery times for cardiac T1 mapping): a comparison with MOLLI
    Glenn S Slavin
    Jeff A Stainsby
    [J]. Journal of Cardiovascular Magnetic Resonance, 15 (Suppl 1)
  • [42] Simultaneous T1 and T2 mapping of the carotid plaque (SIMPLE) with T2 and inversion recovery prepared 3D radial imaging
    Qi, Haikun
    Sun, Jie
    Qiao, Huiyu
    Zhao, Xihai
    Guo, Rui
    Balu, Niranjan
    Yuan, Chun
    Chen, Huijun
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (06) : 2598 - 2608
  • [43] Simultaneous variable flip angle-actual flip angle imaging method for improved accuracy and precision of three-dimensional T1 and B1 measurements
    Hurley, Samuel A.
    Yarnykh, Vasily L.
    Johnson, Kevin M.
    Field, Aaron S.
    Alexander, Andrew L.
    Samsonov, Alexey A.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (01) : 54 - 64
  • [44] A FAST METHOD FOR T1 FITTING
    SPERBER, GO
    ERICSSON, A
    HEMMINGSSON, A
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1989, 9 (01) : 113 - 117
  • [45] MR fingerprinting for rapid simultaneous T1, T2, and T1ρ relaxation mapping of the human articular cartilage at 3T
    Sharafi, Azadeh
    Zibetti, Marcelo V. W.
    Chang, Gregory
    Cloos, Martijn
    Regatte, Ravinder R.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (05) : 2636 - 2644
  • [46] TESTING THE T1 METHOD - A COMPARISON OF T1(MIN) AND STRUCTURAL DATA
    LUO, XL
    CRABTREE, RH
    [J]. INORGANIC CHEMISTRY, 1990, 29 (15) : 2788 - 2791
  • [47] Comparison of T1 mapping and fixed T1 method for dynamic contrast-enhanced MRI perfusion in brain gliomas
    Conte, G. M.
    Altabella, L.
    Castellano, A.
    Cuccarini, V.
    Bizzi, A.
    Grimaldi, M.
    Costa, A.
    Caulo, M.
    Falini, A.
    Anzalone, N.
    [J]. EUROPEAN RADIOLOGY, 2019, 29 (07) : 3467 - 3479
  • [48] Recurrent neural network-based simultaneous cardiac T1, T2, and T1ρ mapping
    Tao, Yiming
    Lv, Zhenfeng
    Liu, Wenjian
    Qi, Haikun
    Hu, Peng
    [J]. NMR IN BIOMEDICINE, 2024, 37 (08)
  • [49] Simultaneous T1, T2, and T1ρ relaxation mapping of the lower leg muscle with MR fingerprinting
    Sharafi, Azadeh
    Medina, Katherine
    Zibetti, Marcelo W., V
    Rao, Smita
    Cloos, Martijn A.
    Brown, Ryan
    Regatte, Ravinder R.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (01) : 372 - 381
  • [50] Comparison of T1 mapping and fixed T1 method for dynamic contrast-enhanced MRI perfusion in brain gliomas
    G. M. Conte
    L. Altabella
    A. Castellano
    V. Cuccarini
    A. Bizzi
    M. Grimaldi
    A. Costa
    M. Caulo
    A. Falini
    N. Anzalone
    [J]. European Radiology, 2019, 29 : 3467 - 3479